3B Medical 3B AER X User Guide
Quick Start Guide
- Place the Aer X into the Aer X bag, ensure the filter inlet is not covered and air can flow freely.
- Slide battery into battery compartment by opening cover located on the top of your Aer X. The charging port is located just above the filter inlet. Your Aer X may be charged during use.
- Insert the power cord into the Aer X if not using battery power or when charging battery.
- Connect cannula to the oxygen port.NOTE: the oxygen port connects magnetically, this is for your safety (figure A)
- Place the nasal cannula in a comfortable position, with the prongs inside your nostrils.
- Press the blue Power button to begin oxygen flow (figure B)
- An extended beep indicates your device is powering on. Please wait until start up is complete before changing any settings.
- Your Aer X will now be providing oxygen when you inhale
QUICK TROUBLE SHOOTING TIPS
If the 3B™ AER X should shut down during start up or, the Yellow light is illuminated after start up;
- Check to make sure the 3B AER X cannula is attached. USING A DIFFERENT CANNULA THAN THE CANNULA SHIPPED WITH YOUR 3B AER or HAVING NO CANNULA ATTACHED may cause the 3B AER X to shut down during startup.
- Make sure that the battery is charged. A DEPLETED BATTERY will cause the 3B AER X to shut down during startup.
- Power down your 3B AER X and repeat the startup process.If the 3B AER X fails to start again;
- Disconnect the battery and connect the power supply.
- Repeat the startup process.If the 3B™ AER X continues to fail to start up, please contact your equipment provider for assistance.
[xyz-ips snippet=”download-snippet”]