Luna ® G3 BPAP 25A (LG3700)Quick Operation Manual
Home Button: Return to the previous menu or main interface Knob: Adjust device settings Start / Stop Button: Start / Stop delivering air
Setting up the Device
①: Heated Tubing Port②: Air Outlet③: DC Inlet④: Filter Cap (Air Inlet)A. Connect one end of the tube to the assembled mask.B. Connect one end of the tube to the air outlet of the device, as shown in the figure above. Or connect the heated tubing joint to the air outlet of the device, and then insert the power plug into the heated tubing port on the back of the device.C. Connect the power adapter to the DC inlet of the main device.D. Connect the power cord to the power adapter.E. Plug the power cord into the power outlet.
Adjusting Device Settings
A. Connect the power cord and power adapter properly, The screen displays the Main The interface shown in the figure left below, press the Start / Stop Button, and the device will start delivering air, the screen displays the Main Interface shown the figure right below.
B. When the cursor is on the icon, press the Knob to access the Initial Setup Interface of the Patient Menu, as shown in the figure left below. The Heated Tubing option can only be adjusted when the device is connected to the heated tube, as shown in the figure right below.
C. Press the Knob to select the option which is going to be changed, as shown in the figure below. D. Turn the Knob to adjust an option. E. Press the Knob to confirm your adjustment to the option. F. The cursor switches among the options as you turn the Knob, as shown in the figure below. Refer to steps C-E to adjust other options. Press the Home Button to exit the Setup Interface.
Starting Treatment
After you finish adjusting device settings and return to the Main Interface, press the Start /Stop Button to start treatment.
Usage score of Luna ® G3 BPAP 25A
After each treatment, Luna ® G3 BPAP 25A will provide a Usage Score (for CPAP and AutoCPAP mode only) and Usage Summary on the screen for the user to get a quick understanding of the usage for the recent one day (from 12:00 noon of the previous day to 12:00 noon of the day).The total score is 100 points, with the below breakdown:
Total score |
Status |
Display |
≥ 75 | Good | |
55 – 74 | Normal | |
≤ 54 | Need to improve |
Used time (60 points)The more time of treatment the user received, the more points the user scores. If the device is used for 30 minutes, get 5 points, or for 1 hour, get 10 points. The user can score up to 60 points.
Used time (min) | ≥360 | 330-359 | 300-329 | … | 60-89 | 30-59 | 0-29 |
Score | 60 | 55 | 50 | … | 10 | 5 | 0 |
Mask seal (20 points)If the user is wearing the mask well without extra leakage during treatment, then the user can score up to 20 points. If the average leakage is less than 10 L/min, it scores 20 points, and 5 points will be deducted for every 5 L/min increase in leakage.
Leak (L/min) | <10 | 10-14 | 15-19 | 20-24 | ≥25 |
Score | 20 | 15 | 10 | 5 | 0 |
Respiratory events (10 points)The fewer the respiratory events (apneas and hypopneas), the higher the score.If AHI is less than 5, it scores 10 points, when AHI ≥ 5 and for every 5 increase in AHI, 2 points is deducted.
AHI | <5 | 5-9 | 10-14 | 15-19 | 20-24 | ≥25 |
Score | 10 | 8 | 6 | 4 | 2 | 0 |
[xyz-ips snippet=”download-snippet”]