maxtec MaxBlend 2 Instruction Manual

MaxBlend™2LOW FLOW / HIGH FLOW

Instructions for UseR229M01 REV. HMaxtec2305 South 1070 WestSalt Lake City, Utah 84119USAphone: (800) 748.5355fax: (801) 973.6090email: [email protected]web: www.maxtec.comAUTHORIZED REPRESENTATIVE:QNET BVKantstraat 19, NL-5076 NP HaarenThe NetherlandsConforms to:AAMI STD ES60601-1, ISO STD80601-2-55, IEC STDS 606011-6, 60601-1-8 & 62366Certified to:CSA STD C22.2 No. 60601-1

NOTE: The latest edition of this operating manual can be downloaded from our website at www.maxtec.comRead this entire manual before attempting to operate or service the MaxBlend 2. Attempting to operate the MaxBlend 2 without fully understanding its features and functions may result in unsafe operating conditions.

CLASSIFICATION

Protection class ……………………………………………………………………………………………………… II, Type BProtection against water ………………………………………………………………………………………………… IPX1Mode of operation ………………………………………………………………………………………………. ContinuousSafety of application in the presence of a flammable anesthetic mixture ……….. See section 9.4Power specification ……………………………………………………. 7.5V(MAX) 1.9W/250mA(MAX)CAUTION: Federal law restricts this device to sale by, or on the order of, a medical professional.Product Disposal Instructions:The sensor, batteries, and circuit board are not suitable for regular trash disposal.Return sensor to Maxtec for proper disposal or disposal according to local guidelines. Follow local guidelines for disposal of other components.

WARRANTY

The MaxBlend 2 is designed for air/oxygen delivery. Under normal operating conditions, Maxtec warrants the MaxBlend 2 to be free from defects of workmanship or materials for a period of three years from the date of receipt from Maxtec, provided that the unit is prop­erly operated and maintained in accordance with Maxtec’s operating instructions. Based on Maxtec’s product evaluation, Maxtec’s sole obligation under the foregoing warranty is limited to making repairs or issuing credit, or replacement for equipment found to be defective. This warranty extends only to buyers purchasing new equipment directly from Maxtec, or through Maxtec’s designated distributors and agents. Maxtec warrants the MAX-550E oxygen sensor in the MaxBlend 2 to be free from defects in material and workmanship for a period of two years from Maxtec’s date of shipment in a MaxBlend 2 unit. Should a sensor fail prematurely, the replacement sensor is warranted for the remainder of the original sensor warranty period. Routine maintenance items, such as batteries, are excluded from the warranty. Maxtec, and any other subsidiaries, shall not be liable to the purchaser or other persons for incidental or con­sequential damages to equipment that has been subject to abuse, misuse, misapplication, alteration, negligence, or accident.THESE WARRANTIES ARE EXCLUSIVE AND IN LIEU OF ALL OTHER WARRANTIES, EXPRESSED OR IMPLIED, INCLUDING A WARRANTY OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE.NOTE: In order to obtain optimum performance from your MaxBlend 2, all operations and maintenance must be performed in accordance with this manual. Please read the manual thoroughly before using the MaxBlend 2, and do not attempt any repair or procedure that is not described herein. Maxtec cannot warranty any damage resulting from misuse, unauthor­ized repair, or improper maintenance of the instrument.EMC NoticeThis equipment uses, generates, and can radiate radio frequency energy. If not installed and used in accordance with the instructions in this manual, electromagnetic interference may result. The equipment has been tested and found to comply with the limits set forth in IEC 60601-1-2 for medical products. These limits provide reasonable protection against electromagnetic interference when operated in the intended use environments described in this manual.MRI NoticeThis equipment contains electronic and ferrous components, whose operation can be affected by intense electromagnetic fields. Do not operate the MaxBlend 2 in an MRI environment or in the vicinity of high-frequency surgical diathermy equipment, defibrillators, or shortwave therapy equipment.Electromagnetic interference could disrupt the operation of the MaxBlend 2.WARNINGSIndicates a potentially hazardous situation that, if not avoided, could result in death or serious injury.The following warnings apply any time you operate or service the MaxBlend 2:

