BW GasAlertMicro 5/PID/IR User Manual

GasAlertMicro 5/PID/IR

 

1, 2, 3, 4, 5 Gas DetectorUser Manual

Limited Warranty and Limitation LiabilityBW Technologies LP (BW) warrants the product to be free from defects in material and workmanship under normal use and service for a period of two years, beginning on the date of shipment to the buyer. This warranty extends only to the sale of new and unused products to the original buyer. BW’s warranty obligation is limited, at BW’s option, to refund of the purchase price, repair or replacement of a defective product that is returned to a BW authorized service center within the warranty period. In no event shall BW’s liability hereunder exceed the purchase price actually paid by the buyer for the Product. This warranty does not include:a) fuses, disposable batteries or the routine replacement of parts due to the normal wear and tear of the product arising from use; b) any product which in BW’s opinion, has been misused, altered, neglected or damaged, by accident or abnormal conditions of operation, handling or use; c) any damage or defects attributable to repair of the product by any person other than an authorized dealer, or the installation of unapproved parts on theproduct; or The obligations set forth in this warranty are conditional on:a) proper storage, installation, calibration, use, maintenance and compliance with the product manual instructions and any other applicable recommendations of BW;b) the buyer promptly notifying BW of any defect and, if required, promptly making the product available for correction. No goods shall be returned to BW until receipt by the buyer of shipping instructions from BW; andc) the right of BW to require that the buyer provide proof of purchase such as the original invoice, bill of sale or packing slip to establish that the product is within the warranty period.THE BUYER AGREES THAT THIS WARRANTY IS THE BUYER’S SOLE AND EXCLUSIVE REMEDY AND IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESS OR IMPLIED, INCLUDING BUT NOT LIMITED TO ANY IMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. BW SHALL NOT BE LIABLE FOR ANY SPECIAL, INDIRECT, INCIDENTAL, OR BASED ON CONTRACT, TORT OR RELIANCE OR ANY OTHER THEORY. Since some countries or states do not allow limitation of the term of an implied warranty, or exclusion or limitation of incidental or consequential damages, the limitations and exclusions of this warranty may not apply to every buyer. If any provision of this warranty is held invalid or unenforceable by a court of competent jurisdiction, such holding will not affect the validity or enforceability of any other provision.

Warranty Registrationhttp://www.honeywellanalytics.com/support/product-registration

Contacting BW Technologies by Honeywell

USA & Canada: 1-888-749-8878

Europe: 00800-333-222-44 www.honeywellanalytics.com

Other countries: 1-403-248-9226

Table of Contents

Title

Page

Introduction ……………………………………………………………………………………………………………………………………………………. 1 Gases Monitored …………………………………………………………………………………………………………………………………………. 2Safety Information – Read First………………………………………………………………………………………………………………………… 2 aCautions …………………………………………………………………………………………………………………………………………………….. 3 Sensor Poisons and Contaminants …………………………………………………………………………………………………………………. 6 Getting Started ……………………………………………………………………………………………………………………………………………….. 8 Parts of the GasAlertMicro 5/PID/IR …………………………………………………………………………………………………………………. 9 Screen Elements …………………………………………………………………………………………………………………………………………… 10 Buttons ………………………………………………………………………………………………………………………………………………………… 11 Activating the Detector ……………………………………………………………………………………………………………………………………………………….. 12Self-Tests………………………………………………………………………………………………………………………………………………….. 12 Battery Test ………………………………………………………………………………………………………………………………………….. 12 Datalogging (Optional)……………………………………………………………………………………………………………………………. 13 Pump Test ……………………………………………………………………………………………………………………………………………. 15 Due-Lock Enabled …………………………………………………………………………………………………………………………………. 17 Force Calibration Enabled ………………………………………………………………………………………………………………………. 17 Bump Daily Enabled ………………………………………………………………………………………………………………………………. 18Self-Test Pass …………………………………………………………………………………………………………………………………………… 19 Self Test Fail……………………………………………………………………………………………………………………………………………… 19 Battery Test ………………………………………………………………………………………………………………………………………………. 20 Datalogger Operation (Optional)…………………………………………………………………………………………………………………… 20 Deactivating the Detector………………………………………………………………………………………………………………………………. 21i

GasAlertMicro 5/PID/IR User Manual

Title

Page

User Options Menu ……………………………………………………………………………………………………………………………………….. 21 Exit User Options Menu ………………………………………………………………………………………………………………………………. 22 Options Menu…………………………………………………………………………………………………………………………………………….. 22 Backlight ………………………………………………………………………………………………………………………………………………. 23 Confidence Beep …………………………………………………………………………………………………………………………………… 23 Due-Lock ……………………………………………………………………………………………………………………………………………… 23 Latched Alarms ……………………………………………………………………………………………………………………………………… 24 Passcode Protect…………………………………………………………………………………………………………………………………… 24 Safe Display………………………………………………………………………………………………………………………………………….. 25 Sensor Configuration ………………………………………………………………………………………………………………………………….. 25 Sensor Enable/Disable …………………………………………………………………………………………………………………………… 26 Span Gas Value…………………………………………………………………………………………………………………………………….. 27 STEL Period …………………………………………………………………………………………………………………………………………. 28 TWA Method …………………………………………………………………………………………………………………………………………. 28 Resolution …………………………………………………………………………………………………………………………………………….. 29 %Vol CO2 (CO2 Sensors Only)………………………………………………………………………………………………………………… 29 %Vol CH4 (LEL Sensors Only)…………………………………………………………………………………………………………………. 30 Correction Factor (CF) ……………………………………………………………………………………………………………………………. 30 Logger Option ……………………………………………………………………………………………………………………………………………. 33 Clock Option ……………………………………………………………………………………………………………………………………………… 33 Language Selection ……………………………………………………………………………………………………………………………………. 34 Tech Mode ………………………………………………………………………………………………………………………………………………… 35 Sensors………………………………………………………………………………………………………………………………………………… 36 Initialize ………………………………………………………………………………………………………………………………………………… 37 Force Calibration …………………………………………………………………………………………………………………………………… 37 Bump Daily …………………………………………………………………………………………………………………………………………… 38 Stealth Mode…………………………………………………………………………………………………………………………………………. 38

ii

GasAlertMicro 5/PID/IR User Manual

Title

Page

Sleep Mode…………………………………………………………………………………………………………………………………………… 39 Alarms ………………………………………………………………………………………………………………………………………………………….. 40Gas Exposures Computed …………………………………………………………………………………………………………………………… 43 Viewing Gas Exposures ………………………………………………………………………………………………………………………………. 43 Clearing Gas Exposures ……………………………………………………………………………………………………………………………… 44 Gas Alarm Setpoints …………………………………………………………………………………………………………………………………… 44 Viewing the Alarm Setpoints ………………………………………………………………………………………………………………………… 44 Resetting Gas Alarm Setpoints …………………………………………………………………………………………………………………….. 45 Stopping a Gas Alarm …………………………………………………………………………………………………………………………………. 46 Sensor Alarm …………………………………………………………………………………………………………………………………………….. 46 Pump Alarm ………………………………………………………………………………………………………………………………………………. 46 Low Battery Alarm………………………………………………………………………………………………………………………………………. 47 Automatic Deactivation Alarm ………………………………………………………………………………………………………………………. 47 Bump Test…………………………………………………………………………………………………………………………………………………….. 47 Performing a Bump Test ……………………………………………………………………………………………………………………………… 47 Calibration and Setting Alarm Setpoints ………………………………………………………………………………………………………… 48 Guidelines …………………………………………………………………………………………………………………………………………………. 48 Diagnostics Testing…………………………………………………………………………………………………………………………………….. 49 Applying Gas to the Sensors ……………………………………………………………………………………………………………………….. 50Single Gas Calibration Cap …………………………………………………………………………………………………………………….. 50 Calibration Procedure …………………………………………………………………………………………………………………………………. 51Start Calibration …………………………………………………………………………………………………………………………………….. 52 Auto Zero and Oxygen (O2) Sensor Calibration …………………………………………………………………………………………. 52 Zero CO2 (GasAlertMicro 5 IR only) …………………………………………………………………………………………………………. 52 Passcode Protect Activated …………………………………………………………………………………………………………………….. 53 Auto Span …………………………………………………………………………………………………………………………………………….. 54 Successful Span ……………………………………………………………………………………………………………………………………. 56

iii

GasAlertMicro 5/PID/IR User Manual

Title

Page

Setting the Calibration Due Date ……………………………………………………………………………………………………………… 57 Alarm Setpoints …………………………………………………………………………………………………………………………………….. 58 Setting the Remaining Alarm Setpoints …………………………………………………………………………………………………….. 60 Finish Calibration …………………………………………………………………………………………………………………………………… 60 Verification ……………………………………………………………………………………………………………………………………………. 61 Unsuccessful Span ………………………………………………………………………………………………………………………………… 61 Pump ……………………………………………………………………………………………………………………………………………………………. 64 ………………………………………………………………………………………………………………………………………………………………… 64 Identifying the Pump …………………………………………………………………………………………………………………………………… 64 Generation 1: 116885-L3 (yellow) and 118933-L3 (black)……………………………………………………………………………. 64 Generation 2: 130916-L3 (yellow) and 130917-L3 (black)……………………………………………………………………………. 65 Installing the Pump Module …………………………………………………………………………………………………………………………… 66 Replacing the Pump Filter (Generation 2 Pump) …………………………………………………………………………………………….. 67 Replacing the Pump Nozzle (Generation 2 Pump) ………………………………………………………………………………………….. 68 Attaching the Auxiliary Filter ………………………………………………………………………………………………………………………… 68 Confined Space Sampling …………………………………………………………………………………………………………………………… 70 Maximum Hose Length when Confined Space Sampling…………………………………………………………………………….. 71 Datalogger ……………………………………………………………………………………………………………………………………………………. 72 MMC/SD Card Compatibility…………………………………………………………………………………………………………………………… 72 Inserting the MMC/SD Card …………………………………………………………………………………………………………………………. 72 MMC/SD Card Troubleshooting ……………………………………………………………………………………………………………………… 73 Restoring Datalog Files……………………………………………………………………………………………………………………………….. 74 Reformatting the MMC/SD Card …………………………………………………………………………………………………………………… 75 Import Datalogs to Fleet Manager II ……………………………………………………………………………………………………………….. 76 Minimum PC Requirements …………………………………………………………………………………………………………………………. 76 Importing from MicroDock II to Fleet Manager II……………………………………………………………………………………………… 76 Import to Fleet Manager II Using a Card Reader …………………………………………………………………………………………….. 76

iv

GasAlertMicro 5/PID/IR User Manual

Title

Page

View Datalog Files in Spreadsheets ……………………………………………………………………………………………………………….. 77 Example of a Datalog Spreadsheet ………………………………………………………………………………………………………….. 78Maintenance …………………………………………………………………………………………………………………………………………………. 81 Battery Cautions ……………………………………………………………………………………………………………………………………….. 81 Charging the Battery …………………………………………………………………………………………………………………………………… 81 Replacing the Alkaline Batteries …………………………………………………………………………………………………………………… 82 Replacing the Lithium Battery Pack ………………………………………………………………………………………………………………. 83 Replacing a Sensor or Sensor Filter ……………………………………………………………………………………………………………… 83 Photoionization Detector (PID) …………………………………………………………………………………………………………………….. 85 Clean or Replace the Lamp …………………………………………………………………………………………………………………….. 85 Replace the Lamp………………………………………………………………………………………………………………………………….. 86 Replace the Electrode Stack …………………………………………………………………………………………………………………… 87 WEEE Battery Directive ………………………………………………………………………………………………………………………………. 87 Removal and Disposal of the Battery Pack………………………………………………………………………………………………… 87 Removal and Disposal of the Coin Cell …………………………………………………………………………………………………….. 87Troubleshooting ……………………………………………………………………………………………………………………………………………. 89 Replacement Parts and Accessories ……………………………………………………………………………………………………………… 95 Specifications ……………………………………………………………………………………………………………………………………………….. 97General Specifications for Datalogger Units …………………………………………………………………………………………………. 100 GasAlertMicro 5/PID/IR Downloadable Datalogger ……………………………………………………………………………………….. 101 PID Correction Factor (CF) Library……………………………………………………………………………………………………………….. 102

v

GasAlertMicro 5/PID/IR User ManualTitle

Page

vi

List of Figures

Figure Title

Page

1.

Parts of the GasAlertMicro 5/PID/IR………………………………………………………………………………………………………. 9

2.

Screen Elements ………………………………………………………………………………………………………………………………. 10

3.

Applying Gas to the Sensors ………………………………………………………………………………………………………………. 50

4.

Single Gas Calibration Cap ………………………………………………………………………………………………………………… 51

5.

Removing the Single Gas Calibration Cap ……………………………………………………………………………………………. 51

6.

Installing the Pump Module ………………………………………………………………………………………………………………… 66

7.

Replacing the Pump Filter (Generation 2 Pump)……………………………………………………………………………………. 67

8.

Replacing the Pump Nozzle ……………………………………………………………………………………………………………….. 68

9.

Attaching the Auxiliary Filter ………………………………………………………………………………………………………………. 68

10.

Attaching the Filter Cord……………………………………………………………………………………………………………………. 69

11.

Attaching the Sample Probe……………………………………………………………………………………………………………….. 70

12.

Inserting/Removing the MMC/SD Card ………………………………………………………………………………………………… 73

13.

Replacing the Alkaline Batteries………………………………………………………………………………………………………….. 82

14.

