Thank you for purchasing our product.Please read this manual carefully before use.Please keep this manual properly after reading
Safety Instructions
Please read the following precautions carefully
Caution There may be a risk of death or serious injury.
Caution There may be a risk of personal injury or property loss.
Do not input signals exceeds the measurement range of this product. Please select the correct test position and range to avoid damage to the instrument or personal injury. “ ” will be shown on the display when out of range.
When the voltage to be measured exceeds 36V DC or 25V AC, the operator shall be careful to avoid electric shock.
Check the function position before measuring.
Disconnect the test leads from the circuit before changing the mode.
For your safety, please read this manual carefully before use. Please fully understand the instructions and use this product correctly.
Caution Do not measure circuits that exceed the maximum input rated value 1000V.
Measurement Notices
Common Notice
Caution
DC/AC high voltage circuits are very dangerous, please becareful to measure.
Do not to add AC/DC voltages that exceeds the maximumrated value between the ground terminal and test terminal.
Do not add voltage that exceeds the allowable value.
Do not operate this product with wet hands to avoid the risk of electric shock.
Do not use the product around explosive gas, vapor, or in damp or wet environments.
Do NOT touch the input terminals when measuring.
Do not use test leads with damaged coatings.
AC/DC Voltage Measurement Notice
Caution
The measured voltage should not exceed the rated maximum test value, otherwise it may damage the product and endanger personal safety.
Do not measure voltages that exceed the allowable value.
AC/DC Current Measurement Notice
Caution
The measured current should not exceed the rated maximum test value, otherwise it may damage the product and endanger personal safety.
Use the 10A Terminal and the Mode to judge range and choose the right function position when measure an unknown current.
It is strictly forbidden to input voltage in this measuring state.
Resistance Measurement Notice
Caution
Before measuring the resistance, please make sure that all capacitors in the circuit to be measured are fully discharged.
It is strictly forbidden to input voltage in this measuring stat
Caution
Please do not measure the continuity with a voltage circuit or wire.
It is strictly forbidden to input voltage in this measuring state.
Other Notices
Battery Change Notice
Caution
Please turn off the power.
Install the battery cover and tighten the screws before measuring.
Test Leads Notice
Caution
Do not touch the pins of the test leads during the measurement. Do not use test leads with damaged coatings. Please do not stretch the cables to avoid breaking the test lead. Dust and water are not supported.
Caution
Except for our company or our authorized individuals, please do not attempt to repair or modify the circuit board to avoid the danger caused by improper operation.
Introduction
This product is a multi-functional, auto-ranging digital multimeter with a 6000 counts LCD display.
Specifications are subject to change without notice.
Batteries included are samples.
Battery life may be shorter than regular commercial batteries.If the backlight and buzzer are used frequently, battery life will be shortened.
Electrical Specifications
Function
Range
Resolution
Accuracy
Max
DC Voltage
(V)
6.000V
0.001V
±(0.5%+3)
1000V
60.00V
0.01V
600.0V
0.1V
1000V
1V
DC Voltage
(mV)
60.00mV
0.01mV
600mV
600.0mV
0.1mV
AC Voltage
(V)
6.000V
0.001V
±(1.0%+3)
750V
60.00V
0.01V
600.0V
0.1V
750V
1V
AC Voltage
(mV)
60.00mV
0.01mV
600mV
600.0mV
0.1mV
Function
Range
Resolution
Accuracy
Max
DC Current
(A)
6.000A
0.001A
±(1.2%+3)
10A
10.00A
0.01A
DC Current
(mA)
60.00mA
0.01mA
600mA
600.0mA
0.1mA
AC Current
(A)
6.000A
0.001A
±(1.5%+3)
10A
10.00A
0.01A
AC Current
(mA)
60.00mA
0.01mA
600mA
600.0mA
0.1mA
Resistance
600.0Ω
0.1Ω
±(0.5%+3)
60MΩ
6.000kΩ
0.001kΩ
60.00kΩ
0.01kΩ
600.0kΩ
0.1kΩ
6.000MΩ
0.001MΩ
60.00MΩ
0.01MΩ
±(1.5%+3)
Capacitance
9.999nF
0.001nF
±(5.0%+20)
9.999mF
99.99nF
0.01nF
±(2.0%+5)
999.9nF
0.1nF
9.999μF
0.001μF
99.99μF
0.01μF
999.9μF
0.1μF
9.999mF
0.001mF
±(5.0%+5)
Frequency
99.99Hz
0.01Hz
±(0.1%+2)
9.999MHz
999.9Hz
0.1Hz
9.999kHz
0.001kHz
99.99kHz
0.01kHz
999.9kHz
0.1kHz
9.999MHz
0.001MHz
Duty Cycle
1%~99%
0.1%
±(0.1%+2)
Temperature
(-20~1000)℃
1℃
±(2.5%+5)
1000℃
(-4~1832)℉
1℉
1832℉
Diode
√
Continuity
√
Preparations for Measurement
Preparations
Please check the product carefully before use and confirm if there is cracks or missing items. If you have any questions, please contact the store you purchased.
Disconnect the test leads from the circuit before changing the mode.
Please note that if there are noise-generating devices around, or large temperature difference, the data may become unstable or there may be increasing errors.