  • Always confirm prescribed flow before administering to patient and monitor flow on a frequent basis.
  • Always follow ANSI and CGA standards for medical gas products, flowmeters, and oxygen handling.DO NOT operate the MaxBlend 2 unless qualified personnel is in attendance to promptly respond to alarms, inoperative conditions, or sudden malfunctions.
  • Patients on life-support equipment should be visually monitored at all times.DO NOT ignore audible alarms of the MaxBlend 2. Alarms indicate conditions that require your immediate attention.DO NOT use parts, accessories, or options that have not been authorized for use with the MaxBlend 2. Using unauthorized parts, accessories, or options may be harmful to the patient or damage the MaxBlend 2.
  • Check all audible and visual alarms periodically to ensure they are operating properly. If an alarm fails to activate, contact your Maxtec Certified Service Technician.DO NOT operate the MaxBlend 2 with the monitor powered off or without setting the adjustable alarms. All adjustable alarms must be set to ensure safe operation.DO NOT steam autoclave or otherwise subject the MaxBlend 2 to temperatures above 122°F (50°C).
  • If either the air or oxygen source pressure is reduced or increased sufficiently to create a pressure differential of 20 PSI or more, an audible alarm will sound. This condition may significantly alter the FiO2 and flow output from the MaxBlend 2.DO NOT tape, obstruct, or remove the alarm during clinical use.DO NOT occlude the sensor port on the side of the MaxBlend 2.
  • An air inlet/water filter is recommended for use with the MaxBlend 2. See section 6.1.DO NOT use lubricants on the MaxBlend 2.DO NOT use the MaxBlend 2 until the correct performance has been verified. See section 3.0.
  • If a condition is detected that could possibly prevent the monitor from continuing to operate safely, it will sound an alarm. If, at any time, EOx (i.e. E02, E04, etc.) appears on the LCD refer to section 4.0, or contact a Maxtec Certified Service Technician.
  • All services should be referred to a Maxtec Certified Service Technician.
  • Maxtec recommends that the MaxBlend 2 be serviced by Maxtec at a minimum of every three years, or if a leak or other performance problem is suspected.
  • If the MaxBlend 2 is dropped, follow the procedures outlined in section 3.0 for a performance check before reusing the device.
  • When the unit is going to be stored (not in use for more than 30 days), always remove the batteries to protect the unit from potential damage due to battery leakage.
  • Always replace batteries with high-quality, name-brand “AA” alkaline batteries.
  • The low oxygen alarm can be set below 18%, and as low as 15%. See section 2.3 on setting alarms. This is in accordance with IEC 80601-2-55.
  • In the event of exposure to an ELECTROMAGNETIC DISTURBANCE, the analyzer may display an E06 or E02 error message. If this occurs, refer to Section 4.0 for instructions to resolve the problem.To prevent the risk of burns, fire, or injury to person(s):
  • Mixed gases continuously bleed into the atmosphere at the oxygen concentration rate set at the control knob. Bleeding oxygen into any closed area could increase the risk of fire or explosion.DO NOT operate this device in the presence of any flame or source of ignition, or when using devices such as electrosurgical equipment or defibrillators.DO NOT operate the MaxBlend 2 in the presence of flammable anesthetics or in an atmosphere of explosive gases. Operating the MaxBlend 2 inflammable or explosive atmospheres may result in fire or explosion.
  • Galvanic 02 sensor electrolyte gel is acidic and may cause skin or eye irritation and/or burns. Take care when handling or replacing exhausted or damaged disposable 02 sensors. Be sure to dispose of expired sensors in accordance with hospital and/or governmental regulations (02 Sensor SDS available upon request from Maxtec).DO NOT smoke in an area where oxygen is being used.CAUTIONSIndicates a potentially hazardous situation that, if not avoided, could result in minor or moderate injury and property damage.DO NOT store the MaxBlend 2 in hot areas for prolonged periods of time. Temperatures above 80°F (27°C) can shorten battery life.To minimize the potential for electrostatic shock, DO NOT use antistatic or electrically conductive hoses with the MaxBlend 2.When cleaning the MaxBlend 2:DO NOT use harsh abrasives.DO NOT immerse the MaxBlend 2 in liquid sterilizing agents or liquids of any kind.DO NOT spray cleaning solution directly onto the sensor port, bleed muffler, or buzzer opening.DO NOT allow cleaning solution to pool on the front panel, sensor port, or bleed muffler.DO NOT sterilize the MaxBlend 2. Standard sterilization techniques may damage the blender.DO NOT clean or dry the MaxBlend 2 with a high-pressure air gun. Applying high-pressure air to the MaxBlend 2 may damage components and render the system inoperable.DO NOT over-clean the MaxBlend 2. Repeated use of a cleaning agent can cause residue build-up on critical components. Excessive residue buildup can affect the MaxBlend 2’s performance.