Replacing the Lithium Battery Pack …………………………………………………………………………………………………….. 83

15.

Replacing a Sensor or Sensor Filter ……………………………………………………………………………………………………. 84

16.

Parts of the PID ………………………………………………………………………………………………………………………………… 86

vii

GasAlertMicro 5/PID/IR User Manualviii

List of Tables

Table Title

Page

1.

Gases Monitored ………………………………………………………………………………………………………………………………… 2

2.

Sensor Poisons and Contaminants ……………………………………………………………………………………………………….. 6

3.

International Symbols ………………………………………………………………………………………………………………………….. 7

4.

Parts of the GasAlertMicro 5/PID/IR………………………………………………………………………………………………………. 9

5.

Screen Elements ………………………………………………………………………………………………………………………………. 10

6.

Button ……………………………………………………………………………………………………………………………………………… 11

7.

Alarms …………………………………………………………………………………………………………………………………………….. 40

8.

Computed Gas Exposures …………………………………………………………………………………………………………………. 43

9.

Gas Alarm Setpoints …………………………………………………………………………………………………………………………. 44

10.

OSHA Sample Factory Alarm Setpoints……………………………………………………………………………………………….. 45

11.

Applying Gas to the Sensors ………………………………………………………………………………………………………………. 50

12.

Single Gas Calibration Cap ………………………………………………………………………………………………………………… 51

13.

Time Required to Span ……………………………………………………………………………………………………………………… 55

14.

Installing the Pump Module ………………………………………………………………………………………………………………… 66

15.

Generation 1 Pump Maximum Hose Length …………………………………………………………………………………………. 71

16.

Generation 2 Pump Maximum Hose Length …………………………………………………………………………………………. 71

17.

Datalog Spreadsheet Example……………………………………………………………………………………………………………. 78

18.

Datalog Status Codes ……………………………………………………………………………………………………………………….. 79

19.

Datalog Gas and Correction Factor Sensor Codes ………………………………………………………………………………… 80

20.

Replacing the Alkaline Batteries………………………………………………………………………………………………………….. 82

21.

Replacing a Sensor or Sensor Filter …………………………………………………………………………………………………… 85

22.

Parts of the PID sensor ……………………………………………………………………………………………………………………… 85

23.

Troubleshooting………………………………………………………………………………………………………………………………… 89

ix

GasAlertMicro 5/PID/IR User Manual

24.

Replacement Parts and Accessories …………………………………………………………………………………………………… 95

25.

PID Corrections Factor (CF) Library…………………………………………………………………………………………………… 102

x

ISO 9001

GasAlertMicro 5/PID/IR IntroductionGasAlertMicro 5/PID/IRIntroductiona Warning To ensure personal safety, read the Safety Information Read First and Cautions before using the detector. The GasAlertMicro 5, GasAlertMicro 5 PID, and GasAlertMicro 5 IR gas detectors (“the detector”) warn of hazardous gas at levels above user-defined alarm setpoints. The detector is a personal safety device. It is your responsibility to respond properly to the alarm.Note Unless reference is made to a specific detector model, the GasAlertMicro 5, GasAlertMicro 5 PID, and GasAlertMicro 5 IR detectors are referred to as GasAlertMicro 5/PID/IR. Refer to Gases Monitored prior to operating the detector.1

GasAlertMicro 5/PID/IR User Manual

Gases MonitoredThe following table lists the gases that are monitored by the detector. Table 1. Gases Monitored

Gas Detected Oxygen (O2)Combustible gases (LEL)Carbon monoxide (CO) Hydrogen sulfide (H2S) Phosphine (PH3) Sulfur oxide (SO2) Chlorine (Cl2) Ammonia (NH3) Nitrogen dioxide (NO2) Hydrogen cyanide (HCN) Chlorine dioxide (ClO2) Ozone (O3) Volatile organic compounds (VOC) Carbon dioxide (CO2) 2

Unit of Measure% volume a) percent of lower explosivelimit (%LEL) b) percent by volumemethane 0-5.0% v/v parts per million (ppm) parts per million (ppm) parts per million (ppm) parts per million (ppm) parts per million (ppm) parts per million (ppm) parts per million (ppm) parts per million (ppm) parts per million (ppm) parts per million (ppm)parts per million (ppm)parts per million or %vol CO2

CAUTION: FOR SAFETY REASONS, THIS EQUIPMENT MUST BE OPERATED AND SERVICED BY QUALIFIED PERSONNEL ONLY. READ AND UNDERSTAND THIS USER MANUAL COMPLETELY BEFORE OPERATING AND SERVICING.Safety Information – Read FirstUse the detector only as specified in this user manual, otherwise the protection provided by the detector may be impaired.International symbols used on the detector and in this user manual are defined in Table 3.Read the Cautions on the following pages before using the detector.ec WarningThis instrument contains batteries. Do not mix with the solid waste stream. Spent batteries must be disposed of by a qualified recycler or hazardous materials handler. Dispose of lithium cells immediately. Do not disassemble and do not dispose of in fire. Do not mix with the solid waste stream. Spent batteries must be disposed of by a qualified recycler or hazardous materials handler.

aCautions· Warning: Substitution of components may impair Intrinsic Safety.· Caution: For safety reasons, this equipment must be operated and serviced by qualified personnel only. Read and understand this user manual completely before operating or servicing.· Do not use the detector if it is damaged. Inspect the detector before using. Look for cracks and/or missing parts.· If the detector is damaged or parts are missing, contact BW Technologies by Honeywell immediately.· Use only sensor(s) that are specifically designed for the GasAlertMicro 5/PID/IR detectors. Refer to Replacement Parts and Accessories.· Calibrate the detector before first-time use and then on a regular schedule, depending on use and sensor exposure to poisons and contaminants. Sensors must be calibrated regularly and at least once every 180 days (6 months).· BW recommends to bump test the sensors, before each day’s use, to confirm their ability to respond to gas by exposing the detector to a gas concentration that exceeds the alarm setpoints. Manually verify that the audible and visual alarms are activated. Calibrate if the readings are not within the specified limits.· BW recommends the combustible sensor be checked with a known concentration of calibration gas after any known exposure to catalyst contaminants/poisons (sulfur compounds, silicon vapors, halogenated compounds, etc).

GasAlertMicro 5/PID/IR aCautions· The combustible sensor is factory calibrated to 50% LEL methane. If monitoring a different combustible gas in the % LEL range, calibrate the sensor using the appropriate gas.· Warning: High off-scale LEL readings may indicate an explosive concentration.· Only the combustible gas detection portion of this instrument has been assessed for performance by CSA International.· Protect the combustible sensor from exposure to lead compounds, silicones, and chlorinated hydrocarbons. Although certain organic vapors (such as leaded gasoline and halogenated hydrocarbons) may temporarily inhibit sensor performance, in most cases, the sensor will recover after calibration. aCautions· Before using common products around sensors, refer to Sensor Poisons and Contaminants.· High concentrations of certain toxic gases, for example H2S, may have an adverse effect on the LEL sensor. This effect, known as inhibition, is usually temporary but in extreme circumstances can impair the sensitivity of the LEL sensor. After any gas exposure that causes an alarm in the toxic gas sensors, the LEL sensor should be verified with a bump test, and recalibrated if necessary.· Any rapid up-scaling reading followed by a declining or erratic reading may indicate a gas concentration beyond upper scale limit, which may be hazardous.3

GasAlertMicro 5/PID/IR User Manual

· Calibrate only in a safe area that is free of hazardous gas, in an atmosphere of 20.9% oxygen.· Use only BW approved batteries for the GasAlertMicro 5/ PID/IR detectors. Refer to Replacement Parts and Accessories.· Charge the detector before first-time use. BW recommends the detector be charged after every workday.· Charge the battery pack immediately when a low battery alarm occurs.· Read and adhere to the battery cautions provided in Battery Cautions.· Charge the GasAlertMicro 5/PID/IR batteries using the recommended charging adapter only. Do not use any other charging adapter. Failure to adhere to this caution can lead to fire and/or explosion.· Read and adhere to all instructions in the charger user manual. Failure to do so can result in fire, electrical shock, personal injury, and/or property damage.· Extended exposure of the GasAlertMicro 5/PID/IR detectors to certain concentrations of combustible gases and air may stress a detector element, which can seriously affect its performance. If an alarm occurs due to high concentration of combustible gases, recalibration should be performed, or if needed, the sensor replaced.aCautions · Protect the PID sensor from exposure to silicone vapors.

· When calibrating O3 and ClO2 sensors that are located in the Toxic 2 sensor position, a single gas calibration cap must be used to ensure accurate calibration. For more information, refer to Single Gas Calibration Cap.· Replace the CO2 sensor only in a safe and non-hazardous area that is free of hazardous gas.· The optional pump (M5-PUMP) is certified for use with the GasAlertMicro 5/PID/IR detectors only.· Do not immerse the detector in liquids. · Do not test the combustible sensor’s response with abutane cigarette lighter; doing so will damage the sensor. · Do not expose the detector to electrical shock or severecontinuous mechanical shock. · Do not attempt to disassemble, adjust, or service thedetector unless instructions for that procedure are provided in the user manual and/or that part is listed as a replacement part. Use only BW Technologies by Honeywell Replacement Parts and Accessories. · The detector warranty is void if customer, personnel, or third parties damage the detector during repair attempts. Repair attempts made by non-BW Technologies by Honeywell repair/service personnel voids this warranty. Lithium Battery Packs · Warning: The lithium battery (M5-BAT08) may present a risk of fire or chemical burn hazard if misused. Do not disassemble, heat above 100°C (212°F), or incinerate.

4

· Do not use any other lithium batteries with the GasAlertMicro 5/PID/IR detectors. Use of any other cell can cause fire and/or explosion. To order and replace the M5BAT08 battery, refer to Replacement Parts and Accessories.· Warning: Lithium polymer cells exposed to heat at 130°C (266°F) for 10 minutes can cause fire and/or explosion.· Dispose of used lithium cells immediately. Do not disassemble and do not dispose of in fire. Do not mix with the solid waste stream. Spent batteries must be disposed of by a qualified recycler or hazardous materials handler.· Keep lithium cells away from children. · Products may contain materials that are regulated fortransportation under domestic and international dangerous goods regulations. Return product in compliance with appropriate dangerous goods regulations. Contact freight carrier for further instructions.

GasAlertMicro 5/PID/IR aCautions5

GasAlertMicro 5/PID/IR User ManualSensor Poisons and ContaminantsSeveral cleaners, solvents, and lubricants can contaminate and cause permanent damage to sensors. Before using cleaners, solvents, and lubricants in close proximity to the detector sensors, read and adhere to the following caution and table.a Caution Use only the following BW Technologies by Honeywell recommended products and procedures: · Use water based cleaners. · Use non-alcohol based cleaners. · Clean the exterior with a soft, damp cloth. · Do not use soaps, polishes, or solvents.6

The following table lists common products to avoid using around sensors.Table 2. Sensor Poisons and Contaminants

Cleaners and LubricantsBrake cleanersLubricantsRust inhibitorsWindow and glass cleaners DishsoapsCitrus based cleaners Alcohol based cleaners Hand sanitizers Anionic detergents Methanol (fuels and antifreezes)

SiliconesSilicone cleaners and protectantsSilicone based adhesives, sealants, and gelsHand/body and medicinal creams that contain siliconeTissues containing siliconeMold releasing agentsPolishes

Aerosols Bug repellents and sprays LubricantsRust inhibitorsWindow and glass cleaners

Table 3. International Symbols

Symboln

Description Approved to both U.S. and Canadian Standards by CSA International

GasAlertMicro 5/PID/IR Sensor Poisons and Contaminants

g

European Explosives Protection

X

Conforms to European Union Directives

ATEX IECEx EAC Ex

Conforms to European ATEX DirectivesInternational Electrotechnical Commission Scheme for Certification to Standards for Electrical Equipment for Explosive AtmospheresConforms to Russian Custom Union Certification and Declaration

Conforms to Korea Testing Laboratory (KTL) Certification

Conforms to Brazilian InMetro Certification

Australian Regulatory Compliance Mark 7

GasAlertMicro 5/PID/IR User ManualGetting StartedThe list below provides the standard items included with the detector. If the detector is damaged or parts are missing, contact the place of purchase immediately.· Batteries: Three replaceable alkaline cells with battery pack, or one rechargeable lithium battery pack· Sensors: O2, combustible (LEL), toxic, H2S/CO (TwinTox sensor), PID, or CO2· Calibration cap and hose · Single gas calibration cap · Screwdriver · Quick reference guide · Quick reference card · Technical Reference Guide and training on CD-ROM · Fleet Manager II (if applicable)To order parts, refer to Replacement Parts and Accessories.

The detector is shipped with the sensors, and battery packs installed. To replace sensors, the pump, or the battery pack, refer to Replacement Parts and Accessories.To become oriented with the features and functions of the detector, refer to the following figures and tables:· Figure 1. and Table 4. describe the detector’s components.· Figure 2. and Table 5. describe the detector’s screen elements.· Table 6. describes the detector’s button.

