All safety messages are identified by the following, “WARNING” or “CAUTION”, of ANSI Z535.4 (American National Standard Institute: Product Safety Signs and Labels). The meanings are as follows:
WARNING
A potentially hazardous situation that, if not avoided, could result in death or serious injury.
CAUTION
A potentially hazardous situation that, if not avoided, may result in minor or moderate injury.
This manual is subject to change without notice at any time to improve the product.
Product specifications are subject to change without any obligation on the part of the manufacturer.
When using the SK-WP, the following safety precautions should always be followed.
WARNING Use only the specified AC adapter for the scale. Other adapters may cause damage. (AC adapter is optional.) Internal service or adjustment to this product should be performed by a qualified person.CAUTION Avoid installing the scale in direct sunlight, which may cause discoloration or malfunctions.Do not mix battery types or new and old batteries. Replace with all new batteries at the same time.If the scale is not to be used for a long period of time, remove all batteries from the battery compartment to avoid leakage.Avoid overloading to the scale.Avoid using the weighing platform to move the scale, as that could cause damage to the scale.Avoid chemical solvents. Clean the scale with water.When the scale is used with an AC adapter, the scale does not conform to IP65 protection.
IP65
Degrees of protection against water: Protected against water jets.
degrees of protection against solid foreign objects: Dust-tight.
International Protection of IEC529.
PARTS DESCRIPTION
DISPLAY
PREPARATION
Installing Batteries / Connecting the AC AdapterInstalling Batteries
CAUTION When replacing batteries, be careful not to apply too much force to the weighing platform. An excessive force could damage the scale.
Unlock and remove the battery cover. (See illustration at the right) Insert six new batteries (R20P / LR20 / D size) into the battery compartment, taking extreme care to that the plus and minus polarities are observed. Be sure to install and lock the battery cover after installing the batteries.
AC Adapter (Optional)
CAUTION When using an AC adapter, the scale doesn’t comply with IP65.The AC input requirement could be 100, 120, 220, 230, or 240 volts (50/60Hz) depending on the area where used, so please verify that the adapter is the correct type.Plug the AC adapter into the AC adapter socket inside the battery compartment.
Setting up the Scale
CAUTIONAvoid installing the scale in direct sunlight, which may cause discoloration or malfunctions. Place your SK-WP on a firm weighing table so that the scale is level. The scale will not perform accurately when it is not level. Place the scale on a firm surface and adjust the feet to level the scale, while using the sprit level, for accurate weighing.
WEIGHING
Before Weighing
Automatic Power-Off Function.If the scale is left ON and the stable indicator is displayed, the automatic power-off function turns the power off after approximately five minutes (factory setting). To disable or enable the function, press the [ON/OFF] key while pressing the [RE-ZERO] key. The scale stores this setting. To verify this, press both the [RE-ZERO] and [UNITS] keys while in the weighing mode. (See “CALIBRATION” about the [UNITS] key.) poff0 Function disabled poff1 Function enabled
Error Message
E
Overload
—–
The scale zero is out of range
Lb
Low battery
Calibration May be Required Before WeighingRead “CALIBRATION” first and if necessary, calibrate your scale for accurate weighing.
Weighing Procedures
Turn on the scale.Press the [ON/OFF] key to turn on the scale. Then the scale displays all segments for a few seconds and then 0 will be displayed.
Start weighingIf you do not use a container for weighing,Press the [RE-ZERO] key to display (0)Verify the reading is (0 ). Place objects on the weighing platform to weigh.When the reading becomes stable, the stable indicator is displayed.If you use a container for weighing,Place an empty container on the platform. Wait for the stable indicator to be displayed and press the [RE-ZERO] key. Place the objects to be weighed in the container. When the reading becomes stable, the stable indicator is displayed.Select the weighing unit, — Only for the “(k)g-lb” model—If you use the (k)g-lb model, press the [UNITS] key to select weighing with either (k)g or lb as the unit.Normally scales power up in-lb when shipped from the factory. To power up in (k)g, turn the scale off. Press and hold the [UNITS] key, then press the [ON/OFF] key. The scale will start with (k)g. To return to the original setting, repeat this instruction.