  • Dropping or severely jarring the sensor after calibration may shift the calibration point enough to require recalibration.
  • Always operate the MaxBlend 2 with clean, dry medical-grade gases. Contaminants or moisture can cause defective operations. Oxygen should have a minimum dewpoint of -80°F (-62°C) or moisture content less than 7.9 ppm (0.0059mg/L). Oxygen “purity”  should be at least 99.0% and the air used should be medical-grade. Water vapor content must not exceed a dew point of 5°F (-15°C) below the lowest ambient temperature to which the delivery system is exposed.
  • Particulate content must not exceed that which would be found immediately downstream of a 15-micron absolute filter. Refer to CGA commodity specifications G-4.3 and G7.1 for more information. The water vapor content of medical air or 02 supply to the blender must not exceed 5.63 x 103 milligrams H2O per cubic meter of non-condensable gas.DO NOT disassemble the MaxBlend 2. All service should be performed by a Maxtec Certified Service Technician.
  • Be sure the MaxBlend 2 is securely mounted. This device is usually mounted to a hospital rail system or an infusion stand. Dropping the device may cause injury or damage to the device.
  • The outlets of this device are capable of delivering pressure as high as the inlet pressure.Ensure that the devices that carry the gas from the blender to the patient prevent excessive pressure on the patient.DO NOT use the oxygen sensor if it appears to be damaged or is leaking. The oxygen sensors contain a weak acidic solution encapsulated in plastic housing. Under normal operating conditions the solution (electrolyte) is never exposed.NOTESIndicates supplemental information to assist in the use of the device.
  • The MaxBlend 2 is tested for compliance with IS011195 and meets requirements regarding reverse gas flow as delivered.
  • Applicable parts used in the MaxBlend 2 have been cleaned and degreased for oxygen service. Any lubricants used are designed specifically for the application.
  • As long as the absolute pressure of the gas mixture being monitored is constant, the MaxBlend 2 will accurately read oxygen concentrations. However, if the absolute pressure varies the reading will fluctuate proportionately, as the sensor actually measures the partial pressure of oxygen in the mixture. The sensor readings will also change proportionately with barometric pressure changes; because of this, frequent calibration of the sensor is recommended.
  • Users are advised to use pressure regulators which display the outlet pressure.
  • All specifications assume the following standard environmental conditions unless specified otherwise: ambient and sample gas temperatures of 77°F (25°C), barometric pressure of 30inHg (760mmHg), sea level altitude, ambient relative humidity of 50%, sample gas relative humidity of 0%.
  • It is important to note that the oxygen concentration selection scale is provided only as a guideline for selecting 02 concentrations. The clinician should use the MaxBlend 2’s display panel to adjust the 02 concentration to the desired setting.
  • The alarm limits can be set to levels that would render them useless for a particular patient’s clinical condition. Ensure that the delivered oxygen level and flow rate are set to values prescribed by the patient’s physician. Also, ensure that the high and low alarm limits are set to levels such that they will sound if the oxygen level is outside of safe limits.Be sure to review and, if necessary, re-set the alarm limits when the patient’s clinical condition changes, or when the patient’s physician prescribes a change in oxygen therapy.
  • This device does not contain automatic barometric pressure compensation.
  • Gas leaks that cause room air to mix with the gas sample may cause inaccurate oxygen readings. Ensure the 0-rings on the sensor and flow diverter are in place and intact prior to use.