8

Parts of the GasAlertMicro 5/PID/IRFigure 1. Parts of the GasAlertMicro 5/PID/IR

GasAlertMicro 5/PID/IR Parts of the GasAlertMicro 5/PID/IR

Table 4. Parts of the GasAlertMicro 5/PID/IR

Item 1 2 3 456 7 8 9 10 11

DescriptionLiquid crystal display (LCD) Buttons Audible alarms Toxic 2 sensor Toxic 1/PID sensor (GasAlertMicro 5 PID) or Toxic 1/IR (CO2) sensor (GasAlertMicro 5 IR) Visual alarm indicators (LEDs) LEL sensor Oxygen sensor Buttons Battery pack Alligator clip

9

GasAlertMicro 5/PID/IR User ManualScreen ElementsFigure 2. Screen Elements NoteIf enabled, the backlight automatically activates for 8 seconds when there is an alarm condition and whenever there is insufficient light to view the LCD. Any button reactivates the backlight in low light conditions. 10

Table 5. Screen Elements

Item 1 2 3 4 5 6 7 8 9 10 111213

DescriptionAlarm condition Automatically span sensor Gas cylinder Gas type Battery life indicator Passcode lock Data transmission Clock Stealth mode Pump indicator (optional) MMC indicator (optional) Alarm condition (low, high, TWA, STEL, or multi alarm) or view TWA, STEL and peak (MAX) gas exposures Automatically zero sensor

ButtonsButtonAGH C

GasAlertMicro 5/PID/IR ButtonsTable 6. ButtonDescription · To activate the detector press A. · To deactivate the detector, press and hold A until the countdown is complete. · To increment the displayed value or scroll up, press G. · To enter the user options menu, press and hold G and H simultaneously until the countdown is complete. · To clear the TWA, STEL, and peak (MAX) gas exposure readings, press and hold C and G simultaneouslyuntil the countdown is complete. · To view the date and time, alarm setpoints (TWA, STEL, low, high) of all sensors, and the LEL/PID correction factor(if applicable) press G. · To decrement the displayed value or scroll down, press H. · To initiate calibration and to define alarm setpoints, press and hold C and H simultaneously until the countdownis complete. · To view the TWA, STEL, and peak (MAX) gas exposure readings, press C. · To acknowledge latched alarms, press C. Refer to Latched Alarms on page 24.

11

GasAlertMicro 5/PID/IR User ManualActivating the DetectorIf using the pump module, attach it and the pump accessories prior to activating the detector. For illustrations and procedures, refer to the following:·. · Confined Space Sampling · Replacing the Pump Filter (Generation 2 Pump) · Replacing the Pump Nozzle (Generation 2 Pump) · Attaching the Auxiliary FilterNote Only activate the detector in a safe atmosphere that is free of hazardous gas in an atmosphere of 20.9% oxygen.To activate the detector, press A.Self-TestsWhen the detector is activated, it performs several self-tests. Confirm the following tests occur.Note If an error message displays during the self-test, refer to Troubleshooting.

Battery TestThe detector performs a battery test during startup. If the battery has insufficient power to operate, the following screen displays before deactivating.Replace the batteries and reactivate the detector. 1. All of the LCD elements display simultaneously as the detector beeps, flashes, vibrates, and briefly activates the backlight.

12

2. The version and serial number of the detector displays. 3. The date and time displays.

GasAlertMicro 5/PID/IR Activating the DetectorNoteIf there is a problem with the MMC/SD card, Datalogger disabled displays. The detector then automatically continues with the self-test.If the card requires formatting, the following screen displays as the card is automatically formatted.

Datalogging (Optional)4. If the detector is a datalogging unit, it determines if · a MultiMediaCard (MMC) or Secure Digital (SD) card is inserted, · the detector can communicate with the card, · the detector supports the size of the card, and · the card requires formatting.

5. The detector then runs a self-test to verify the sensors and power supply are operating correctly.13

GasAlertMicro 5/PID/IR User ManualSelf-test Successful: If successful, the following screen displays.

7. The TWA, STEL, low, and high alarm setpoints then display in the following order (left to right).

TWA

STEL

Self-test Unsuccessful: If a sensor fails the self-test, a warning displays indicating which sensor(s) has failed.

Low

High

6. If correction factors are set in the user options, the LEL or PID (custom) correction factors display.

NoteThe alarm setpoints may vary by region. Refer to Resetting Gas Alarm Setpoints.

14

Pump Test a WarningIf the pump module is installed on the detector, the following three things must occur during start-up. If any one of the conditions below does not occur, discontinue use of the detector and contact BW Technologies by Honeywell immediately. · The detector prompts for a pump test during start-up · The pump module passes the pump test at start-up whenthe pump inlet or sample chain inlet is blocked · The J icon displays on the LCD 8. If the pump module is attached to the detector, the followingscreens display.When the following screen displays, block the pump inlet.

GasAlertMicro 5/PID/IR Activating the Detector15

GasAlertMicro 5/PID/IR User ManualIf the pump inlet is not blocked within 10 seconds or the pump test fails, the following screens display.

9. Unless disabled in user options, the oxygen (O2) sensor is calibrated automatically.

If C is not pressed or the pump is not removed within 25 seconds, the detector performs the pump test again.NoteThe pump may require a new filter. Refer to Replacing the Pump Filter (Generation 2 Pump).If the pump alarm persists, remove the battery pack to deactivate the detector and then refer to the “Pump Operation” section of Troubleshooting.If the pump test is successful, the following screen displays and the self-test continues.

If the calibration is successful, the detector beeps twice.Note If the automatic O2 calibration feature has been disabled, Automatic O2 span disabled displays.10. The number of days remaining before calibration is due displays for all sensors.

16

If any sensor is over due for calibration, the LCD displays the name of the sensor and the number of days past due.

GasAlertMicro 5/PID/IR Activating the DetectorIf no passcode is entered, or it is entered incorrectly, the following screen displays.

Due-Lock EnabledThe Due-lock option ensures that a passcode must be entered when calibration is past due, otherwise the detector automatically deactivates.11. If Due-Lock is enabled in the user options, the following screen displays.

To enable/disable this option, refer to Due-Lock in the user options menu. Also refer to Passcode Protect.Force Calibration Enabled12. If Force cal (force calibration) is enabled in tech mode, calibration is mandatory before the detector enters normal operation.Refer to Force Calibration in Tech Mode to enable/disable, and refer to Calibration Procedure to calibrate.

Enter the correct passcode and press C to confirm.Note If any sensor is overdue, Y displays continually until calibration is performed.

If C is not pressed to start calibration, the following screen displays and the detector deactivates.17

GasAlertMicro 5/PID/IR User Manual

Successful Bump Test: If the bump test passes, the following screens display.

Bump Daily Enabled a CautionBW recommends that a bump test to all sensors be performed every 24 hours prior to the beginning of the work shift.13. If Bmp Daily (bump daily) is enabled in tech mode, the following screens display.

The detector waits for the sensor(s) to clear (30 seconds) and then enters normal operation.Unsuccessful Bump Test: If the bump test is unsuccessful or the bump test is not performed, the following screen displays and the detector deactivates.

If a bump test of the LEL and O2 sensor is not performed, the detector will deactivate.Apply the LEL gas and then apply the O2 (a higher or lower percentage than the default 20.9%, such as 18% O2).18

If additional sensors require a bump test but are not mandatory, the following screens display.

GasAlertMicro 5/PID/IR Activating the DetectorSelf-Test PassIf the detector passes the self-test, it enters normal operation and displays the ambient gas readings.

Press C Yes to accept and proceed to normal operation.OrIf A No is pressed, or no buttons are pressed, the sensor(s) that is past due displays as FAIL when the detector enters normal operation. In the following screen example, only the SO2, CO, and O2 sensors are operational.

The detector begins recording the peak (MAX) gas exposure and calculating the short-term exposure level (STEL) and time-weighted average (TWA) exposures.Self Test FailIf a sensor fails, FAIL flashes above that sensor on the normal operation screen. For possible problems and solutions, refer to Troubleshooting.

The self-test is now complete. 19

GasAlertMicro 5/PID/IR User Manual Battery TestThe batteries are tested when the detector is activated and continuously thereafter. The battery power icon displays continually during normal operation. If battery power is low, flashes.Datalogger Operation (Optional) a CautionDo not remove the battery pack while the detector is activated. Doing so will prevent the datalogger from logging correctly. Datalogger operation is automatic and requires no settings. During normal operation the card is tested every 20 seconds.Note The MMC/SD card icon (S) displays continuously on the LCD when the card is inserted. The card is not required for operation; however, if the card is not inserted the detector will not record data.20

GasAlertMicro 5/PID/IR Deactivating the Detector

Deactivating the DetectorTo deactivate the detector, press and hold A while it beeps and flashes to the corresponding countdown.At the end of the countdown, the detector emits an extended beep and flash then displays 0 before deactivating.Note If A is not held down for the complete countdown, the detector remains activated.

User Options MenuIf the detector is passcode protected, a passcode must be entered to access the user options menu. For more information, refer to Passcode Protect.The available user options are as follows:1. Exit2. Options: Backlight, confidence beep, due-lock, latch, passcode, and safe mode.3. Sensors: Sensor enable/disable, span gas, STEL period, TWA method, resolution, % vol CH4, correction factor, automatic O2 calibration, and % vol CO2 (CO2 sensor only)4. Logger5. Clock6. Language: English, French, German, Spanish, and Portuguese.7. Tech mode: Sensors, initialize, forced calibration, daily bump test, stealth, and zero level (CO2 sensor only).NoteTech mode is not visible in the user options menu. To access this option, refer to Tech Mode.

21

GasAlertMicro 5/PID/IR User Manual1. To enter the user options menu, press and hold G and H simultaneously as the detector beeps and flashes to the corresponding countdown.

Note If no buttons are pressed within 20 seconds, the detector returns to normal operation.Exit User Options MenuTo exit the user options menu and return to normal operation, scroll to Exit and press C. The following screen displays.

G and H must be held down for the entire countdown to access the user options menu. When the countdown is complete, the revision/serial number screen displays followed by the options menu.2. To scroll through the options, press H or G. When the cursor displays beside the desired option, press C.3. To return to the previous menu, scroll to Back and press C or press A.

The user options menu can also be exited by repeatedly pressing A until the detector returns to normal operation.

Options Menu

Each feature within the Options menu is enabled/disabled by pressing C to toggle between the checkboxes.

Enabled

Disabled

22

BacklightThe Backlght (backlight) option enables the LCD backlight to activate automatically in low-light conditions. If disabled, the backlight activates only when the detector is in alarm mode.NoteIn stealth mode, the backlight does not activate.The detector is shipped with the Backlght option enabled.Confidence BeepThe Confibeep (confidence beep) option provides continuous confirmation that the detector is operating properly. When confidence beep is enabled, the audible alarm beeps once every 10 seconds.NoteConfidence beep is automatically disabled during a low battery alarm.The detector is shipped with the Confibeep option disabled.

GasAlertMicro 5/PID/IR User Options MenuDue-LockIf the Due-lock (calibration user lockout) option is enabled and a sensor is overdue for calibration upon startup, the passcode must be entered to access normal operation.Note If the correct passcode is not entered, the detector deactivates. The detector is shipped with the Due-lock option disabled.23

GasAlertMicro 5/PID/IR User ManualLatched AlarmsIf enabled, the Latch (latched alarms) option causes the low and high gas alarms (audible, visual, and vibrator) to persist until they are acknowledgedby pressing C and the gas concentration is below the low alarm setpoint. The peak concentrations display continually until the alarm condition no longer exists. Local regulations may require Latch option be enabled.The detector is shipped with the Latch option disabled.Passcode ProtectThe Passcode option prevents unauthorized access to the user options menu, the calibration function, and to adjusting the alarm setpoints.Note The passcode is provided separately in the supplementary booklet.24

If passcode protect is enabled and the Enter passcode: 1000 screen displays, press G or H to scroll to the correct passcode and then press C to confirm.The detector is shipped with the Passcode option disabled.If an incorrect passcode is entered or C is not pressed within 5 seconds to confirm the correct passcode, Passcode incorrect displays. The alarm beeps three times and the detector either· resumes normal operation (for user options), or· deactivates (for calibration if Due Lock is enabled).

Safe DisplayWhen enabled, the Safe option confirms that conditions are normal and there are no gas hazards present. When all gas levels are normal or below the alarm setpoints, Safe displays continually on the LCD.Safe does not display if any active sensor has failed, if any active sensor has an overdue calibration, and/or if any active sensor has failed a bump test.The detector is shipped with the Safe option disabled.

GasAlertMicro 5/PID/IR User Options MenuSensor ConfigurationThe Sensor option provides access to additional options and functions for each sensor. Depending upon the sensor that is selected, some or all of the following options are available for configuration:· Enabling/disabling a sensor · Setting the span gas value · Adjusting the STEL period (not applicable to LEL and O2 sensors) · Selecting the TWA method (not applicable to LEL and O2 sensors) · Resolution setting (not applicable to CO, LEL, O2, and CO2sensors) · % vol CO2 (CO2 sensor only) · % vol CH4 (LEL sensor only) · Selecting the correction factor (LEL and PID sensors only) · Automatic calibration (O2 sensor only)

25

GasAlertMicro 5/PID/IR User Manual1. From the option menu screen, scroll to Sensors and press C to access the following screen.2. Press G or H to scroll to the desired sensor. Press C to confirm and to access the menu options specific to the selected sensor. For all sensor options, if a value is changed but not confirmed within 5 seconds, the detector emits an audible alarm and displays the following error message.

Sensor Enable/Disable a WarningDisabling a sensor should only be performed with extreme caution. The disabled sensor cannot measure and alarm against the applicable gas. If a sensor fails, disabling the sensor deactivates the fail alarm. The sensor should be replaced and enabled as soon as possible. The detector will function normally with the remaining enabled sensors.NoteDetectors that are configured for a 1, 2, 3, or 4 gases may contain a dummy sensor in one of the four sensor locations.After selecting the desired sensor, the following screen displays.