CALIBRATION
When to Calibrate
Calibration may be required when the scale is initially installed if it is moved a substantial distance, or in accordance with local regulations. This is necessary because the weight of a mass in one location is not necessarily the same in another location. Also, with time and use, mechanical deviations can occur.
The Position of the [UNITS] key for the CalibrationThe (k)g model has no [UNITS] key, but the key for the calibration is located under the overlay
Calibration Using a Calibration Weigh
Calibrate the scale with the following calibration weight.SK-1000WP: 1000g ± 0.1gSK-2000WP: 2000g ± 0.2gSK-5000WP: 5000g ± 0.5gSK-5001WP: 5000g ± 0.5gSK-10KWP: 10kg ± 1gSK-20KWP: 20kg ± 2gSK-30KWP: 30kg ± 2g
Enter calibration mode.Press the [ON/OFF] key to turn the power off. While pressing both the [RE-ZERO] and [UNITS] keys, press the [ON/OFF] key. Then,[ Cal ]will be displayed.
Zero calibrationPress the [RE-ZERO] key to calibrate zero. [Cal0] will be displayed.When you have proceeded to the display other than[ Cal0], press the [ON/OFF] key to turn the power off and start over again.Wait for the stable indicator to be displayed and press the [RE-ZERO] key. The calf will be displayed in a few seconds. If the zero calibration only is to be done, press the [UNITS] key. The scale will show the[ end ]and automatically return to the weighing mode.
Span calibration Place the weight at the center of the platform for accurate calibration. Wait for the stable indicator to be displayed and press the [RE-ZERO] key. The scale will show the [end ]and the scale will automatically return to the weighing mode.
Calibration by Gravity CompensationIf you have calibration weights at the capacity of the scale, you can calibrate the scale with the weights instead of compensating for the gravity acceleration.
The scale is calibrated at 9.798 m/sec 2 at the factory. If the gravity acceleration at your location is not the same, calibrate the scale by compensating for the gravity acceleration. (See “The Value of Gravity at Various Locations” and “World Map” and find your location value.)
Enter calibration mode.Press the [ON/OFF] key to turn the power off. While pressing both the [RE-ZERO] and [UNITS] keys, press the [ON/OFF] key. Then, Cal will be displayed.
Select a digit to be changed.Press the [UNITS] key. The previously set value, for example, [9798] will be displayed. A digit after the comma “,” will be changed by pressing the [RE- ZERO] key. Press the [UNITS] key to move the comma “,” and select a digit to be changed.
Set the new value.Change the value of the gravity acceleration using the [RE-ZERO] and [UNITS] keys.
Save the value in memory.Press and hold the [UNITS] key until the end is displayed. The scale will automatically return to the weighing mode.
SPECIFICATIONS
Model
SK-1000WP
SK-2000WP
SK-5000WP
SK-5001WP
SK-10KWP
SK-20KWP
SK-30KWP
Capacity
1000 g
2000 g
5000 g
5000 g
10 kg
20 kg
30 kg
2.2 lb
4.4 lb
11 lb
—
22 lb
44 lb
66 lb
Resolution
0.5 g
1 g
2 g
1 g
0.005 kg
0.01 kg
0.02 kg
0.001 lb
0.002 lb
0.005 lb
—
0.01 lb
0.02 lb
0.05 lb
Non-linearity
±1g
±2g
±4g
±4g
±0.01 kg
±0.02 kg
±0.04 kg
Repeatability
0.5g
1g
2g
2g
0.005 kg
0.01 kg
0.02 kg
Span drift
±0.015% / °C TYP (5°C~35°C / 41°F~95°F)
Operating
-10°C~40°C / 14°F~104°F, Less than 85%RH
temperature
Display
25 mm / 0.98 inches, 7 segment liquid crystal display
Display
Approximately 10 times per second
update
Power
6 x R20P / LR20 / “D” size batteries or AC adapter
Battery life
Approximately 600 hours with manganese type
1200 hours with alkaline cells at 20°C / 68°F
Platform size
232 (W) x 192 (D) mm / 9.13 (W) x 7.56 (D) in.
Dimensions
266 (W) x 280 (D) x 146 (H) mm / 10.48 (W) x 11.03 (D) x 5.75 (H) in.