INTRODUCTION

The MaxBlend 2 is a compact air/oxygen gas mixing device that incorporates the use of a battery-powered oxygen monitor. The gas mixing device (blender) provides precise mixing of medical-grade air and oxygen, while the monitor measures the selected oxygen concentra­tions from the blender’s gas flow and displays these measured concentrations on a digital display. The monitor provides high and low alarm limits which, when exceeded, cause an audible and visual alarm.1.1 Indication for UseThe MaxBlend 2 is designed to provide a continuous air/oxygen gas mixture and to continu­ously monitor the concentration of oxygen being delivered to the infant, pediatric, and adult patients. It is a restricted medical device intended for use by qualified, trained personnel, under the direction of a physician, in professional healthcare settings, i.e., hospital, subacute, and nursing-care facilities where the delivery and monitoring of air/oxygen mixtures is required. This is not intended as a life-supporting device.1.2 MAX-550E Oxygen SensorThe MAX-550E is a galvanic, partial pressure sensor that is specific to oxygen. It consists of two electrodes (a cathode and an anode), a Teflon membrane, and an electrolyte. Oxygen diffuses through the Teflon membrane and immediately reacts electrochemically at a gold cathode. Concurrently, oxidation occurs electrochemically at a lead anode, generating an electrical cur­rent and providing a voltage output. Electrodes are immersed in a unique gelled weak acid electrolyte, which is responsible for the sensor’s long life and motion insensitive characteristic. Since the sensor is specific to oxygen, the current generated is proportional to the amount of oxygen present in the sample gas. When no oxygen is present, there is no electrochemical reaction, and therefore, the negligible current is produced. In this sense, the sensor is self-zeroing.CAUTION: The MAX-550E oxygen sensor is a sealed device containing a mild acid electro­lyte and lead (Pb). These materials are hazardous waste constituents and should be disposed of properly, or returned to Maxtec for proper disposal or recovery.CAUTION: Dropping or severely jarring the sensor after calibration may shift the calibra­tion point enough to require recalibration.1.3 Symbol GuideThe following symbols and safety labels are found on the MaxBlend 2 and/or labeling:

POWER SUPPLY SYMBOL GUIDEThe following symbols and safety labels are found on the MaxBlend 2 power supply (sold separately):