The detector retains the previous setting and returns to the user options menu.

Press C to toggle between enable/disable (sensor can be enabled at any time).Enabled Disabled

26

If disabled, the readings and the gas type for the sensor do not display when in normal operation.If a sensor is enabled but it is not installed in the detector, FAIL flashes above the gas type of the missing sensor.

Disabled

Enabled/not installed

If all the sensors are disabled, the following screens display.

Enable one or more sensors to exit and access normal operation.

GasAlertMicro 5/PID/IR User Options Menu

Span Gas Value

The Span gas option increases or decreases the gas concentration for calibration (must match the gas concentration listed on the gas cylinder).

After selecting the sensor, press H to scroll to Span gas and press C within 20 seconds to confirm.

Depending upon the sensor selected, a

screen similar to A or B displays. Press

A

G or H to scroll to the desired value and

press C within 5 seconds to confirm.

If C is not pressed within 5 seconds to confirm the new value, the detector retains the previous value and returns to the user options menu.

Note

BW recommends that span concentration

B

values be set between specific ranges.

Refer to the Resetting Gas Alarm

Setpoints.

27

GasAlertMicro 5/PID/IR User ManualSTEL PeriodThe short-term exposure limit (STEL period) option provides protection for workers from over exposure to high concentrations of gas, and is based on 5-15 minute intervals.The STEL period option is available for every toxic sensor. After selecting the desired sensor, press H to scroll to STEL period and press C within 20 seconds to confirm. The STEL period can be set from 5 to 15 minutes. Press G or H to scroll to the required value, and then press C within 5 seconds to confirm. If C is not pressed within 5 seconds to confirm the new value, the detector retains the previous value and returns to the user options menu. The detector is shipped with the STEL period set to 15 minutes.28

TWA MethodThe TWA method (time-weighted average) option is a safety measure used to calculate accumulated averages of gases to notify the user when the maximum average is accumulated. The options are the OSHA or ACGIH calculating method.OSHA Method: 8 hour moving average–oldest value (first hour) is replaced by the newest value (ninth hour). ACGIH Method: Infinite accumulated average to 8 hours–total accumulation, whether it is 2 hours or 8 hours.After selecting the desired sensor, press H to scroll to TWA method. Press C within 20 seconds to confirm.A check displays in the checkbox of the currently selected method. To select the other method, press H to move the check to the other method. Press C to confirm the selection.The detector is shipped with the OSHA method enabled.NoteIf the TWA method has been changed, the TWA, STEL, and MAX values must be reset to ensure the TWA is calculated correctly. Refer to Clearing Gas Exposures.

ResolutionThe Resolution option displays the gas measurement using Regular or Extra resolution.Regular: Displays gas measurement in 1 ppm increments.Extra: Displays the gas measurement in 0.1 ppm increments.NoteRegular resolution for O3 and ClO2 sensors is 0.1 ppm, while extra resolution is 0.01 ppm. The Resolution option is not available for CO, O2, LEL, PID, and CO2 sensors.After selecting the desired sensor, press H to scroll to Resolution. Press C within 20 seconds to confirm.A check displays in the checkbox of the currently selected resolution. To select the other resolution, press H to move the check to other resolution. Press C to confirm the selection.The detector is shipped with Regular resolution enabled for applicable sensors.

GasAlertMicro 5/PID/IR User Options Menu%Vol CO2 (CO2 Sensors Only)If the %vol CO2 is enabled, the detector displays the carbon dioxide (CO2) readings as %vol (0.0). From the Sensors option menu, select CO2.Press H to scroll to %vol CO2. Press C to toggle between enable and disable. Confirmation is not required. If no buttons are pushed, after 20 seconds the detector returns to the sensor selection screen. The change is saved automatically. The detector is shipped with %vol CO2 disabled.Note Correction factors are not applicable to the CO2 sensor.

29

GasAlertMicro 5/PID/IR User Manual%Vol CH4 (LEL Sensors Only)If %vol CH4 is enabled, any currently enabled correction factor is ignored and the detector operates assuming a methane (CH4) calibration.From the Sensors menu, select LEL and press C to confirm.Next, press H to scroll to %vol CH4 and press C within 20 seconds to confirm.Press C to toggle between enable and disable.Confirmation is not required. If no buttons are pushed within 20 seconds, the detector returns to the sensor selection screen. The change is saved automatically.NoteIf changing the measurement unit from % LEL to % Vol. or from % Vol. to % LEL, a calibration must be completed and the alarm setpoints changed. For calibration information and for alarm setpoint information refer to Calibration and Setting Alarm Setpoints on page 48.The detector is shipped with %vol CH4 disabled.30

Correction Factor (CF)Depending upon the selected sensor, refer to the following sections LEL Sensor or PID Sensor for more information.Note Corrections factors are not applicable to CO2 sensors.

GasAlertMicro 5/PID/IR User Options Menu

LEL SensorThis option is used to enter compensation factors for hydrocarbons other than methane. The factor can only be applied if the LEL sensor has been calibrated with methane.After selecting the LEL sensor, press H to scroll to Correction. Press C within 20 seconds to confirm and access the LEL correction library.Scroll to the required gas type and press C. A check displays in the corresponding checkbox. The detector automatically applies the correction factor.To disable the Correction option, press H to scroll to None or to Methane. A check displays in the corresponding checkbox. If required, select a different gas type correction factor.Custom: To enter a correction factor that is not listed in the library, press H to scroll to Custom and press C within 5 seconds to confirm.The Custom LEL correction screen displays. Press G or H to select the required value, and press C within 5 seconds to confirm.

PID SensorThis option is used to enter compensation factors for selected gas types. The factor can only be applied if the PID sensor has been calibrated with isobutylene.After selecting the PID sensor, press H to scroll to Correction. Press C within 20 seconds to confirm and access the PID correction library.Scroll to the required gas type and press C. A check displays in the corresponding checkbox. The detector automatically applies the correction factor.To disable the Correction option, press H to scroll to None or to Isobutyl. A check displays. If required, select a different gas type correction factor.Custom: To enter a correction factor for a custom PID sensor, scroll to Custom and press C. Press G or H to scroll to the required value, and press C within 5 seconds to confirm. Refer to the PID Correction Factor PID Correction Factor (CF) Library Library for gas types and corresponding correction factor values.

31

GasAlertMicro 5/PID/IR User ManualNoteThe PID sensor is not sensitive enough to detect VOCs Benzene, Butadiene, and Vynil Chloride before they exceed their toxic threshold limit value. The MicroDock II is still able to set the Micro 5PID/IR to one of these VOCs. In this case the Micro5/PID/IR will warn that the selection is not supported and ask if you wish to continue. If the warning is not acknowledged the Micro 5/PID/IR will fail the PID Sensor until you reboot and accept, or choose a different VOC.

Automatic Oxygen (O2) CalibrationWhen the Autocal option is enabled, it forces the detector to automatically calibrate the oxygen sensor during startup.If the Autocal option is enabled, ensure the detector is activated in a safe area that is free of hazardous gas in an atmosphere of 20.9% oxygen.From the Sensor menu, press H to scroll to O2 and press C within 20 seconds to confirm.Press H to scroll to Autocal. Press C to toggle between enable/ disable.The detector is shipped with the Autocal option enabled.

32

Logger OptionThe Logger option is used to define how often the detector records a datalog sample (once every 1 to 127 seconds).From the user options menu, press H to scroll to Logger. Press C within 20 seconds to confirm.

GasAlertMicro 5/PID/IR User Options MenuIf C is not pressed within 5 seconds, the following screen displays.

Press G or H to change the current logger rate. When the desired value displays, press C within 5 seconds to confirm the new value.

The detector is shipped with the datalog sample time set to 5 seconds.Clock OptionThe Clock option is used to change the date and time. From the user options menu, press H to scroll to Clock. Press C within 20 seconds to confirm.

33

GasAlertMicro 5/PID/IR User Manual

The screen displays showing the month highlighted indicating it is selected to set.Press G or H to scroll to the desired month and press C within 20 seconds to confirm. To bypass and retain the current setting, press C. Continue setting/bypassing the remaining options.

Language SelectionThe detector is shipped with English selected as the default language. The available languages are as follows:· Français (French) · Deutsch (German) · Español (Spanish) · Prtuguês (Portuguese)Press H to scroll to Language and press C within 20 seconds to confirm.

The date/time options are set in the following order: · month · day · year · hour · minutesWhen the settings have been set or bypassed, the detector beeps twice and returns to the user options menu.NoteThe detector is shipped with the date and time set to Mountain Standard Time (MST).

Press G or H to scroll to the desired language and press C. A check displays in the checkbox of the selected language.

34

Wait for 20 seconds until the detector returns to the user options menu, or press G to scroll to Back (English), Retour (French), Zurück (German), Regreso (Spanish), or Retornar (Portuguese). All screens now display in the selected language.Tech Modea Warning Tech mode should only be accessed by authorized personnel. Tech mode can only be accessed from the Language option. Press H to scroll to Language. Do not press C until instructed.

GasAlertMicro 5/PID/IR User Options MenuIn the following order, press and continue to hold each button until Tech mode displays below the Language option.1. Press and hold H with the right index finger. 2. Press and hold G with the right middle finger. 3. Press and hold C with the left thumb.

4. Press C to enter Tech mode. The options are as follows:· Sensors · Initialize · Force Calibration (Force calibration) · Bump Daily (Bump test daily) · Stealth Mode · Sleep Mode

35

GasAlertMicro 5/PID/IR User ManualSensors a CautionTo reconfigure the sensor type, physically change the sensor prior to entering Tech mode.When a toxic sensor is physically removed and replaced by another toxic sensor, the detector must be reconfigured to recognize the change.Note If a sensor is replaced, the detector will classify the sensor as overdue for calibration. Calibrate the new sensor immediately.1. Press H to scroll to Sensors. Press C within 20 seconds to confirm and access the toxic sensor menu.2. Press G or H to scroll to Toxic 1 or Toxic 2 and press C within 20 seconds to confirm.

A corresponding list of toxic sensors displays. A checkbox displays beside the current toxic sensor.NoteToxic 1: List includes the PID and CO2 sensors.Toxic 2: List includes the H2S/CO COSH sensor.3. Press G or H to scroll to the new sensor and press C to confirm. A checkbox displays beside the new sensor. To reconfigure, exit the user options menu.The following screen displays. The detector deactivates and immediately reactivates. It performs the reconfiguration during the startup.The new sensor must be calibrated. Until the new sensor is calibrated, the due date displays as OL (over limit) when in normal operation.4. Calibrate the new sensor immediately.

36

InitializeThe Initialize option restores the original factory default settings of the detector. 1. Press H to scroll to Initialize andpress C within 20 seconds to confirm.2. From the Initialize? screen perform one of the following options within 5 seconds. · Press A No to exit, or · Press C Yes to initialize.If No is selected, the following screen displays and the detector exits the initialize option.

GasAlertMicro 5/PID/IR User Options MenuIf Yes is selected, the following screen displays while performing the initializing process.When initializing is complete, the following screen displays. The detector deactivates and then immediately reactivates. The detector then reconfigures to the default settings while it performs the self-test. Verify all settings and alarm setpoints, and then calibrate the sensors.Force CalibrationIf enabled, the Force cal option automatically forces the detector to enter calibration if a sensor is overdue upon startup. Press H to scroll to Force cal. Press C to toggle between enable and disable.The detector is shipped with the Force cal option disabled.37

GasAlertMicro 5/PID/IR User ManualBump DailyIf enabled, the Bmp daily option forces the detector to perform a daily bump test to ensure that it is responding to the test gas. Press H to scroll to Bmp daily. Press C to toggle between enable/disable.When enabled, the following screen displays during the startup self-test. The bump test must be performed for the LEL and O2 sensors, otherwise the detector will deactivate. Refer to Bump Daily Enabled for procedures.The detector is shipped with the Bmp daily option disabled.Note If Bmp daily is enabled, the O2 and LEL sensors require daily bump tests whenever the detector is activated (or reactivated) following 00:00 hours (midnight).

Stealth ModeThe Stealth option disables the backlight, visual alarms, and audible alarms when concealment is required.Only the vibrator and the LCD activate during an alarm condition.Press H to scroll to Stealth. Press C to toggle between enable and disable.The detector is shipped with the Stealth option disabled.

38

Sleep ModeNoteBW recommends the Sleep option be enabled when using rechargeable battery packs.If Sleep is enabled, sensor circuits remain active after the detector is deactivated and inserted into the charging cradle.Press H to scroll to Sleep. Press C to toggle between enable/disable.NoteEnabling the Sleep option may require the battery be charged more frequently.The detector is shipped with the Sleep option disabled.

GasAlertMicro 5/PID/IR User Options Menu39

GasAlertMicro 5/PID/IR User Manual

AlarmsThe following table describes the detector alarms and corresponding screens.· During an alarm condition, the detector activates the backlight and displays the current ambient gas reading.

