ROSCOE nebGO
ULTRASONIC HANDHELD NEBULIZERManual
MODEL: NEB-GO
Read this manual before operating the nebulizer.Save these instructions for future reference.
INTRODUCTION
Thank you for purchasing the Roscoe Medical branded nebGO Ultrasonic Handheld Nebulizer. With proper care, it will provide you with many years of reliable treatment.
This product is designed for the successful treatment of asthma, allergies and other respiratory disorders. Using vibrating mesh technology, the nebulizer converts the prescribed medication into an aerosol mist for easy inhalation.
Your ultrasonic handheld nebulizer should be used under the supervision of a licensed physician and/or a respiratory therapist. Any use of this product other than its intended use should always be avoided.
INDICATIONS FOR USE: This device is an ultrasonic vibrating mesh nebulizer system designed to aerosolize physician-prescribed liquid medications for inhalation to a patient (except for Pentamidine). The device may be used with patients 5 years and older in the home, hospital and sub-acute care settings.
ILLUSTRATION OF PARTS
- Medication Cup Cap
- Medication Cup
- Mouthpiece
- ON / OFF ButtonPress ON / OFF Button to turn on the power.
- ElectrodeMedication Cup Release Button
- Power Indicator*
Main Unit
- Press Release Button to remove the medication cup.
Light IndicationsGreen light on = During operationOrange light on / flash = Low power / MalfunctionNo lights = Malfunction
BOTTOM VIEW
- Battery Cover
- Power Cable Jack
- Power CableJack Cover
SAFETY SYMBOLS USED ON PRODUCT AND IN THIS MANUAL | |
![]() |
Indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury. |
![]() ![]() ![]() |
Indicates a potentially hazardous situation which, if not avoided, could result in serious injury and equipment damage. |
![]() ![]() ![]() ![]() |
Indicates a potentially hazardous situation which, if not avoided, could result in minor or moderate injury to the user or patient or damage to the device or other property. |
![]() ![]() |
Class II Equipment |
![]() ![]() |
Type BF Equipment |
![]() ![]() |
Consult Instructions for Use |
![]() ![]() |
On/Off Button |
![]() ![]() |
Serial Number |
IMPORTANT SAFETY PRECAUTIONS AND WARNINGS
![]() ![]() ![]() (USA only) |
U.S. Federal Law restricts this device to sale by, or on the order of, a physician or licensed practitioner. This device should only be used under the continued supervision of a physician or licensed practitioner. |
![]() ![]() ![]() |
THIS MEDICAL DEVICE IS NOT INTENDED TO BE USED AS A LIFE SUSTAINING OR LIFE SUPPORTING DEVICE. |
This device has been thoroughly tested and inspected to assure proper performance and operation.
Specifications put forth in this manual were in effect at the time of publication. However, to ensure continual improvement measures, changes to these specifications may be made at any time without obligation on the part of manufacturer.
Upon receiving your nebulizer, inspect the packaging for signs of damage in transit. Inspect the unit thoroughly. If damage is noted, external or internal, DO NOT use, and report to your Homecare Dealer immediately.
When using electrical products, especially when children are present, basic safety precautions should ALWAYS be followed. Read all instructions before using.
![]() ![]() |
To reduce the risk of serious or fatal injury from electrocution, fire, burns and to reduce the risk of damage or malfunction of the unit:
|
![]() ![]() |
|
![]() ![]() |
|
![]() ![]() |
|
![]() ![]() |
|
![]() ![]() |
|
![]() ![]() |
WARNING: ALWAYS USE CLOSE ADULT SUPERVISION WHEN ADMINISTERING TREATMENT TO A CHILD. DO NOT LEAVE CHILD UNATTENDED DURING TREATMENT.
|
![]() ![]() |
|
![]() ![]() ![]() ![]() |
Medication cup and accessories are NOT dishwasher safe. |
NOTE |
The exterior surfaces of the unit and the air outlet connector may become slightly warm to the touch during use. Warm air can also be felt exiting the vents on the sides of the case of the unit. This is normal. |
GETTING STARTED
- Check to ensure the following components of the nebulizer are enclosed:• Main Unit with Medication Cup• Carry Bag• Mouthpiece• Lanyard• Adult and Pediatric Masks• Power Cord with USB Wall Outlet
- It is recommended to keep extra mouthpieces and/or masks on hand at all times in case original parts become lost, damaged, or soiled.
