Masimo LNC Series LNCS Patient Cables User Guide

LNCS® Patient CablesLNC Series

DIRECTIONS FOR USE

ReusableNot made with natural rubber latexNon-Sterile

INDICATIONS

The LNC Patient Cable series have the same indications for use as the applicable sensors. Please see the compatible sensor’s directions for use for indications for use and prescribed information.

DESCRIPTION

The LNC Patient Cable Series are used with Masimo LNCS® sensors. The LNC Patient Cable and Masimo LNCS Sensors have been validated with Masimo SET® Rad-5® Pulse Oximeter.WARNINGS, CAUTIONS, AND NOTES

  • Ensure the cable is physically intact, with no broken or frayed wires or damaged parts. Visually inspect the cable and discard if cracks or discoloration are found.
  • Carefully route patient cable to reduce the possibility of patient entanglement or strangulation.
  • Do not attempt to reprocess, recondition or recycle Masimo sensors or patient cables as these processes may damage the electrical components, potentially leading to patient harm.
  • Masimo sensors and cables are designed for use with devices containing Masimo SET® oximetry or licensed to use Masimo sensors.
  • All sensors and cables are designed for use with specific monitors. Verify the compatibility of the monitor, cable and sensor before use, otherwise degraded performance and/or patient injury can result.
  • Failure to properly connect the sensor or the oximeter module to the cable will result in intermittent readings, inaccurate results, or no reading.
  • To avoid damage to the cable, always hold it by the connector rather than the cable when connecting or disconnecting either end.
  • Always refer to the oximeter module operator’s manual for complete instructions or additional instructions.
  • To prevent damage, do not soak or immerse the sensor or cable in any liquid solution. Do not attempt to sterilize the cable or sensor.
  • Caution: Replace the cable when a replace cable message is displayed, or when a low SIQ message is consistently displayed while monitoring consecutive patients after completing the low SIQ troubleshooting steps identified in the monitoring device operator’s manual.
  • Note: The cable is provided with X-Cal® technology to minimize the risk of inaccurate readings and unanticipated loss of patient monitoring. The cable will provide up to 17,520 hours of patient monitoring time. Replace the cable when the patient monitoring time is exhausted. Some legacy monitoring devices do not distinguish between the cable and the sensor. In these devices, a replacement sensor or similar message could pertain either to the sensor or the cable.

INSTRUCTIONSA) Attach the LNC Patient Cable to the device

  1. Orient the connector of the cable to mate with the connector on the patient cable connector of the Masimo SET instrument and insert.B) Attaching the LNC Cable to the LNCS Sensor 1. Insert the sensor connector completely into the patient cable connector (Fig. 1).
  2. Completely close the protective cover (Fig. 2).

C) Disconnecting the LNC Cable from the LNCS Sensor

  1.  Lift the protective cover to gain access to the sensor connector (Fig. 3).
  2. Pull firmly on the sensor connector to remove from the patient cable (Fig. 4).

WARNINGDo not use the LNC-10 or LNC-14 Patient cables with the PXC2-30 Extension cable.

CLEANING

Clean the LNC Patient Cable by wiping it with a 70% isopropyl alcohol pad and allowing it to dry thoroughly.

COMPATIBILITY

LNC Patient Cables are for use only with devices containing Masimo SET® technology and with compatible sensors. Consult individual oximetry system manufacturers for compatibility of particular devices and sensor models. Each device manufacturer is responsible for determining whether their devices are compatible with each sensor and/or cable model. For Compatibility Information Reference: www.masimo.com

WARRANTY

Masimo warrants to the initial buyer only that this product when used in accordance with the directions provided with the Products by Masimo, will be free of defects in materials and workmanship for a period of six (6) months. Single-use products are warranted for single-patient use only.THE FOREGOING IS THE SOLE AND EXCLUSIVE WARRANTY APPLICABLE TO THE PRODUCTS SOLD BY MASIMO TO BUYERS. MASIMO EXPRESSLY  DISCLAIMS ALL OTHER ORAL, EXPRESS OR IMPLIED WARRANTIES, INCLUDING WITHOUT LIMITATION ANY WARRANTIES OF MERCHANTABILITY  OR FITNESS FOR A PARTICULAR PURPOSE. MASIMO’S SOLE OBLIGATION AND BUYER’S EXCLUSIVE REMEDY FOR BREACH OF  ANY WARRANTY SHALL BE, AT MASIMO’S OPTION, TO REPAIR OR REPLACE THE PRODUCT.

WARRANTY EXCLUSIONS

This warranty does not extend to any product that has been used in violation of the operating instructions supplied with the product, or has been subject to misuse, neglect,  accident or externally created damage.This warranty does not extend to any product that has been connected to any unintended device or system, has been modified, or has been disassembled or reassembled. This warranty does not extend to sensors or patient cables that have been reprocessed, reconditioned or recycled.IN NO EVENT SHALL MASIMO BE LIABLE TO THE BUYER OR ANY OTHER PERSON FOR ANY INCIDENTAL, INDIRECT, SPECIAL, OR CONSEQUENTIAL DAMAGES (INCLUDING WITHOUT LIMITATION LOST PROFITS), EVEN IF ADVISED OF THE POSSIBILITY THEREOF. IN NO EVENT SHALL MASIMO’S  LIABILITY ARISING FROM ANY PRODUCTS SOLD TO THE BUYER (UNDER A CONTRACT, WARRANTY, TORT OR OTHER CLAIM) EXCEED THE AMOUNT  PAID BY THE BUYER FOR THE LOT OF PRODUCT(S) INVOLVED IN SUCH CLAIM. IN NO EVENT SHALL MASIMO BE LIABLE FOR ANY DAMAGES  ASSOCIATED WITH A PRODUCT THAT HAS BEEN REPROCESSED, RECONDITIONED OR RECYCLED. THE LIMITATIONS IN THIS SECTION SHALL  NOT BE DEEMED TO PRECLUDE ANY LIABILITY THAT, UNDER APPLICABLE PRODUCTS LIABILITY LAW, CANNOT LEGALLY BE PRECLUDEDBY CONTRACT.

NO IMPLIED LICENSEPurchase or possession of this LNC Patient Cable does not carry any express or implied license to use this cable with any device that is not an authorized device or separately authorized to use LNC Patient Cables.CAUTION: FEDERAL LAW (U.S.A.) RESTRICTS THIS DEVICE TO SALE BY OR ON THE ORDER OF A PHYSICIAN.For professional use. See instructions for use for full prescribing information, including indications, contraindications, warnings, precautions, and adverse events. If you encounter any serious incident with the product, please notify the competent authority in your country and the manufacturer.

The following symbols may appear on the product or product labeling:

http://www.Masimo.com/TechDocs

Patents: http://www.masimo.com/patents.htmMasimo, SET, , LNCS, X-Cal, and Rad-5 are federally registered trademarks of Masimo Corporation.

© 2021 Masimo Corporation

Manufacturer:Masimo Corporation52 DiscoveryIrvine, CA 92618USAwww.masimo.com

Authorized representative for Masimo Corporation:MDSS GmbHSchiffgraben 41D-30175 Hannover, Germany3490L-eIFU-0321