Alarms Low Alarm · Fast beep · Slow flash · L and gas type flash · Vibrator alarm activates High Alarm · Constant beep · Fast flash · L and gas type flash · Vibrator alarm activates

Screen`

· If more than one type or level of alarm exists simultaneously, a multi alarm results.· To change the factory-set alarm setpoints, refer to Calibration and Setting Alarm Setpoints.Table 7. Alarms

Alarms TWA Alarm · Fast beep · Slow flash · L and gas type flash · Vibrator alarm activates STEL Alarm · Constant beep · Fast flash · L and gas type flash · Vibrator alarm activates

Screen

40

Table 7. AlarmsAlarms Multi Alarm · Alternating low and high alarmbeep and flash· L and gas types flash· Vibrator alarm activates Sensor Alarm · One beep every 15 seconds · FAIL flashes above the failedsensorLow Battery Alarm · One beep and two flashesevery 25 seconds· flashes

Screen

GasAlertMicro 5/PID/IR Alarms

Alarms Over Limit (OL) Exposure Alarm · Fast beep and flash · L and gas type flash· Vibrator alarm activates Automatic Deactivation Alarm · Eight beeps and flashes· displays· Vibrator alarm temporarily activatesNormal Deactivation · Three beeps and flashes

Screen

41

GasAlertMicro 5/PID/IR User ManualTable 7. AlarmsAlarms Confidence Beep · One beep, one flash, and onevibrate every 10 seconds

Screen

AlarmsPump Alarm· Two fast beeps and alternating flashes· Vibrator alarm activates· L and J flashNote If enabled, during an alarm condition the Latch (latching alarms) option causes the low and high gas alarms (audible, visual, and vibrator) to persist until the alarm is acknowledged by pressing C and the gas concentration is below the low alarm setpoint. The peak concentrations display continually until the alarm condition no longer exists. Local regulations may require the Latch Alarms option be enabled.42

Alarms MMC/SD Fail Alarm · One beep every 5 seconds · S flashes

Screen

Screen

If the Stealth option is enabled, the detector only vibrates during an alarm; the audible and visual alarms are disabled.a Caution: If the detector emits a low battery alarm, charge the lithium battery pack immediately.

Gas Exposures Computeda WarningTo avoid possible personal injury, do not deactivate the detector during a work shift. TWA and STEL readings reset if the detector is deactivated for more than 5 minutes.Table 8. Computed Gas Exposures

Gas Exposure

Description

TWA (toxic only)

Time-weighted average (TWA) based on accumulated exposure to toxic gases averaged over a work day according to OSHA or ACGIH method.

STEL (toxic only)

Short-term exposure limit (STEL) to gas based on a 5-15 minute user selectable period.

*Peak (MAX)

Peak (MAX) concentration exposure during work shift.

*For oxygen, it is the highest or the lowest concentrationencountered.

Viewing Gas ExposuresPress and hold C until the peak (MAX) gas exposures displays.The TWA gas exposures display next.The STEL gas exposures display last.

GasAlertMicro 5/PID/IR Alarms

43

GasAlertMicro 5/PID/IR User Manual Clearing Gas ExposuresThe exposures automatically clear after 5 minutes when the detector is deactivated. To clear the MAX, TWA, and STEL exposure readings immediately, press and hold C and G simultaneously. The detector displays the following screen during the countdown.Note Hold C and G for the entire countdown, otherwise the MAX, TWA, and STEL exposure readings will not clear.Gas Alarm SetpointsThe gas alarm setpoints trigger the gas alarms and are described in Table 9..

Table 9. Gas Alarm Setpoints

Alarm Low alarmHigh alarmTWA alarm STEL alarm Downscale alarmMulti alarm

ConditionToxics and combustibles: Ambient gas level above low alarm setpoint. Oxygen: Ambient gas level may be set above or below 20.9%.Toxics and combustibles: Ambient gas level above high alarm setpoint. Oxygen: Ambient gas level may be set above or below 20.9%.Toxic only: Accumulated value above the TWA alarm setpoint.Toxic only: Accumulated value above the STEL alarm setpoint.Toxic: If sensor reading is negative (half of the TWA setpoint). LEL: If sensor reading is negative (half of the low alarm setpoint).Two or more gas alarm conditions.

Viewing the Alarm SetpointsTo view the current alarm setpoints for all of the sensors, press G during normal operation.

44

GasAlertMicro 5/PID/IR Alarms

The time/date, TWA, STEL, low, and high alarm setpoint screens display in the following order left to right:.If a correction factor has been applied to a sensor, one of the following screens display indicating the sensor and gas type.

Resetting Gas Alarm SetpointsNote Standard factory alarm setpoints may vary by region.The following table lists the factory alarm setpoints according to the Occupational Safety and Health Association (OSHA) settings.Table 10. OSHA Sample Factory Alarm Setpoints

GasO2 LEL CO H2S PH3 SO2 Cl2 NH3 NO2 HCN ClO2 O3 VOCCO2

TWAN/A N/A 35 ppm 10 ppm 0.3 ppm 2 ppm 0.5 ppm 25 ppm 2.0 ppm 4.7 ppm 0.1 ppm 0.1 ppm 50 ppm 5,000 ppm

STELN/A N/A 50 ppm 15 ppm 1.0 ppm 5 ppm 1.0 ppm 35 ppm 5.0 ppm 10.0 ppm 0.3 ppm 0.1 ppm 100 ppm 30,000 ppm

Low19.5% vol. 10% LEL 35 ppm 10 ppm 0.3 ppm 2 ppm 0.5 ppm 25 ppm 2.0 ppm 4.7 ppm 0.1 ppm 0.1 ppm 50 ppm 5,000 ppm

High23.5% vol. 20% LEL 200 ppm 15 ppm 1.0 ppm 5 ppm 1.0 ppm 50 ppm 5.0 ppm 10.0 ppm 0.3 ppm 0.1 ppm 100 ppm 30,000 ppm

45

GasAlertMicro 5/PID/IR User ManualTo change the factory-set alarm setpoints, refer to Calibration and Setting Alarm Setpoints.Note To disable an alarm, set the alarm setpoint to 0 (zero).Stopping a Gas AlarmThe low and high alarms stop when the ambient gas level returns to below the low alarm setpoint.Note If alarms are set to latch, press C to acknowledge the alarm. The alarms will persist until the alarm is acknowledged and the gas concentration is below the low alarm setpoint.a Caution Follow all safety procedures as defined by your employer. Confirm with your supervisor before clearing TWA and STEL alarms.The TWA and STEL alarms can be stopped either by · clearing the MAX, TWA, and STEL exposures (refer to Clearing Gas Exposures), or · deactivating the detector for 5 minutes (minimum) and then reactivating it again.46

Sensor AlarmThe detector tests for missing or defective sensors during the activation self-test. If a sensor fails the self-test, FAIL flashes on the LCD above the failed sensor. Refer to Troubleshooting.Pump AlarmThe external pump draws air over the sensors continually. If the pump stops operating or becomes blocked, the detector activates the pump alarm. The following screens display.The pump alarm continues until the blockage is cleared or it is acknowledged by pressing C. If C is pressed, the detector automatically launches a pump test to reset the pump module.

GasAlertMicro 5/PID/IR Bump Test

Refer to Pump Test for more information. If the pump test is successful, the detector returns to normal operation, otherwise the pump alarm continues. If the pump alarm persists, refer to the Pump Operation section in Troubleshooting.Low Battery AlarmThe detector tests the batteries upon activation and continuously thereafter. Battery power is continually displayed during normal operation. If the battery voltage is low, the detector activates the low battery alarm.The low battery alarm continues until the batteries are replaced/charged, or until the battery power is almost depleted. If the battery voltage becomes too low, the detector deactivates.NoteTypically, the low battery alarm continues for 30 minutes before the detector automatically deactivates.Automatic Deactivation AlarmIf the battery voltage is in immediate danger of falling below the minimum operating voltage, the audible alarm beeps eight times and the visual alarm flashes eight times. After 3 seconds, the LCD dims and the detector deactivates.To replace or charge the batteries, refer to the following sections:· Replacing the Alkaline Batteries · Replacing the Lithium Battery Pack

Bump TestA bump test is the process of applying a small amount of test gas to force the detector into alarm. A bump test should be performed regularly to confirm the sensors are responding correctly to gas, and that the audible, visual, and vibrator alarms activate during an alarm condition. Calibrate if the readings are not within specified limits.Performing a Bump Testa Caution BW recommends to bump test the sensors before each day’s use to confirm their ability to respond to gas by exposing the sensors to a gas concentration that exceeds the alarm setpoints.To perform a manual bump test refer to Applying Gas to the Sensors on page 50 and the following steps 1-6. Follow this procedure when Bump Daily is enabled.NoteThe Calibration icon will flash if a sensor has failed a bump test.NoteIf performing a bump test on a pump unit, connect the calibration hose directly to the pump module.

47

GasAlertMicro 5/PID/IR User Manual1. Connect the calibration hose to the 0.5l/min regulator on the gas cylinder.NoteOnly use the calibration cap for bump tests and calibrations.2. Connect the calibration hose to the intake inlet on the calibration cap. Arrows on the calibration cap indicate the direction of gas flow.3. Attach the calibration cap and tighten the knob. Refer to Applying Gas to the Sensors on page 50.NoteEnsure the knob is securely fastened before applying gas. Do not overtighten the calibration cap. Ensure strong air currents are kept away from the exhaust side of the calibration cap, as this can affect the bump test accuracy.4. Apply gas. Verify the visual, audible, and vibrator alarms activate.5. Close the regulator and remove the calibration cap from the detector. The detector will temporarily remain in alarm until the gas clears from the sensors.6. Disconnect the calibration hose from the cap and regulator.

Calibration and Setting Alarm SetpointsGuidelinesWhen calibrating the detector, adhere to the following guidelines:Recommended gas mixture: CO: 50 to 500 ppm balance N2 H2S: 10 to 100 ppm balance N2 PH3: 1 to 5 ppm balance N2 SO2: 10 to 50 ppm balance N2 Cl2: 3 to 25 ppm balance N2 NH3: 20 to 100 ppm balance N2 NO2: 5 to 50 ppm balance N2 HCN: 5 to 20 ppm balance N2 ClO2: 0.1 to 1.0 ppm balance N2 O3: 0.1 to 1.0 ppm balance N2 VOC:100 ppm isobutylene LEL: 50% LEL or 2.5% by vol. methane balance air CO2: 5000 ppm balance air O2: 18.0%· To ensure accurate calibration, use a premium-grade calibration gas. Gases approved by the National Institute of Standards and Technology (NIST) improves the validity of the calibration.· Do not use a gas cylinder beyond its expiration date.

48

GasAlertMicro 5/PID/IR Calibration and Setting Alarm Setpoints

· Calibrate a new sensor before use. Install the sensor, activate the detector, and allow the sensor to stabilize before starting calibration. Used sensor: 60 seconds / New sensor: 5 minutes· Calibrate the detector at least once every 180 days (every 90 days for HCN sensors) depending upon use and sensor exposure to poisons and contaminants. For a list of common products that can damage sensors, refer to Sensor Poisons and Contaminants.· Calibrate the detector if the ambient gas varies during startup.· Calibrate the sensor before changing the alarm setpoints.· Calibrate only in a safe area that is free of hazardous gas in an atmosphere of 20.9% oxygen.· To disable an alarm, set the alarm setpoint to 0 (zero).· If the Auto cal option is enabled, the oxygen (O2) sensor calibrates automatically every time the detector is activated. Activate the detector in a safe area that is free of hazardous gas in an atmosphere of 20.9% oxygen.· After activating the detector, allow it to stabilize for 1 minute before performing a calibration or bump test.· If a certified calibration is required, contact BW Technologies by Honeywell.NoteA generator must be used for O3 and ClO2 sensors.

Diagnostics TestingThe detector tests the ambient air (auto zero) and the test gas that is applied (auto span) to ensure it meets expected values. Auto zero sets the zero-gas level of the sensor. If ambient gas is present, the zero level will be incorrect. If excessive target gas is present, the detector displays an error message and lists the affected sensor.In auto span, if the target gas is not detected or does not meet expected values, a message displays that the detector is exiting calibration mode. The detector retains the previous set values.

49

GasAlertMicro 5/PID/IR User Manual Applying Gas to the SensorsThe calibration cap, single gas calibration cap, and hose are shipped with the detector. Refer to Figure 3. and Table 11. for installation.Note The calibration cap and single gas calibration cap must only be used during the calibration span process.Figure 3. Applying Gas to the Sensors

Table 11. Applying Gas to the Sensors

Item1 2 3 4 5

DescriptionDetector with calibration cap Calibration hose Regulator and gas cylinder Toxic 2 sensor position Single gas calibration cap

Single Gas Calibration Cap a CautionIf an O3 or ClO2 sensor is located in the Toxic 2 position (refer to Figure 3. and Table 11.), a single gas calibration cap must be used to ensure accurate calibration.To calibrate O3 and ClO2 sensors using the single gas calibration cap, refer to Figure 4., Table 12., and complete the following procedures.1. Insert the cap into the Toxic 2 sensor position on the detector (refer to Figure 3.). Press firmly until the release tabs click.2. Connect the calibration hose to the gas cylinder and to the intake inlet on the cap.3. Proceed to Calibration Procedure.

50

Table 12. Single Gas Calibration Cap

Item1 2 3 4

Description Intake inlet Calibration hose Gas flow direction arrow Output outlet

GasAlertMicro 5/PID/IR Calibration and Setting Alarm SetpointsRemoving the Single Gas Calibration Cap Using the thumb, push forward against both the inlet and the outlet simultaneously to remove the cap from the detector.

Figure 4. Single Gas Calibration CapNote The arrow on the cap indicates the direction of gas flow from intake to outtake.