![]() ![]() ![]() |
If you did not receive the components listed above, please contact the dealer where you purchased your nebulizer. |
SETUP AND OPERATING THE HANDHELD NEBULIZER
![]() ![]() ![]() ![]() |
Accessories are NOT sterile from package. Prior to initial operation, or if device has been stored/unused for >1 month, all accessories should be thoroughly cleaned. |
1. Wash hands with soap and warm water and dry completely with a clean towel.2. Make sure all accessories have been cleaned and/or disinfected before first use and after last daily treatment. Refer to the Cleaning of the Nebulizer Components and Accessories section on page 10.3. If using battery power, push back and pull up the battery cover to open.4. Insert two AA batteries according to the battery polarity indicated inside the battery compartment.5. Close and push forward the battery cover to lock.
![]() ![]() |
DO NOT use old and new batteries together or mix batteries with differing cell chemistries. |
6. If using the power adapter, make sure to remove batteries before use. The adapter does NOT charge batteries.
![]() ![]() ![]() |
The compressor is designed for intermittent use only. For optimal performance, operate the device for up to 20 minutes, then leave off for 20 minutes. |
7. To insert the medication cup into the device, align the three prongs on the back of the medication cup with the three trenches of the main device and press down. There will be an audible click.8. Flip open the medication cup cap.9. Fill nebulizer cup using eyedropper or syringe with prescribed medication. DO NOT fill more than 6 ml per treatment.10. Close the nebulizer cap, ensuring a tight seal.11. Press the mouthpiece or mask directly onto the medication cup opening.12. If using the mouthpiece, place it between your teeth, with lips firmly sealed around the mouthpiece. If using the mask, place it over the mouth, firmly sealed over the area around the mouth.13. Press the ON/OFF button on the top of the device to power the unit on. The light indicator will be green.14. Breathe in and out slowly and deeply through the mouth until aerosol treatment is complete.NOTE: Tilt the device slightly to ensure all residual medication contacts the mesh and is completely aerosolized.15. Press ON/OFF button to power the device off. The light indicator will be unlit.16. Unplug the power adapter from the electrical outlet (if using).17. Detach the mouthpiece or mask from the device.18. Remove the medication cup from the device by pressing the button under the cup and carefully lifting the cup up simultaneously.19. Clean and disinfect ALL nebulizer components before the next treatment. See pages 10-11.
![]() ![]() ![]() |
|
CLEANING
CLEANING THE NEBULIZER COMPONENTS AND ACCESSORIES
- If there is any residual moisture (condensation) in the medication cup, wipe it out with gauze or a lint-free towel.
- Pour distilled water into the medication cup.
- Turn on the device to nebulize the distilled water for 1 to 2 minutes to clean the mesh.
- Power off the device, then remove batteries or disconnect the adapter from the main unit.
- Remove the medication cup from the main unit.
- Wash and rinse the medication cup with distilled water.
- Shake off excess water and allow parts to be fully air dried on a clean, dry towel.
- Use a clean towel slightly dampened with warm water to clean the outside of the device.
- Wash the mouthpiece and mask accessories with a mild soap and warm water or soak in 1 part white vinegar to 3 parts hot tap water for one hour. Rinse thoroughly and allow to air dry completely before use. Perform disinfection per procedures on page 11.
- Make sure that all cleaned parts are completely dry before storing or used.
![]() ![]() |
|
![]() ![]() |
|
DISINFECTING THE NEBULIZER ACCESSORIES
Method 1 : Boiling
- Rinse the medication cup with distilled water.
- Bring a saucepan of distilled water to a boil.
- Carefully immerse the medication cup into the boiling water for a maximum of 10 minutes.
- Carefully remove the medication cup from the boiling water and shake off excess water.
- Allow to cool down and air dry completely on a clean, dry towel. Ensure all cleaned parts are completely dry before use or storage.
Method 2 : Alcohol
- Rinse the medication cup, mouthpiece and/or mask(s) with distilled water.
- Immerse the medication cup, mouthpiece and/or mask(s) in 75% ethyl alcohol for 1 minute.
- Rinse the medication cup, mouthpiece and/or mask(s) with distilled water again, shake off excess water.
- Allow parts to air dry completely on a clean, dry towel. Ensure all cleaned parts are completely dry before use or storage.
![]() ![]() |
|
MAINTENANCE
The following replacement parts and accessories are available:(DO NOT use unauthorized / non-Compass Health Brands replacement parts)
Description | SKU |
Medication Cup | X1GOSH01 |
Mouthpiece | X1GOSH02 |
Adult Mask | X1GOSH03 |
Pediatric Mask | X1GOSH04 |
Carry Bag | X1GOSH05 |
Power Cord | X1GOSH06 |
Wall Outlet | X1GOSH07 |
With proper care, the medication cup should be replaced every 12 months, and the mouthpiece and/or masks should be replaced every 3 months, depending on frequency of use.