1.4 Component Identification

  1. LOW ALARM LED — In a low alarm condition, the “LOW ALARM” LED will flash twice a second, accompanied by the audio buzzer.
  2. HIGH ALARM LED — In a high alarm condition, the “HIGH ALARM” LED will flash twice a second, accompanied by the audio buzzer.
  3. ON/OFF KEY —This key is used to turn the device on or off. To turn the device OFF, the button must be held while a rapid 3-2-1 countdown takes place to prevent accidental power-off.
  4. UNLOCK KEY — The unlock key is used to lock and unlock the instrument.
  5. UP (ALARM HIGH) KEY — The up key is used in setting the high FiO2 alarm limit. The device must be in the unlocked state for the key to operate. See section 2.3.2 for instructions on setting the high FiO2 alarm limit.
  6. CALIBRATION KEY — This key is used to calibrate the device. The device must be in the unlocked state for the key to operate. See section 2.8 for instructions on calibrating.
  7. O2 INLET FITTING — An O2 fitting for connection to the O2 inlet hose from the gas source.
  8. OXYGEN CONCENTRATION SELECTOR CONTROL — A knob that allows for selections of mixed oxygen concentrations from 21% to 100%.
  9. MEDICAL AIR INLET CONNECTOR — An air fitting for connection to an air inlet hose from the air gas source.
  10. FLOWMETER OUTLET — A fitting for connection to the patient delivery tubing.
  11. BLEED TOGGLE SWITCH — The Bleed Toggle Switch turns on the auxiliary bleed. For the blender to accurately mix the oxygen and air, the bleed must be turned on whenever the total flow delivered to the patient is less than 15 LPM for the high flow model or less than 3 LPM for the low flow model.
  12. FLOW CONTROL KNOB — The flow control knob controls the flow of gas exiting the flowmeter.
  13. OXYGEN FLOWMETER — Measures the flow of mixed gas exiting the flowmeter outlet.
  14. SMART ALARM KEY — The Smart Alarm key is used to help set the High-Low Alarm window quickly.See section 2.3.3 for instructions on using the Smart Alarm setting.
  15. DOWN (LOW ALARM) KEY —The down key is used in setting the low FiO2 alarm limit. The device must be in the unlocked state for the key to operate. See section 2.3.1 for instructions on setting the low FiO2 alarm limit.
  16. BACKLIGHT KEY — The backlight key will manually activate the backlight for 30 seconds.See section 2.6 for more information on backlighting operation.
  17. SILENT KEY — In an alarm condition, pressing the SILENT key will deactivate the audio alarm for 2 minutes.
  18. LCD DISPLAY — The digit liquid crystal display (LCD) provides a direct readout of oxygen concentrations. The digits also display error codes, alarm set modes, and calibration codes as necessary.1.5 Display Indicators
  19. OXYGEN CONCENTRATION — Current oxygen concentration percentage from the oxygen sensor.
  20. HIGH ALARM INDICATOR —The high alarm setting is displayed at all times just below the “HIGH” icon on the LCD readout. The indicated value represents the oxygen percentage at which the high alarm will be activated.
  21. OXYGEN HIGH ALARM LIMIT — High oxygen alarm setpoint. Audible and visual alarms will trigger when this limit is exceeded.
  22. CALIBRATION REMINDER — The calibration reminder symbol is located at the bottom of the display. This symbol will display after one week has elapsed from the previous calibration.
  23. SLEEP MODE INDICATOR — The sleep mode indicator is used to help with battery consumption.
  24. LOW BATTERY INDICATOR —The Low Battery Indicator is located in the middle of the display and is only activated when the voltage on the batteries is below a normal operating level.
  25. ALARM SILENCE/SMART ALARM INDICATOR — When the Silent key is pressed the indicator will display with crossbars to alert the condition. When the Smart Alarm key is pressed the indicator will display with T-bars to alert the condition.
  26. OXYGEN LOW ALARM LIMIT — Low oxygen alarm setpoint. Audible and visual alarms will trigger when this limit is exceeded.
  27. LOW ALARM INDICATOR —The low alarm setting is displayed at all times just below the “LOW” icon on the LCD readout. The indicated value represents the oxygen percentage at which the low alarm will be activated.
  28. <18% ALARM INDICATOR — The <18% alarm indicator is located above the Low Alarm Indicator digits. When the low alarm setting is set below <18%, the indicator will flash each second to alert the operator of this special condition. See section 2.3.1 for setting this low alarm condition.

1.6 Back View33. BATTERY COMPARTMENT — The MaxBlend 2 is powered by four “AA” alkaline batteries.34. SENSOR CABLE — The cable connects the MaxBlend 2 to the MAX-550E Sensor.35. SENSOR WITH DIVERTER — The sensor with flow diverter is designed to fit into a port behind the flowmeter.36. O2 SENSOR PORT — A sampling port for the oxygen sensor. It allows mixed gas from the blender to flow over the sensor membrane.37. PRESSURE DIFFERENTIAL REED ALARM — An audible alarm that, when activated, indicates that an unacceptable pressure differential exists between the two gas source pressures.38. AUXILIARY MIXED GAS OUTLET — The outlet can be used as a power take-off.39. AUXILIARY MIXED GAS OUTLET — The outlet can be used to add additional flowmeters.40. MOUNT ADAPTER — An adapter that allows the MaxBlend 2 to be mounted onto a bracket for attaching to a rail or ventilator system.41. EXTERNAL POWER SUPPLY PORT — The port provides a connection for the external power adapter. See section 2.7 for more information on the power supply.1.7 Requirements for Operating the BlenderAll operator-detachable inlet pressure hoses supplied with the gas mixer comply with ASTM/ISO 5359.PRESSURIZED OXYGEN: The compressed oxygen source must provide clean, dry, medical-grade oxygen at the pressure specified in Section 8.0.PRESSURIZED AIR: The compressed air source must provide clean, dry, medical-grade air at the pressure specified in Section 8.0.