Figure 5. Removing the Single Gas Calibration CapCalibration ProcedureTo calibrate the detector and set the alarm setpoints, perform the following procedure.Note To bypass a step during the calibration process (after auto zero), press A.

51

GasAlertMicro 5/PID/IR User ManualStart CalibrationNoteVerify that the calibration gas being used matches the span concentration value(s) that are set for the detector. Refer to Span Gas Value. Correction factors are not applied during calibration. Correction factors that were set prior to calibration are restored when the detector returns to normal operation.NoteThe calibration icon will flash if a sensor has an overdue calibrationa Caution Calibrate only in a safe area that is free of hazardous gas in an atmosphere of 20.9% oxygen. 1. To enter calibration, press and hold C and H simultane-ously as the detector beeps, flashes, and vibrates to the corresponding countdown.

Auto Zero and Oxygen (O2) Sensor Calibration

2.

flashes while the detector automatically zeroes the

toxic and combustible sensors, and calibrates the O2 sensor.

52

Depending upon the detector being calibrated, the auto zero screens display differently.GasAlertMicro 5/PID GasAlertMicro 5 IRNote Do not apply calibration gas during this process, otherwise the auto zero step will fail.Zero CO2 (GasAlertMicro 5 IR only)If calibrating using the GasAlertMicro 5 or the GasAlertMicro PID, proceed to Auto Span step #5. If the detector is passcode protected, proceed to Passcode Protect step #4.3. For GasAlertMicro 5 IR models, there is a second auto zero that is performed for the CO2 sensor. When auto zero for the other sensors is complete, the following screen displays.

GasAlertMicro 5/PID/IR Calibration and Setting Alarm Setpoints

Press C Yes to zero the CO2 sensor.OrPress A No to bypass the CO2 zero and proceed to Auto Span step #5. If the detector is passcode protected, refer to Passcode Protect Activated step #4.If C Yes is pressed, the following screen displays.

flashes whiles the detector zeros the CO2 sensor (approximately 30 seconds).Passcode Protect Activated4. When auto zero is complete and if the passcode protect option is enabled, the following screen displays.

a Warning Use nitrogen (N2) to zero the CO2 sensor.Apply N2 to zero the CO2 sensor. The following screen displays.

The passcode must be entered to proceed. Press G or H to scroll to the correct passcode and then press C within 5 seconds to confirm. If entered correctly, the detector beeps twice and proceeds to Auto Span step #5. Incorrect Passcode: If the passcode is incorrect or is not confirmed within 5 seconds by pressing C, the following screens display.53

GasAlertMicro 5/PID/IR User ManualThe detector saves the calibration and returns to normal operation.Auto Span5. When auto zero is complete, the following screens display. GasAlertMicro 5 and PIDGasAlertMicro 5 IR

.Note Span sensors in the following order:· Exotics (NH3, ClO2, O3, Cl2, and CO2)· Single gas· Quad gas (H2S, CO, LEL, and O2)· PIDNote A generator must be used for O3 and ClO2 sensors. To ensure accurate calibration, a single gas calibration cap must be used to calibrate O3 and ClO2 sensors.· ClO2: Use a Tedlar bag as a buffer between the generator and the detector (using the single gas calibration cap) to regulate the flow rate to ensure accurate readings.Allow the Tedlar bag to fill for several minutes before initiating calibration.· O3: Calibrate using the generator and the single gas calibration cap. Do not use a Tedlar bag.Set the generator to 0.5 ppm at a flow rate of 0.5 l/min.Depending upon the gas cylinder being used, one or all four sensors can be calibrated at one time. Apply Span Gas Now

54

Attach the calibration cap (or single gas calibration cap for O3 and ClO2) and apply gas to the sensor(s). To attach caps, refer to Figure 3..Refer to the following flow rates:

· 1000 ml/min. for NH3 and Cl2

· 500 ml/min. for CO2

· 250-500 ml/min. (all other sensors)

K flashes as the detector initially detects the calibration gas.

GasAlertMicro 5 and PID

GasAlertMicro 5 IR

GasAlertMicro 5/PID/IR Calibration and Setting Alarm Setpoints

Exotic toxic gases

5 minutes

LEL (combustibles)

30 seconds

PID gases

2 minutes

Insufficient Level: If a sensor does not attain a sufficient level of expected gas, it is cleared from the LCD and is not spanned.

After 30 seconds the detector beeps and K stops flashing. flashes while spanning the sensors until the detectorhas attained a sufficient level of the expected gas.Refer to Table 13. for times required to span.Table 13. Time Required to Span

Gas Type Most toxic gases and CO2

Time Required to Span 2 minutes

55

GasAlertMicro 5/PID/IR User Manual

While the detector is spanning the sensor(s), a countdown of time remaining displays in the lower left of the screen.

GasAlertMicro 5 and PID

GasAlertMicro 5 IR

Successful Span6. If the sensor(s) has spanned successfully, the audible alarm beeps three times and the following screens display.

When the span is complete, the following screen displays.Successful Span: Proceed to Successful Span step #6. Unsuccessful Span: Refer to Unsuccessful Span for possible solutions.

If there are more sensors to span, remove the current cylinder and connect the next gas cylinder. Press G and apply gas to span the other sensor(s).Or Press H to end the span and proceed to Setting the Calibration Due Date step #7.When all sensors have spanned, the following screen displays.

If a sensor fails to span, refer to Unsuccessful Span at the end of the calibration section for possible solutions.56

Setting the Calibration Due Date7. When the span is complete, the calibration due date can be set for each sensor that has spanned successfully. The following screen displays.Press C to set the calibration due dates. Or Press A to bypass and proceed to Alarm Setpoints step #8.Note If a new due date is not entered, the detector automatically resets to the previously entered number of days.

GasAlertMicro 5/PID/IR Calibration and Setting Alarm SetpointsThe calibration due dates are set in the following order: · Toxic 1 · Toxic 2 · LEL · O2 If the due date of an unsuccessfully spanned sensor is changed, the following screens display.The detector then automatically proceeds to the next sensor. To change the calibration due date (1-365 days), press H or G until the new value displays. Press C within 5 seconds to confirm. Or Press A to bypass a sensor and proceed to the next sensor.57

GasAlertMicro 5/PID/IR User ManualNote If a value is changed but C is not pressed within 5 seconds to confirm, the following screen displays.

Alarm Setpoints8. When all of the sensor due dates have been set or bypassed, the alarm setpoints need to be set or bypassed. The following screen displays.

The previous value is automatically retained. The detector proceeds to the next sensor due date.Repeat step #7 to set the calibration due dates for the remaining sensors.When all of the due dates are entered, press C to proceed to Alarm Setpoints step #8.OrPress A to bypass setting the alarm setpoints and proceed to Finish Calibration step #10.

Factory alarm setpoints may vary by region. Refer to Resetting Gas Alarm Setpoints for an example.Alarms can be set to any value within the detection range of the selected sensor. Refer to Detection Range in Specifications.NoteTo disable an alarm setpoint, set it to 0 (zero).

58

If a new setpoint is entered but not confirmed within 5 seconds by pressing C, the following screen displays.

GasAlertMicro 5/PID/IR Calibration and Setting Alarm SetpointsSetting the TWA Alarm Setpoint The current TWA alarm setpoint displays for the selected sensor (if applicable).

The previous setpoint is retained and the detector proceeds to the next setpoint. The setpoints are set in the following order:· TWA (if applicable)· STEL (if applicable)· low· high To bypass a setpoint, press C to save the current value and proceed to the next setpoint.

Press H or G to change the value for the TWA alarm setpoint. When the required value displays, press C to confirm.Setting the STEL Alarm SetpointThe current STEL alarm setpoint displays for the selected sensor (if applicable).

Press H or G to change the value for the STEL alarm setpoint. When the required value displays, press C to confirm.59

GasAlertMicro 5/PID/IR User ManualSetting the Low Alarm Setpoint The current low alarm setpoint displays for the selected sensor.

Setting the Remaining Alarm Setpoints9. Repeat step #8 to set the alarm setpoints for the remaining sensors. When complete, the detector emits two quick beeps and proceeds to Finish Calibration.Finish Calibration10. The detector displays the following screen to indicate that calibration is complete.

Press H or G to change the value for the low alarm setpoint. When the required value displays, press C to confirm.Setting the High Alarm SetpointThe current high alarm setpoint displays for the selected sensor.

Press H or G to change the value for the high alarm setpoint. When the required value displays, press C to confirm. Proceed to step #9 Setting the Remaining Alarm Setpoints.60

Verification1. After calibration is complete and the detector returns to normal operation, verify the calibration by using a gas cylinder other than the one used for calibration.2. The gas concentration should not exceed the sensor’s detection range. Confirm that the LCD displays the expected concentration values.3. To ensure the reading is accurate, apply the verification gas for the same amount of time as was applied to the sensor when it was calibrated.Example: SO2 span time 2 minutes therefore, apply verification gas for 2 minutes.

GasAlertMicro 5/PID/IR Calibration and Setting Alarm SetpointsUnsuccessful SpanIf the sensor(s) did not span successfully, refer to the following sections for possible solutions:· Failed Span · No Gas Detected · Did Not Reach Target Span · Large SpanFailed Span If a sensor fails the span, the following error message displays.

If the sensor is not calibrated, it displays as FAIL in normal operation the next time the detector is activated.

61

GasAlertMicro 5/PID/IR User ManualIf all sensors fail the span, the following screen displays.

No Gas DetectedIf the detector does not detect any gas within 30 seconds, the following screens display.

1. Ensure the regulator valve is open and that the gas cylinder is not past the expiration date. Replace the gas cylinder and/ or regulator if required.2. Replace the sensor(s) and sensor filter if required.3. Press A to exit and then calibrate again in a safe area that is free of hazardous gas in an atmosphere of 20.9% oxygen.4. If the span fails a second time, reactivate the detector to test the sensors.If all sensors fail the span, the calibration due dates cannot be reset. The following screens display.

1. Ensure the regulator valve is open and the calibration hose is free of debris.2. Replace the gas cylinder and/or regulator if required.3. Replace the sensor(s) and sensor filter if required. 4. Press G to reapply gas or press H to end the span.Did Not Reach Target Span If the target span was not achieved, as defined in the Span Gas Value option, the following screens display.

62

Not reaching the target span can result from · a problem with the span gas, · the gas cylinder being past the expiry date, or · a problem with the sensor. Accept Current Span: If the gas cylinder, regulator, and sensor are operating correctly, press C to accept the current span. Reject Current Span: Press A to reject. Verify the gas cylinder, detector settings, and sensor are operating correctly. Calibrate the sensor again.Large Span If the span adjustment is unusually large (more than 15%), the following screens display.Ensure the correct gas cylinder being used and that the span concentration value(s) of the detector matches the span concentration value(s) of the gas cylinder. Refer to Span Gas Value.Adjustment Expected: If the calibration adjustment is expected, press C to accept the span. Adjustment Not Expected: If the calibration adjustment is not expected or the span gas value does not match the gas cylinder, press A to reject the span and calibrate that sensor again.

GasAlertMicro 5/PID/IR Calibration and Setting Alarm Setpoints63

GasAlertMicro 5/PID/IR User Manual

PumpWhen using the pump module, attach it and the pump accessories prior to activating the detector.The pump module can be used independently to measure gas or it can be used with the sample probe to test for gases in confined spaces. To measure gas using the sample probe, refer to Confined Space Sampling.a Warning If the pump module is installed on the detector, the following three things must occur during start-up. If any one of the conditions below does not occur, discontinue use of the detector and contact BW Technologies by Honeywell immediately.· The detector prompts for a pump test during start-up· The pump module passes the pump test at start-up when the pump inlet or sample chain inlet is blocked· The J icon displays on the LCDTo ensure accurate gas detection, the sensors must be calibrated immediately when the pump module is replaced by the diffusion cap and vice versa. When using the sample probe at -10°C to +0°C (14°F to 32°F), keep the sample probe in your hand.

a Caution A demand flow regulator must be used to manually calibrate the GasAlertMicro 5/PID/IR detector when the pump module is installed. The calibration cap is designed for use with the diffusion cap only. It cannot be used with the pump module.Identifying the PumpThere are two generations of pump. The best way to differentiate the Generation 2 pump from the Generation 1 pump is the inline filter that is visible on the Generation 2 pump on the left-hand side.Each model of pump has different operating specifications. Refer to Maximum Hose Length when Confined Space Sampling on page 71.Generation 1: 116885-L3 (yellow) and 118933-L3 (black)

64

Generation 2: 130916-L3 (yellow) and 130917-L3 (black)

GasAlertMicro 5/PID/IR Pump

65

GasAlertMicro 5/PID/IR User ManualInstalling the Pump ModuleFigure 6. Installing the Pump Module

Table 14. Installing the Pump Module

Item 1 2 3 4

Description Pump module Sensor filter Detector Machine screws (2)

To install the pump module, refer to Figure 6., Table 14., and the following procedures.

1. Deactivate the detector.2. Remove the two machine screws and the sensor cover. Remove the sensor filter from the sensor cover and insert it into the pump module.3. Attach the pump module to the detector and replace the two machine screws.4. If sampling in a confined space, attach the sample probe. Refer to Confined Space Sampling.If using an auxiliary filter as an additional filtration system, refer to Attaching the Auxiliary Filter.5. Activate the detector. The detector performs a pump test during the startup self-tests. Refer to Pump Test.