STORAGE
When not in use, store the device and all accessories in the included carry bag or in a dry, cleanenvironment.
![]() ![]() |
|
![]() ![]() ![]() |
|
TROUBLESHOOTING
Please consult the table below to resolve any issues experienced with the device:
Problem | Possible Cause | Action |
Low Atomization | Low Battery Power | Replace batteries or use the power adapter |
Faulty Connection | Use rubbing alcohol to clean power adapter | |
Clogged Mesh Holes | Refer to Cleaning and Disinfection procedures to clean the medication cup | |
Broken Mesh | Replace the medication cup | |
No Power Indicator / No Atomization | Improperly Inserted Batteries | Follow (+) and (-) markings on the battery cover to reinsert in the correct direction |
Dead Batteries | Replace batteries or use the power adapter | |
Faulty Connection | Check and reconnect the power adapter to the main unit | |
Broken Mesh | Replace the medication cup | |
Power Indicator On / No Atomization | Improperly Inserted Medication Cup | Refer to setup instructions to reinstall the medication cup |
Clogged Mesh Holes | Refer to Cleaning and Disinfection procedures to clean the medication cup | |
Broken Mesh | Replace the medication cup | |
Orange Power Indicator | Low Battery Power | Replace batteries or use the power adapter |
Flashing Orange Power Indicator | Faulty Connection | Use rubbing alcohol to clean power adapter and/or reinstall the medication cup |
Broken Mesh | Replace the medication cup |
If the device does not function properly after performing the recommended actions above, contact an authorized Compass Health Brands dealer.
![]() ![]() |
NEVER operate this device if any of the parts are not functioning properly or have been damaged. |
NEBULIZER SPECIFICATIONS
Power | Two AA Alkaline Batteries; USB Power Adapter |
Power Input | AC100~240V/50~60Hz |
Power Consumption | <1.5W |
Medication Capacity | 6ml |
Vibrating Frequency | 117 kHz (+/- 15%) |
Particle Size (micron) | MMAD < 5μm |
Average Nebulization Rate | ≥0.25ml/min |
Operating Temperature/ Humidity Range | 50˚F to 104˚F (10˚C to 40˚C) / 30~75%RH |
Storage/Transport Temperature/ Humidity Range | -4˚F to 158˚F (-20˚C to 70˚C) / 20~80%RH |
Operating Cycles | 20 minutes on, 20 minutes off |
Dimension (L x W x H) | 3” x 1.6” x 2.9” (76.5 mm x 41 mm x 73 mm) |
Weight | 2.6 oz (74g) |
Standard Accessories | Main Unit, Medication Cup, Mouthpiece, Adult Mask, Pediatric Mask, Carry Bag with Lanyard and USB Power Adapter |
Battery Life | Up to 90 minutes continuous use; 4 days if used 20 minutes per day (batteries not included) |
Warranty | 2 Year Limited Warranty |
LIMITED WARRANTY
Compass Health Brands offers a limited two (2) year warranty on the device and three (3) months on accessories. We warrant that the compressor unit and its parts will be free from defects in workmanship and materials for a period of two (2) years from the date of original purchase and ninety (90) days on all accessories including medication cup. This warranty does not extend to failures resulting from accident, misuse, abuse, alteration, use of unauthorized service, parts or cleaning solutions, or failure to comply with the User’s Manual.
If the warranted equipment should fail during the warranty period, Compass Health Brands, at our option, will repair or replace parts at issue, provided the claim is bona fide. It shall be the responsibility of the purchaser to pack and return the equipment in a manner to avoid shipping damage. The shipment must be prepaid to us. This warranty covers the cost of labor incurred in the removal or replacement of the warranted component parts in the U.S. only.
SERVICE AND REPAIR
If you should have any questions about this nebulizer, please contact an authorized Compass Health Brands dealer. Do not open or disassemble the device at any time. This will void the warranty. If the unit is malfunctioning, discontinue use immediately and return to the original provider for repair or replacement.
DISPOSAL
This product should not be disposed of with other household wastes at the end of its working life. To prevent possible harm to the environment or human health from uncontrolled waste disposal, please separate this from other types of wastes and recycle it responsibly to promote the sustainable reuse of material resources.
Household users should contact their local government for details on appropriate and safe disposal.