SNIPER STERILE STAPLE SYSTEM

The Sniper Sterile Staple System is a premium, precise, and consolidated single kit composed of a dynamic compression staple, disposable deployment device, and instrumentation.

STAPLE SYSTEMS

  • Fine trigger dispensing mechanism for accurate deployment without disruption to osteotomy/fracture site
  • Low profile, high-strength, super elastic nitinol compression staple implants designed with dual-barbs intended to mitigate migration without increasing difficulty of revision
  • A robust, stainless steel drill guide utilizes unique snap-off combo drill posts to allow for ease of use in staple placement
  • Simple intra-operative staple reloading capabilities intended to minimize implant waste
  • Available in 8x8x8mm and 10x10x10mm kit sizes
SNIPER STERILE STAPLE SYSTEM
Staple Size*

8.0 x 8.0mm

10.0 x 10.0mm

Wire Size

1.6 x 1.3mm

1.6 x 1.3mm

Part Number

500-08-001

500-10-001

*Bridge Length x Leg Length

Sterile Kit Contents

SURGICAL TECHNIQUE

STEP 1: Place a bone clamp to create the necessary compression across the osteotomy of the fusion site (when applicable).

STEP 2: Place drill guide in desired position across the osteotomy or fusion site.

STEP 3: Drill the first hole in the bone using the snap-off drill post provided. Toggle the proximal shaft of the snapoff drill post to remove the drill shaft and create a post.

STEP 4: Pivot position of the drill guide if desired.

STEP 5: Drill second hole using additional snap-off drill post provided.

STEP 6: Remove drill guide and post. Insert staple legs into pre-drilled holes until the distal tip of the deployment device abuts the osteotomy site.

STEP 7: Press trigger to deploy the staple.

STEP 8: Fully seat the staple.

STEP 9 (OPTIONAL): The distal section of the deployment device may be utilized as an impactor to fully seat the staple if deemed necessary by the surgeon.

STAPLE RELOADING (IF NECESSARY)

STEP 1: Insert staple legs into the applicable holes in the safety cap.

STEP 2: Press firmly until staple is fully seated and legs are in the parallel orientation.

STEP 3: Press trigger to open deployment device and align staple bridge over distal groove of the dispenser.

STEP 4: Release the trigger and push barrel forward to reload the staple.

STEP 5: Remove staple from the safety cap.

FDA cleared 510(k) K162354.Trilliant products are made in the U.S.A.

T 800.495.2919 F 877.778.3864727 North Shepherd Drive, Suite 100 | Houston, TX 77007 | U.S.A.djoglobal.com

Copyright © 2021 by DJO, LLC900-00-083 Rev G

Individual results may vary. DJO, LLC is a manufacturer of orthopedic implants and does not practice medicine. Only an orthopedic, or foot and ankle surgeon can determine what treatment is appropriate.The contents of this document do not constitute medical, legal, or any other type of professional advice.This material is intended for the sole use and benefit of the DJO, LLC sales force and physicians.It is not to be redistributed, duplicated, or disclosed without the express written consent of DJO, LLC.For more information on risks, warnings, and possible adverse side effects refer to the Instructions for Use provided with the device.