66

Replacing the Pump Filter (Generation 2 Pump)To replace the Generation 2 pump filter, refer to Figure 7. and the following steps 1-8.NoteIf using the auxiliary filter on the Generation 1 pump or as an additional filtration system on the new pump module, refer to Attaching the Auxiliary Filter.

GasAlertMicro 5/PID/IR Installing the Pump Module1. Deactivate the detector.2. Remove the filter window screw.3. Remove the filter window.4. Remove the old filter from the filter window.Important! Ensure the filter cavity and filter window are clean and free of debris.5. Insert a new filter.6. Replace the filter window and screw.7. Activate the detector. The detector performs a pump test during the startup self-tests to verify the pump module is operating correctly.8. To order additional filters, refer to Replacement Parts and Accessories. a CautionFilters may need to be replaced more frequently in high particulate areas. BW Technologies by Honeywell recommends that the auxiliary filter be used as a additional filtration in high particulate areas.

Figure 7. Replacing the Pump Filter (Generation 2 Pump) 67

GasAlertMicro 5/PID/IR User Manual

Replacing the Pump Nozzle (Generation 2 Pump)To replace the pump nozzle for the Generation 2 pump, refer to Figure 8. and step 1-4.

3. Insert the new nozzle. Ensure the nozzle post inserts correctly into the nozzle gasket.4. Activate the detector. The detector performs a pump test during the startup self-tests to verify the pump module is operating correctly.

Attaching the Auxiliary FilterThe Pump Module Auxiliary Filter (“the filter”) is used to filter out particulates that can decrease the life span of the pump.

Figure 8. Replacing the Pump NozzleNoteIf the nozzle is damaged, replace it immediately to ensure accurate pump flow. To order additional pump nozzles, refer to Replacement Parts and Accessories.1. Deactivate the detector.2. Gently insert a medium-sized flathead screwdriver into the nozzle slot. Twist the screwdriver slightly to lift and remove the damaged nozzle.68

Figure 9. Attaching the Auxiliary Filter

a CautionThe filter is designed to protect the pump. For Generation 1 pumps, the filter must be connected when the pump is activated. Only remove the filter to perform calibrations, bump tests, and when using the sample probe.For new model pumps, the auxiliary filter can be used as an additional filtration system in high particulate areas.To insert the auxiliary filter, complete the following:1. Loop the filter cord through the ring on the alligator clip and pull the filter through the loop of the filter cord.

GasAlertMicro 5/PID/IR Installing the Pump ModuleNoteAttaching the filter cord to the alligator clip ensures the filter remains with the detector when not in use.2. Pull the filter cord tight. Ensure the loop knot is secure. 3. Align the pump quick connector with the filter quick connectorand turn clockwise. Ensure the connection is secure. 4. Change the filter as required. Environments with heavy air-borne particulates require the filter to be changed frequently. To order additional filters, refer to Replacement Parts and Accessories.

Figure 10. Attaching the Filter Cord 69

GasAlertMicro 5/PID/IR User Manual Confined Space SamplingThe sample probe is used to safely test for gas in confined spaces before entering. Attach the sample probe to the pump module prior to activating the detector.a Warning To measure hazardous gas in a confined space, the sample probe must be used with the pump module.70

Figure 11. Attaching the Sample Probea Caution To prevent the Teflon lining inside the Tygon tubing from causing a blockage when connecting it to the sample probe, the end of the tubing must be flared. Refer to steps 3 and 4.NoteThe detector and pump module can also be used with a sintered filter.1. Deactivate the detector.2. Attach the pump connector end of the teflon-lined Tygon tubing to the nozzle on the pump module.3. Gently insert needle nose pliers into the other end of the teflonlined Tygon tubing. Using a circular motion, flare the end of the tubing.4. Connect the flared end of the tubing to the sample probe. Ensure the teflon lining does not separate from the Tygon tubing, as it will block the tube and generate a pump alarm.5. Activate the detector. Ensure all connections are secure before sampling.6. Insert the sample probe into the confined space.a Warning Depending upon the length of the tubing and the type of gas in the confined space, allow a minimum of 3

seconds per foot of tubing to ensure the readings stabilize before entering the area. Example: 10 ft. = 30 secondsMaximum Hose Length when Confined Space SamplingMaximum sample hose length is dependent on pump model, operating temperature, and sampling accessory. Refer to the tables below.a Caution Maximum sample hose length is dependent on an inner 1/16″ hose diameter. Table 15. Generation 1 Pump Maximum Hose Length

Sintered filter Sample Probe

-10°C to +0°C (14°F to 32°F) 3 m (10 ft.)3 m (10 ft.)

0°C to 50°C (32°F to 122°F) 9.1 m (30 ft.)3 m (10 ft.)

a WarningWhen using the sample probe at -10°C to +0°C (14°F to 32°F), keep the sample probe in your hand.

GasAlertMicro 5/PID/IR Installing the Pump Module

Table 16. Generation 2 Pump Maximum Hose Length

Sintered filter Sample Probe

-20°C to 50°C (-4°F to 122°F) 20 m (66 ft.)3 m (10 ft.)

71

GasAlertMicro 5/PID/IR User Manual

DataloggerDetectors equipped with the datalogger option record information that can be compiled to create a report. To set how often the detector records a sample (1-127 seconds), refer to Logger Option in the user options menu. The following information is recorded in a datalog:· Date and time · Serial number of the detector · Type of gas the detector monitors · Gas reading(s) that display · STEL and TWA readings · Sensor status · Detector status · Passcode protect enabled/disabled · STEL period setting · Confidence beep enabled/disabled · Automatic backlight enabled/disabled · Stealth mode is enabled/disabled · Latching alarm enabled/disabled · Calibration past due option enabled/disabled · Language the detector is set to display

MMC/SD Card CompatibilityFor a list of compatible memory cards, please contact BW Technologies by Honeywell.Inserting the MMC/SD CardTo insert the MMC/SD card into the detector, refer to the following steps 1-5 and Figure 12..1. Deactivate the detector. 2. Release the latch and remove the battery pack. 3. Insert the MMC/SD card (pins face down). 4. Replace the battery pack and secure the latch. 5. Activate the detector. The MMC/SD card is automatically for-matted during the startup self-test.

72

Figure 12. Inserting/Removing the MMC/SD Card

GasAlertMicro 5/PID/IR MMC/SD Card TroubleshootingMMC/SD Card TroubleshootingThe MMC/SD card is not required for operation in detectors equipped with datalogging. However, the following two screens display if the card is not inserted during startup.

A new MMC/SD card is automatically formatted when it is inserted in the detector. When the detector is activated, it begins the self-test and then displays the following screen.

73

GasAlertMicro 5/PID/IR User ManualRestoring Datalog FilesIf the MMC/SD card has been accidentally reformatted or erased by the computer application, the following screens display when the card is inserted into the detector.

If the detector successfully restores the logfile, the following screen displays and the startup tests continue.

a CautionOnly erased data files can be restored using the detector. Computer applications sometimes write data over erased files, and that erased data cannot be restored by the detector. Always create back up files on the computer.To restore the logfile, complete the following:1. From the detector, press H to restore the logfile. The following screen displays.

2. From the computer, verify that the logfile has been restored. When the normal operation screen displays, deactivate the detector.3. Remove the MMC/SD card and insert it into the card reader.4. From the computer desktop, double-click My Computer to view the list of drives.5. Double-click the Removable Disk drive to access LOGFILE0.csv. Open the logfile and verify that the data has been restored.If LOGFILE0.csv does not display, ensure that the MMC/SD card is inserted in the card reader correctly and that all connections are secure.6. After verifying that the logfile has been restored, re-insert the MMC/SD card into the detector.

74

Reformatting the MMC/SD CardTo reformat the MMC/SD card, complete the following:1. Insert the MMC/SD card into the card reader.2. From the computer desktop, double-click My Computer to view the list of drives.3. Double-click the Removable Disk drive to access LOGFILE0.csv.4. Select LOGFILE0.csv and delete. 5. Insert the MMC/SD card into the detector.6. Activate the detector. The startup self-test begins and the following screens display.

GasAlertMicro 5/PID/IR MMC/SD Card Troubleshooting7. Press G to format the MMC/SD card. The following screen displays.For any additional MMC/SD card errors, refer to Troubleshooting.

75

GasAlertMicro 5/PID/IR User Manual

Import Datalogs to Fleet Manager IINote Refer to the following minimum requirements before importing datalogs to Fleet Manager II.Minimum PC Requirements· 500 MHz Pentium (or equivalent) · 100 MB free hard disk space · Windows XP or Vista · USB portImporting from MicroDock II to Fleet Manager IIIf the detector is used with the MicroDock II Station to import datalogs to Fleet Manager II, refer to the MicroDock II User Manual for complete instructions.

Import to Fleet Manager II Using a Card ReaderTo import a datalog file from the detector to Fleet Manager II, complete the following:1. Deactivate the detector.2. Release the latch and remove the battery pack.3. Remove the MMC/SD card from the detector. Refer to Figure 12. and 1.4. Connect the card reader to the USB port on the computer.5. Insert the MMC/SD card into the card reader (ensure the pins face down).6. Open Fleet Manager II. Click Start > All Programs > BW Technologies > Fleet Manager II > BW Technologies Fleet Manager II.7. From the top task bar, click Help > User Manual. 8. Refer to section “Importing Data into Fleet Manager II”.

76

View Datalog Files in SpreadsheetsThe datalog files can be downloaded from the MMC/SD card into most spreadsheet applications using a card reader.Compatible software applications are· Microsoft® Excel 98 or higher, · Quattro Pro, · Lotus 1-2-3, · Microsoft® Access, and · Microsoft® Word. To view a datalog file in a software spreadsheet, complete the following:1. Deactivate the detector and remove the MMC/SD card (refer to Figure 12.).2. Insert the MMC/SD card into the card reader.3. From the computer desktop, double-click My Computer to view the list of drives.4. Double-click Removable Disk drive. 5. Double-click LOGFILE0.csv. 6. Refer to Table 17. for an example of the datalog spreadsheet.The Unit Config column (far right) in Table 17. contains letter codes. Refer to Table 18. and Table 19. for definitions of the codes.

GasAlertMicro 5/PID/IR View Datalog Files in Spreadsheets77

GasAlertMicro 5/PID/IR User Manual

Example of a Datalog SpreadsheetWhen datalog information is imported into most spreadsheet software, it appears similar to the example below.a Warning: Some compatible software packages have an internal file size limit and may not load the entire file. Check the software limit.

Note Not all columns are included in this example. Additional Toxic TWA and Toxic STEL display on a normal spreadsheet.Table 17. Datalog Spreadsheet Example

Date

Day

dd-mm-yy Mon=1

23-12-05 #4 23-12-05 #4 23-12-05 #4 23-12-05 #4 23-12-05 #4 23-12-05 #4 23-12-05 #4 23-12-05 #4 23-12-05 #4 23-12-05 #4 23-12-05 #4 23-12-05 #4 23-12-05 #4 23-12-05 #4

Time hh:mm:ss9:54:25 9:54:30 9:54:35 9:54:40 9:54:45 9:54:50 9:54:55 9:55:00 9:55:05 9:55:10 9:55:15 9:55:20 9:55:25 9:55:30

Toxic1 ppm5 10 5 10 0 0 0 0 5 10 5 10 0 0

Toxic2 ppm10 15 10 15 0 0 0 0 10 15 10 15 0 0

Toxic3 ppm35 50 35 200 0 0 0 0 35 50 35 200 0 0

LEL %CH4 %LEL10 20 24 24 24 2410 20 24 24

O2

Toxic 1

Toxic 1

Status

%

TWA ppm STEL ppm Codes

Serial Number

Unit Config

33——

S104-000001

44——

S104-000001

19.5 —

1111—-

S104-000001

23.5 —

2222—-

S104-000001

20.9 —

-D-E—D S104-000001 FCEKNL

20.9 0

0

——–

S104-000001 FCEKNL

20.9 0

0

–L—-L

S104-000001 FCEKNL

20.9 0

0

LLH—-M S104-000001 FCEKNL

0

0

LLH—-M S104-000001

0

0

LLL—-M S104-000001

19.5 0

0

-L—–L

S104-000001

23.5 0

0

——–

S104-000001

20.9 0

0

——-B-

S104-000001 FCEKNL

20.9 0

0

——B-

S104-000001 FCEKNL

78

Table 18. Datalog Status Codes

Normal operationL Low alarm H High alarm T TWA alarm U TWA and low alarm (dual alarms) V TWA and high alarm (dual alarms) s STEL alarm u STEL and low alarm (dual alarms) f Fresh air delay @ Zero CO2P Plugged (blocked) – pump alarmBatteries OKL Low alarm H High alarm T TWA alarm

Status Codes G Backlight is on v STEL and high alarm (dual alarms) w TWA and STEL alarm (dual alarms) x TWA, STEL, and low (triple alarms) y TWA, STEL, and high (triple alarms) O Overload / sensor is over-ranged C Calibrating F Failure – sensor failure I Time setPump Codes F Failure / pump failureBattery Status Codes B Low battery alarmAlarm Status Codes M Multi alarm C Calibration Q Off / quit / manual deactivation

Note: TWA readings greater than 99 are recorded as OL.