Please note that when measuring resistance, continuity, current flows through the measured circuit will result in incorrect measurement.
When using this device, it may cause display differences due to external strong noise, etc. If the measurement is not available, please turn off the power and wait for a while, then turn the power on again.
Note:The display changes irregularly under the situations where the test leads are not connected . This is caused by high input sensitivity, not a fault.When connected to the circuit, the correct measurement can be taken after the data become stable.
(a) Power (OFF)(b) AUTO (Voltage/Resistance/ Continuity) (c) Voltage/Frequency/ Duty Cycle (d) Temperature/mV Voltage(e) Diode/Resistance/Continuity/ Capacitance (f) Frequency/Duty Cycle (g) A Current(h) mA Current
Measurements
Steps for measurement
Confirm the measurement content in the table on the right.
Switch function dial according to the item to be measured.
Please remove test leads after measuring.
Turn off the power.
Caution
Do not confuse the item to be measured with the position of the function.
Do not exceed the maximum rated value of each function.
Do not switch functions or modes during measurement.
During the measurement process, hold the insulated part of the test lead and do not hold the pins of the test lead.
Auto Power Off
This feature prevents the battery from running out if you forget to turn off the power.
If the operation is not continued for about 15 minutes, the auto power off function will automatically cut off the power.
If you want to cancel this function, you should press and hold the SEL/NCV button and then turn it on again. It will be cancelled after five beeps.
※ Methods of Connecting Test Leads
Put the black lead into COM terminal and put the red lead to terminal in Picture1
Put the red lead to 10A terminal when test the current in Picture2.
Measurement
Function Position
Data Display
Connection Illustration
Usage and Notice
Auto
Auto-recognition for Voltage /Resistance/ Continuity by connecting test leads
Please rotate dial to AUTO position; Put probes correctly to recognize Voltage/Resistance/Continuity automatically. Only when the voltage is higher than 0.8V, this data will be shown on the display.※ Voltage/Resistance/Continuity can be also measured by switch dial to function position manually.
DC Voltage AC Voltage (Frequency Duty Cycle) Temperature
Please select temperature function in mV by SEL/NCV
Connect the black test lead to the COM Terminal and connect the red test lead to Terminal;
Rotate dial to Mode or Mode;
Press SEL/NCV to toggle between AC/DC;
Touch the correct test points of the circuit to measure the voltage;
Read the measured data on the display.※ If reverse the test lead, it shows “ – ” mark. The frequency is 40-1000Hz in AC Voltage Mode. There is no special link between the measured value and the test leads.
DC Current
Connect the black test lead to the COM Terminal and connect the red test lead to the Terminal or the 10A Terminal (based on the value of current);
Rotate dial to Mode or Mode;
Press SEL/NCV to toggle between AC/DC mode;
Cut off the circuit path to be measured. Then connect the test leads across the circuit and power supply;
Read the measured current on the display.※ Please measure current by 10A terminal within 15 sec When measuring DC current, please access to the circuit. from the red test lead to the black test lead based on the direction of the current flow
AC Current
ResistanceΩ
Press SEL/NCV to toggle between AC/DC mode
Please turn off the power of this component to be tested.
All capacitors are fully discharged.※ There is no special link between the measured value and the test leads
Continuity
The built-in beeper will beep when the resistance is lower than 50Ω, which indicates a short circuit.※ Please measure the resistance range in order to get detailed resistance.
Diode
Connect the red probe to the anode and the black probe to the cathode of the diode being tested;
Read the forward bias value on the display;
If the polarity of the test leads is reversed with diode polarity or the diode is broken and it shows “ – ” mark.a. Do not input voltage at the Diode Mode.b. Disconnect circuit power and discharge all capacitors before you test diode.
Capacitance
Connect the red test lead to the anode and the black test lead to the cathode of the capacitor to be tested;※ Disconnect circuit power and discharge all capacitors before testing capacitance. The electrostatic capacity becomes larger, the measurement time is longer.
Frequency Duty Cycle Hz/%
NCV Function
Data Hold/REL HOLD/REL
Maintenance
Replace Battery
When “ ” is shown on the display, batteries should be replaced as below:
Remove the test leads and turn off the product before replacing the batteries;
Loosen the screw on the battery door and remove the battery door.
Replace the used batteries with new batteries of the same type.
Place the battery door back and fasten the screws
Replace the Fuses
If current measurement is not possible, make sure that the fuse is not blown.If it is blown, please replace the required rated fuse.
Remove the battery cover and battery according to the battery replacement method;
Use a screwdriver to remove the two screws on the back cover.
Remove the back cover;
Replace the fuse;
Put the back cover and battery cover back, tighten the screws.
Clean the Product
Wipe the product with a damp cloth and mild detergent.Do not use abrasives or solvents. Dirt or moisture in the terminals can affect readings.*Remove the input signals before you clean the product.
Calibration
Calibration is regularly performed at the calibration laboratory to ensure accurate measurements.The recommended calibration period is once a year.Please consult your dealer for calibration cost and delivery time
Storage Method
Please turn off the power after use to avoid consuming the built-in battery.If it is not used for a long time, please remove the battery and keep it.
True RMS Voltmeter User Manual – True RMS Voltmeter User Manual –