Mouthpiece ventilationQuick setup guide
Choose your circuit setup
Use the setup assistant to activate Astral’s default settings.When you choose a mouthpiece set up for an adult patient:• PEEP, back-up rate, and most alarms automatically default to OFF.• Other parameters adjust to common settings for MPV.The setup assistant will direct you to the circuit setup screen and prompt you to perform a Learn Circuit afterward.
Mouthpiece ventilation can be set up with valve circuits 1or mouthpiece circuit 2. The mouthpiece circuit is a simple, single circuit in which the mouthpiece can easily be replaced by a vented mask for night leak ventilation, without any need to change the circuit.
Choose your ventilation mode
When the Learn Circuit is complete, you will be directed to the ventilation modes.
The following modes can be set with the mouthpiece circuit:• (A)CV – (Assisted) volume-controlled ventilation• P(A)CV – (Assisted) pressure-controlled ventilation• PS – Pressure support mode
Titrate your patient
Initiating mouthpiece ventilation in (A)CV mode(based on recommendations in the ERS handbook for NIV)Set the tidal volume 1 at 700-1200 mL (1000 mL ) and adjust to optimise patient comfort.Set the PEEP to 0 cmH2O. PEEP setting is disabled
5 trigger options• Touch, for little to no effort: triggers at the touch of the mouthpiece• Standard, for higher effort: triggering sensitivity can be set at low, medium, or high• Off: for patients who are fully dependent on a backup rate
ResMed’s mouthpiece ventilation solution has been thoughtfully designed to reflect the needs of patients and their carers and support a good quality of life.
The EasySpeakTM mouthpiece is a convenient daytime solution for patients using Astral. It helps to make eating, breathing, and talking more natural. Quick to assemble and easy to clean, it features a curved ‘lip’ for a proper seal without the need to bite down. Choose between a mouthpiece that’s conveniently pre-mounted on a thin flexible tube or a stand-alone angled mouthpiece to offer a solution that meets your patient’s needs.
The circuit support arm holds the mouthpiece in a comfortable, accessible position. It provides up to 100cm of flexibility with 180 degrees of movement and four setup orientations suitable for single and double limb circuits.
1 Simonds AK, editor. ERS Practical Handbook of Noninvasive Ventilation. European Respiratory Society, 2015.Refer to the user guide for further instructions including indications, contraindications, warnings, and precautions. See ResMed.com for ResMed locations worldwide. Astral and EasySpeak are trademarks and/or registered trademarks of the ResMed family of companies. Specifications may change without notice. Product availability may vary across regions. © 2018 ResMed Ltd.10110946/1 2018-08ResMed.com/Astral