OPERATING PROCEDURES

2.1 Setup and Installation2.1.1 Battery InstallationAll MaxBlend 2 units are powered by four “AA” alkaline batteries (4 x 1.5 Volts) and are shipped without the batteries installed. Only name-brand batteries should be used. Batteries should be replaced by trained service personnel.To install the batteries:Open the battery drawer by squeezing inward on both tabs as shown in the figure below. If you have difficulty squeezing the tabs in with your fingers, use two flat screwdrivers or two coins. Remove the battery drawer completely from the MaxBlend 2. Install four new, “AA” alkaline batteries into the unit, observing the orientation shown on the plastic inside the drawer. Slide the drawer back in with the batteries facing upward. Press in on the drawer until both tabs latch into place.When batteries are replaced, the unit will start up and perform automatic calibration.Ensure calibration is performed as directed in Section 2.8.WARNING: Battery replacement by inadequately trained personnel could result in a safety hazard.WARNING: Electrical shock or damage to the equipment may occur if an inappropriate external power supply is used.Maxtec recommends using only the Maxtec MaxBlend 2 External Power Supply—R230P10.

2.1.2 MaxBlend 2 Setup

  1. Connect the pressurized air source to the Air Inlet Connector.
  2. Connect the pressurized oxygen source to the O2 Inlet Fitting.
  3. Flush gas at the highest possible flow rate through the blender for at least one minute to eliminate any particulate that may have been introduced into the system during handling and installation.

2.1.3 Sensor Installation

  1. Attach the flow diverter onto the oxygen sensor.
  2. Place the sensor into the sensor port located behind the flowmeter.
  3. Attach the sensor cable directly to the sensor and the sensor jack on the back of the monitor enclosure. Ensure the cable is fully inserted into both connections.
  4. Calibrate the sensor prior to use according to the calibration procedures in section 2.8

2.2 MonitoringBefore use on a patient, the oxygen concentration of the delivered gas should be checked at the setting intended for use.

  1. To begin monitoring, press the ON/OFF key located on the front panel. Monitoring will begin immediately.
  2. Should oxygen level exceed either the HIGH or LOW alarm set points, the alarm indicator on the front panel will illuminate indicating either a high or low oxygen condition. The alarm will cease when oxygen levels are within limits, or the limits are adjusted.

NOTE: The MaxBlend 2 will monitor the oxygen concentration of the gas delivered from the blender regardless of which outlet port is in use. Therefore, the integrated flowmeter has ultimately noeffect on the oxygen sensing.2.3 Alarm Setting Procedure2.3.1 Low Alarm SettingTo adjust the low alarm setting:

  1. Press the Unlock key to unlock the keypad. Note the SMART ALARM, (AL. and HIGH icons will begin to flash indicating the SET OPERATING MODE.
  2. Press the DOWN (Low Alarm) key on the keypadNOTE: The Low Alarm digits begin to flash indicating the Low Alarm manual setting.
  3. Use the UP and DOWN keys to set the low alarm to the desired value. Pressing the arrow keys changes the value in 1% increments. If the keys are held down for more than I second. the display will scroll at a rate of I% per second.

NOTE: If 30 seconds elapse between key actuations, the system will store the latest low alarm setting and will revert to normal operation. If this occurs inadvertently, simply repeat the alarm setting procedure.There is a special condition that allows the low oxygen alarm to be set below 18%. To access this condition. Press the DOWN arrow key for One second while the law alarm reading dis­plays 18%. The alarm setting can now be adjusted to 11.16. or I5%. A bar will blink above the setting to provide further indication that the alarm has been set to this special <18% condition.The low alarm value cannot be set lower than 15%, nor can it be set closer than 1% from the high alarm value, For example, if the high alarm is set at 25%, the system will not accept a low alarm setting greater than 24%.When the low alarm value is set. press the Unlock key again to accept the low alarm setting and return to normal operation.NOTE: The default low alarm setting is 18% 02. Removing the batteries, or shutting the unit OFF will reset the low alarm limit to 18%.2.3.2 High Alarm SettingTo adjust the high alarm setting:

  1. Press the Unlock key to unlock the keypad. Note the LOW. SMART ALARM, (AL and HIGH icons will begin to flash indicating the SET OPERATING MODE.
  2. Press the UP (High Alarm) key on the keypad.NOTE: The High Alarm digits begin to flash indicating the High Alarm manual setting.
  3. Use the UP and DOWN keys to set the high alarm to the desired value. Pressing the arrow keys changes the value in I% increments. if the keys are held down for more than 1 second the display will scroll at a rate of 1% per second.NOTE If 30 seconds elapse between key actuations, the system will store the latest high alarm setting and will revert to normal operation. If this occurs inadvertently, simply repeat the alarm setting procedure.
  4. When the high alarm setting is set above 100%. the high alarm will indicate two dashes – -. This special condition turns off. or deactivates, the high alarm.
  5. When the high alarm value is set. press the Unlock again to accept the high alarm setting and return to normal operation.

NOTE: The default high alarm setting is 50% 02. Removing the batteries will reset the high alarm limit to 50%.

2.3.3 Smart Alarm Mode

  1. Press the Unlock Key to unlock the keypad. Note the LOW, SMART ALARM, CAL, and HIGH icons will begin to flash indicating the SET OPERATING MODE.
  2. Press the Smart Alarm key on the keypad. Note the LOW digits, Alarm Mode and HIGH digits begin a slow flash indicating SMART ALARM MODE. The high alarm will now be set equal to the current oxygen reading +3% (rounded to the nearest integer). The low alarm will now be set equal to the current oxygen reading -3% (rounded to the nearest integer, but never lower than 18%.
  3. Pressing the UP key will add 1% to the high alarm setting and subtract 1% from the low alarm setting. Pressing the DOWN key will subtract 1% from the high alarm setting and add 1% to the low alarm setting. In other words, the UP key widens the alarm band and the DOWN key tightens the alarm band. This feature will not set the alarm levels above 100% or below 18%.
  4. Once the desired alarm settings are attained, press the Unlock key again to save the settings and return to the normal operation mode. If 30 seconds elapse without a key press by the user, the device will automatically save the latest alarm settings and return to the normal operation mode.

2.4 Basic OperationTo check the oxygen concentration of the delivered gas:

  1. Ensure the sensor is connected to the flow diverter and inserted completely in the sensor port behind the flowmeter.
  2. Set the oxygen blender control knob to the desired oxygen setting.
  3. Using the ON/OFF key, make sure the unit is in the power-on mode.
  4. Allow the oxygen reading to stabilize.
  5. Adjust the flowmeter to the desired flow rate. Read the flow rate at the center of the float ball.

2.5 Alarm Conditions and PrioritiesIn the event of either a low alarm or high alarm condition, the corresponding LED will begin to flash, accompanied by the audio buzzer. Pressing the Silent key will deactivate the buzzer, but the LED and the alarm value digits on the display will continue to flash until the alarm condition has been rectified. If the alarm condition still exists 120 seconds after silencing the audio buzzer, the beeper will start to sound again.

ALARM

ALARMPRIORITY LOWALARM LED HIGHALARM HD AUDIBLEALARM

AUDIBLEALARM REPEAT

line Power Plugged ill Informational Off Off 2 Pulses No Repeat
Line PowerUnplugged Informational Single Yellow Pulse Single Yellow Pulse 2 Pulses No Repeat
External DC Power Supply Voltage Out of Range Informational Solid Yellow Solid Yellow 2 Pulses Every 15 Sec.
Battery Voltage too low for device to operate (E04) Medium Pulsing Yellow Pulsing Yellow 3 Pulses Every 25 Set
Oxygen/ pressure level above the high alarm setting Medium ON Pulsing Yellow 3 Pulses Every 25 Set
Oxygen/pressure level below the low alarm setting Medium Pulsing Yellow Off 3 Pulses Every 25 Sec.
Oxygen level below the low oxygen alarm setting and lower than 18% High Pulsing Red Off 5.5 Pulses Every 15 Sec.
  • A low alarm condition will remain until the actual concentration is 0.1% higher than the low alarm setting.
  • A high alarm condition will remain until the actual concentration is 0.1% lower than the high alarm setting.

To differentiate the level of severity, the monitor provides three unique sound bursts.

