DJO AirSelect Standard User Guide

BEFORE USING THE DEVICE, PLEASE READ THE FOLLOWING INSTRUCTIONS COMPLETELY AND CAREFULLY. CORRECT APPLICATION IS VITAL TO THE PROPER FUNCTIONING OF THE DEVICE.

INTENDED USE / INDICATIONS:

The STANDARD walking brace provides compressed air support for stable fracture of foot and/or ankle,  severe ankle sprains, and post-operative use. Effective in fracture healing and the management of edema (swelling).

PERFORMANCE CHARACTERISTICS:

  • Soft-good/ semi-rigid designed to restrict motion through elastic or semi-rigid construction.
  • Providing immobilization or controlled movement of the limb or body segment.
  • Providing mild protection of a limb or body segment for acute and prophylactic care.

WARNINGS AND PRECAUTIONS:

  • Do not over-inflate air cells.
  • Improper Aircell inflation may cause significant skin irritation, especially in people with diabetic neuropathy (a diabetic nerve condition).
  • Check your skin frequently for “hot spots” and skin irritation.
  • If you see any skin changes or feelings of discomfort, reduce the Aircell compression.
  • Do not use this device if you cannot feel sensations whether due to post-op anesthesia, nerve, skin or other conditions.
  • Use caution when walking on slippery or wet surfaces to avoid injury.
  • If there is a feeling of excessive pressure or discomfort while using the device adjust the pressure of the air cells (if equipped)/straps on the device.
  • Do not use over open wounds.
  • If pain, swelling, changes in sensation or other unusual reactions occur while using this product, you should contact your doctor immediately.
  • Do not use the device if it is damaged.

NOTE: Contact manufacturer and the competent authority in case of a serious incident arising due to usage of this device.

APPLICATION INFORMATION:

A cotton or cast sock will enhance comfort – NOT INCLUDED1) PREPARE BRACE –

  • Unfasten straps, remove the front panel, and open the liner.

2) APPLY BRACE –

  • While seated, place foot in brace positioning the heel against the back of the brace. Wrap liner around foot and then leg (Fig. A).
  • Replace front panel: Position front panel so the Aircast logo is over the toes and the upper part of the panel is secured to the liner (Fig. B).
  • Secure straps from bottom to top (Fig. C). Tighten until snug and comfortable.

3) ADJUST AIRCELL COMPRESSION –The STANDARD walking brace air cells are initially deflated

  • Begin inflation on the injured side first. Turn the SELECTOR found on the right side of the brace to the appropriate number (Fig. D). Number 1 inflates the Aircell on the right side and Number 2 inflates the  Aircell on the left side. The “lock” position should be used once both air cells are inflated. This prevents accidental deflation.
  • To inflate Aircell, place finger over the small hole on top of large inflation pump found on the top right side of brace (Fig E). Pump until Aircell feels snug.
  • To inflate other Aircell, rotate SELECTOR to the appropriate number and then press and release a large inflation pump. Pump until Aircell feels snug. When both air cells are inflated, rotate SELECTOR to “lock” position.
  • In the event of over-inflation, deflate air cells.

4) DEFLATE AIRCELLS –

  • To deflate air cells, turn the SELECTOR found on the right side of the brace to the proper Aircell number (Fig. D).
  • Press and hold finger on small release button found on the top right-side of brace (Fig. F).
  • Deflate until pressure is comfortable.

NOTE: Deflate air cells before removing brace for proper fit when reapplying.NOTE: Higher pressure provides more support when walking, lower pressure is more comfortable when sitting or reclining. Pressure can also be adjusted by tightening or loosening straps.NOTE: The shell can be trimmed with a cast saw or cast scissors to accommodate larger calves.

CLEANING INSTRUCTIONS:

THE FOAM LINER MAY BE HAND-WASHED IN 86°F/30°C WATER WITH MILD DETERGENTAND AIR-DRIED. DO NOT PUT IN A DRYER OR USE ANOTHER HEAT SOURCE TO DRY.

HIGH ALTITUDE:

AT HIGH ALTITUDES THE AIRCELLS WILL EXPAND BEYOND THEIR OPTIMAL LEVEL. ADJUST AIRCELL PRESSURE BY TIGHTENING OR LOOSENING THE STRAPS.

WARRANTY:

DJO Global, LLC will repair or replace all or part of the unit and its accessories for material or workmanship defects for a period of six months from the date of sale.NOT MADE WITH NATURAL RUBBER LATEX.

MATERIAL COMPONENTS

Shell: POLYPROPYLENEFront Panel: HIGH-DENSITY POLYETHYLENEStraps: NYLONInflation Pumps: THERMOPLASTIC POLYURETHANELiner: NYLON/POLYURETHANE/POLYESTER (OUTER/FOAM/INNER MESH)Binding: POLYESTEROuter Sole / Insole: EVA/RUBBERSelector dial: NYLON

NOTICE: WHILE EVERY EFFORT HAS BEEN MADE IN STATE-OF-THE-ART TECHNIQUES TO OBTAIN THE MAXIMUM COMPATIBILITY OF FUNCTION, STRENGTH, DURABILITY, AND COMFORT, THERE IS NO GUARANTEE THAT INJURY WILL BE PREVENTED THROUGH THE USE OF THIS PRODUCT.

MDSS GmbHSchiffgraben 4130175 Hannover, Germany

DJO, LLC1430 Decision StreetVista, CA 92081-8553 • the USA