3B Medical 3B AER X User Guide

3B Medical 3B AER X User Guide

 

Quick Start Guide

  1. Place the Aer X into the Aer X bag, ensure the filter inlet is not covered and air can flow freely.                                                                                     
  2. Slide battery into battery compartment by opening cover located on the top of your Aer X. The charging port is located just above the filter inlet. Your Aer X may be charged during use. 
  3. Insert the power cord into the Aer X if not using battery power or when charging battery.
  4. Connect cannula to the oxygen port.NOTE: the oxygen port connects magnetically, this is for your safety (figure A)
  5. Place the nasal cannula in a comfortable position, with the prongs inside your nostrils.
  6. Press the blue Power button to begin oxygen flow (figure B)
  7. An extended beep indicates your device is powering on. Please wait until start up is complete before changing any settings.
  8. Your Aer X will now be providing oxygen when you inhale

 

QUICK TROUBLE SHOOTING TIPS

If the 3B™ AER X should shut down during start up or, the Yellow light is illuminated after start up;

  1. Check to make sure the 3B AER X cannula is attached. USING A DIFFERENT CANNULA THAN THE CANNULA SHIPPED WITH YOUR 3B AER or HAVING NO CANNULA ATTACHED may cause the 3B AER X to shut down during startup.
  2. Make sure that the battery is charged. A DEPLETED BATTERY will cause the 3B AER X to shut down during startup.
  3. Power down your 3B AER X and repeat the startup process.If the 3B AER X fails to start again;
  4. Disconnect the battery and connect the power supply.
  5. Repeat the startup process.If the 3B™ AER X continues to fail to start up, please contact your equipment provider for assistance.

 

[xyz-ips snippet=”download-snippet”]