GasAlertMicro 5/PID/IR View Datalog Files in Spreadsheets1 Alarm setpoint 1 (low alarm) 2 Alarm setpoint 2 (high alarm) 3 Alarm setpoint 3 (TWA alarm) 4 Alarm setpoint 4 (STEL alarm) D Calibration due date (days) E Elapsed / last calibration (days) Z Auto zeroing t TestingC Confidence beep is activeS Automatic shutdown F Failure / self-test fail R RTCC / real-time clock failure

79

GasAlertMicro 5/PID/IR User Manual

Table 19. Datalog Gas and Correction Factor Sensor Codes

A No sensor

B

F

SO2

G

K NH3

L

Q IR

A Acetaldhyde

B

F Diesel

G

K Isobtyln

L

P Styrene

Q

U Custom

A Acetone

B

F Ethyl_Ace

G

K Isobutylene

L

P Octane

Q

U Custom

V LEL in % by Vol CH4

H2S PH3 ClO2

Gas Sensor Codes

C

H2S COSH

D

H

NO2

I

M

O3

O

CO HCN LEL

Correction Factor Codes for PID (if applicable)

Acetone Ethanol JP8 Toluene

C

Ammonia

H

Ethylene

M

Kerosene

R

Turpentine

D

Benzene

I

Gasoline

N

MEK

S

Vinyl_Cl

Benzene Gasoline Isopropanol Pentane

Correction Factor Codes for LEL

C

Butane

D

H

Heptane

I

M

MEK

N

R

Propane

S

Cyclohexane Hexane Methane Toluene

LEL Unit Codes

L

LEL in % LEL

E

CO COSH

J

Cl2

P

PID

E

Butadiene

J

Hexane

O

Naptha

T

Xylene

E

Ethanol

J

Hydrogen

O

Methanol

T

Turpentine

80

MaintenanceTo maintain the detector in good operating condition, perform the following basic maintenance as required.· Calibrate, bump test, and inspect the detector at regular intervals. · Maintain an operations log of all maintenance, calibrations, bumptests, and alarm events. · Clean the exterior with a soft damp cloth. Do not use solvents,soaps, or polishes. Refer to Sensor Poisons and Contaminants. · Do not immerse the detector in liquids.Battery Cautions a WarningTo avoid personal injury and/or property damage, adhere to the following:· Replace the batteries immediately when the detector emits a low battery alarm.· Use only batteries that are recommended by BW Technologies by Honeywell. Refer to Specifications.· Ensure the alkaline batteries are properly installed in the detector battery pack.· Charge batteries using only a recommended BW charger. Do not use any other charger. Failure to adhere to this caution can cause fire and/or explosion.· The detector must be deactivated to charge the battery.

GasAlertMicro 5/PID/IR Maintenance· Do not calibrate the detector during or immediately after charging.· Both the rechargeable lithium battery pack and the alkaline battery pack are user-changeable in hazardous locations, but the alkaline battery cells inside the pack can only be replaced in a safe area that is free of hazardous gas.· Warning: The battery pack (M5-BAT08) that is equipped with a lithium polymer cell may present a risk of fire or chemical burn hazard if misused. Do not recharge, disassemble, heat above 212ºF (100ºC), or incinerate.· Warning: Do not use any other lithium batteries with the Micro5/PID/IR detectors. Use of any other cell can cause fire and/or explosion. To order a replacement lithium battery pack (M5-BAT08), refer to Replacement Parts and Accessories.· Warning: Lithium Polymer cells exposed to heat at 266ºF (130ºC) for 10 minutes can cause fire and/or explosion.· Dispose of used lithium cells immediately. Do not disassemble and do not dispose of in fire. Do not mix with the solid waste stream. Spent batteries must be disposed of by a qualified recycler or hazardous materials handler.· Keep lithium cells away from children.Charging the BatteryTo charge the battery, refer to the GasAlertMicro 5/PID/IR Charger Instruction Sheet.81

GasAlertMicro 5/PID/IR User Manual Replacing the Alkaline BatteriesTo replace the alkaline batteries, refer to Figure 13., Table 20., and the following steps 1-6.Figure 13. Replacing the Alkaline Batteries

Table 20. Replacing the Alkaline Batteries

Item1 2 3 4 5 6 7

DescriptionDetector Latch Battery pack Battery tray Captive screws (2) Alkaline batteries (3) Battery shell

1. Deactivate the detector. 2. Open the latch on the bottom of the detector. 3. Remove the battery pack by lifting the bottom of the pack awayfrom the detector. 4. On the battery pack, remove the two captive screws. Removethe battery shell from the battery tray. 5. Replace the three alkaline batteries. 6. Replace the battery pack and the captive screws. 7. Reinsert the battery pack and secure the latch.

82

Replacing the Lithium Battery PackTo replace the lithium battery pack, refer to Figure 14. and the following steps 1-4.Figure 14. Replacing the Lithium Battery Pack

GasAlertMicro 5/PID/IR Maintenance1. Deactivate the detector. 2. Open the latch on the bottom of the detector. 3. Remove the battery pack by lifting the bottom of the packupward from the detector. a WarningDo not disassemble lithium battery packs. Read and adhere to the cautions in Battery Cautions.4. Replace the battery pack with a fully charged lithium battery pack.5. Close the latch.Replacing a Sensor or Sensor Filtera Warning To avoid personal injury, only use sensors that are specifically designed for the detector. Refer to Specifications.a Caution Replace the sensor in a safe and non-hazardous area that is free of hazardous gas Each sensor has a high degree of resistance to common vapors and gases. To clear a sensor, place the detector in a clean environment and wait 10 to 30 minutes. Do not expose a sensor to vapors from organic solvents such as paint thinners and acetone. For a list of common products that can damage sensors, refer to Sensor Poisons and Contaminants.83

GasAlertMicro 5/PID/IR User ManualFor sensor problems, refer to Troubleshooting.

To replace a sensor or sensor filter, refer to Figure 15., Table 21., and the following procedures 1-7.

Figure 15. Replacing a Sensor or Sensor Filter NoteDetectors that are configured for 1, 2, 3, or 4 gases may contain a dummy sensor in one of the four sensor locations.84

Table 21. Replacing a Sensor or Sensor Filter

Item 1 2 3 45

Description Sensor cover Sensor filter Sensors Detector Machine screws (2)

1. Deactivate the detector.2. Remove the two machine screws on the rear shell and then remove the sensor cover or pump module.3. Remove the sensor filter and/or the sensor(s).4. Insert the new filter and/or sensor. Ensure the sensor posts are aligned correctly.5. Re-assemble the detector.6. If the sensor is changed, such as SO2 to an H2S, the detector must be reconfigured. Refer to Sensors in the Tech Mode option.7. A sensor that has been replaced must be calibrated immediately. Refer to Calibration and Setting Alarm Setpoints.

GasAlertMicro 5/PID/IR Maintenance

Photoionization Detector (PID)Clean or Replace the LampThe PID lamp must be cleaned regularly. Use only the cleaning kit that is supplied by BW Technologies by Honeywell.To clean the PID lamp, refer to the illustrations and procedures that are provided with the PID Lamp Cleaning Kit. To order the kit, refer to Replacement Parts and Accessories.

a CautionTo ensure proper maintenance and continued accurate readings from the sensor, use only the PID Lamp Cleaning Kit that is supplied by BW Technologies by Honeywell.

Table 22. Parts of the PID sensor

Item 1 2 3 45

Description PID sensor Diffusion barrier Sensor cover Electrode stack Lamp

85

GasAlertMicro 5/PID/IR User Manual

Replace the LampReplace the lamp when it falls below the acceptable level. Possible indicators that the lamp requires replacement are as follows:· The detector will not calibrate. · The startup self-test fails. · The ppm levels are incorrect. To replace the lamp, refer to the illustrations and procedures in the PID Lamp Cleaning Kit. If required, contact BW Technologies by Honeywell for more information.

Figure 16. Parts of the PID 86

Replace the Electrode StackReplace the electrode stack when it is contaminated. To replace the electrode stack, refer to Table 22., Figure 16., and the following procedures.a Caution Ensure your fingers do not make contact with the diffusion barrier or the electrodes on the underside of the stack.1. Remove the sensor cover. 2. Remove the old electrode stack. 3. Insert the new electrode stack. 4. Replace the sensor cover.WEEE Battery DirectiveFailure to comply with the following battery removal and disposal instructions may result in battery shorting, battery leakage, and/or other damage. Ensure a qualified technician completes the following procedures.Removal and Disposal of the Battery PackTo remove the alkaline batteries, refer to steps #1 to #3 in Replacing the Alkaline Batteries on page 82. To remove the lithium battery pack, refer to steps #1 and #2 in Replacing the Lithium Battery Pack on page 83. Dispose of the battery pack according to local laws.Removal and Disposal of the Coin Cell

GasAlertMicro 5/PID/IR Maintenance87

GasAlertMicro 5/PID/IR User Manual

Item 1 2 3 4 5 6 7 8 9 10

Description Diffusion cap Sensor filter Sensors Front shell Sensor board Back shell Machine screws (for diffusion cap) Battery pack Datalog card Machine screws (for rear shell)

The detector contains a coin cell to power the real-time clock. Only a qualified technician should complete the following procedure.1. Deactivate the detector.2. Open the latch on the bottom of the detector.3. Remove the battery pack by lifting the bottom of the pack upward from the detector.4. Remove the MMC card.5. Remove the two machine screws on the rear shell and then remove the diffusion cap or pump module.

88

6. Remove the sensors. 7. Remove the four machine screws in the battery pack cavity. 8. Remove the rear shell. 9. The coin cell sits on the sensor board.Gently remove the sensor board. 10. The coin cell is connected to the board by four leads. Clip thefour leads individually to remove the coin cell. a CautionDo not touch two or more leads while disconnecting the battery cell. 11. Dispose of the coin cell according to local laws.Note Image above shows three of the four battery leads that must be clipped.

GasAlertMicro 5/PID/IR Troubleshooting

TroubleshootingIf a problem occurs, refer to the solutions provided in Table 23. If the problem persists, contact BW Technologies by Honeywell.

Table 23. Troubleshooting

Problem

Possible Cause

Solution

Startup Troubleshooting The detector does not activate.The detector immediately enters alarm mode when activated.

No batteries Depleted batteries Damaged or defective detector Sensor needs to stabilize Low battery alarm Sensor alarm Pump alarm

The startup self-test fails.

General fault

Table 21. Troubleshooting

Refer to Replacing the Alkaline Batteries or Replacing the Lithium Battery Pack.Refer to Replacing the Alkaline Batteries, or Replacing the Lithium Battery Pack.Contact BW Technologies by Honeywell.Used sensor: wait 60 seconds. New sensor: wait 5 minutes.Refer to Battery Cautions, Replacing the Alkaline Batteries, Replacing the Lithium Battery Pack.Refer to Replacing a Sensor or Sensor Filter.If the sampling hose is attached, determine if it is obstructed. If not, clean or replace the pump filter. If the pump alarm persists, refer to the “Pump Operation” section in Troubleshooting.Ensure that the sensors and battery pack are installed correctly and then reactivate the detector. If the fault persists, record the error message and contact BW Technologies by Honeywell.

89

GasAlertMicro 5/PID/IR User Manual

Problem

Possible Cause

Detector Operation Troubleshooting

Detector does not display normal ambient gas readings after startup self-test.

Sensor not stabilizedDetector requires calibration Target gas is present

Detector does not respond to buttons.

Batteries are depleted

Detector is performing operations that do not require user input

Detector does not accurately measure gas.

Detector requires calibrationDetector is colder/hotter than ambient gas

Sensor filter is blocked

SolutionUsed sensor: wait 60 seconds New sensor: wait 5 minutes Calibrate the sensors. Refer to Calibration and Setting Alarm Setpoints. Detector is operating properly. Use caution in suspect areas. Refer to Replacing the Alkaline Batteries or Replacing the Lithium Battery Pack. Button function restores automatically when the operation ends.Calibrate the sensors. Refer to Calibration and Setting Alarm Setpoints. Allow the detector to adjust to ambient temperature before using.Clean the sensor filter. Refer to Replacing a Sensor or Sensor Filter.

90

Table 21. Troubleshooting

Problem

Possible Cause

MMC/SD Card Troubleshooting

The MMC/SD card is not inserted.

GasAlertMicro 5/PID/IR TroubleshootingSolutionInsert the MMC/SD card. Refer to Inserting the MMC/SD Card.

The MMC/SD card that is inserted in the detector has a storage size that is not supported by the detector.

Insert an approved MMC/SD card: · 128 MB Delkin MMC · 128 MB Delkin SD card · 128 MB Transcend SD card · 64 MB Unigen SD card

The detector has lost communication with the MMC/ SD card.

Retry communicationInsert a new approved MMC or SD card. Refer to Inserting the MMC/SD Card.Reformat the MMC or SD card in windows and then reinsert into the detector.Contact BW Technologies by Honeywell.

91

GasAlertMicro 5/PID/IR User Manual

Table 21. Troubleshooting

Problem

Possible Cause

Alarms TroubleshootingDetector does not enter alarm mode.

Alarm setpoint(s) are set incorrectly.

Alarm setpoint(s) are set to zero.Detector requires calibration.

Detector intermittently enters alarm without any apparent reason.

Ambient gas levels are near alarm setpoint or the sensor is exposed to a puff of the target gas.

Alarms setpoints are set incorrectly.

Missing or faulty sensor.

Detector enters alarm with LEL reading displaying as OL (over limit).

Pump flow rate is set too high.

Automatic Deactivation Troubleshooting

Detector automatically deactivates.

Automatic shutdown activated because of weak batteries.

Calibration is overdue and the Due-lock (calibration user lock-out) option is enabled.

92

SolutionReset alar

References

[xyz-ips snippet=”download-snippet”]


Posted

in

by