2.6 Backlight OperationTo turn on the backlighting:

  1. When the unit is on, pressing the Backlight button will turn the backlighting on for 30 seconds. An additional press will turn off the backlighting.
  2. If the device is being used in a dark location, any button press will activate the backlight.

CAUTION: Excessive use of the backlight can reduce the life of the batteries.

2.7 External Power Supply OperationTo extend the life of the batteries, an external Maxtec approved 7.5V DC external power supply can be purchased. Once connected to the unit, total power is supplied by the power supply.Batteries are still required in the unit and will provide emergency power in the event main AC power is lost.NOTE: Use only the Maxtec external power supply described in section 10.0.NOTE: The power supply is not a battery charger.DO NOT use rechargeable batteries.2.8 Calibration Procedures2.8.1 Calibration to 100% OxygenThe MaxBlend 2 should be calibrated before being placed into clinical use. Thereafter, Maxtec recommends calibration of the unit on a weekly basis. Frequent calibration will have no adverse effect on the performance of the MaxBlend 2. Calibration should also be performed upon the replacement of a sensor.The sensor is best calibrated while mounted in the MaxBlend 2 sensor port. As in normal operation, the oxygen sensor responds best when installed in a vertical position with the sensor facing down.Changes in barometric pressure can affect the oxygen reading. A 1% change in the barometric pressure results in an error of 1% of the actual reading. (Example: If you are reading a 50% oxygen mix and the barometric pressure drops from 1000bar to 990bar the reading will drop to 50% x (990/1000) = 49.5%). Maxtec recommends that you recalibrate after changing point-of-use elevation by more than 500 feet (150m).It is best to calibrate the MaxBlend 2 using the sensor port with a technical grade oxygen standard (99.0% or better). Calibration of the unit with room air is less accurate over the full FiO2 operating range.To calibrate to 100% oxygen:

  1. Connect the oxygen supply line (pressure differential alarm may sound). Verify the sensor is plugged into the O2 sensor port and connected to the sensor cable. DO NOT connect air supply line at this time.
  2. Using the ON/OFF key, make sure the MaxBlend 2 is in the power-on mode.
  3. Rotate the FiO2 control knob to the 100% stop. Allow a few minutes for the reading to stabilize.
  4. Press the Unlock key to unlock the keypad. Note the LOW, SMART ALARM, CAL, and HIGH icons will begin to flash indicating the SET OPERATING MODE.
  5. Press the CAL (Calibration) key on the keypad. The word “CAL” will appear on the display for approximately 5 seconds and then finish with 100.0%.
  6. The unit is now calibrated and in the normal operating mode.

2.8.2 Calibration to Room AirThe MaxBlend 2 can quickly be calibrated to room air (20.9%).To calibrate to room air:

  1. Connect the air supply line (pressure differential alarm may sound). Verify the sensor is plugged into the O2 sensor port and connected to the sensor cableDO NOT connect the oxygen supply line at this time. (If preferred, room-air calibration may be performed by removing the sensor from the O2 sampling port and detaching the flow diverter. If using this method, the gas-supply lines may remain attached.)
  2. Using the ON/OFF key, make sure the MaxBlend 2 is in the power-on mode.
  3. Rotate the FiO2 control knob to the 21% stop. Allow a few minutes for the reading to stabilize.
  4. Press the Unlock key to unlock the keypad. Note the LOW, SMART ALARM, CAL and HIGH icons will begin to flash indicating the SET OPERATING MODE.
  5. Press the CAL (Calibration) key on the keypad. The word “CAL” will appear on the display for approximately 5 seconds and then finish with 20.9%.
  6. The unit is now calibrated and in the normal operating mode.

2.9 Flowmeter Operation

  1. Adjust the flowmeter to the desired setpoint as read by the center of the float ball.• To increase flow—turn knob counter-clockwise• To decrease flow—turn knob clockwise
  2. Check the toggle bleed switch to ensure it is in the appropriate position.• For a low flow blender the bleed toggle switch should be turned on (up position) if the total combined flow from the blender is less than 3LPM.• For a high flow blender the bleed toggle switch should be turned on (up position) if the total combined flow from the blender is less than 15LPM.• At delivered flows greater than the above limits, the bleed toggle switch can be turned off (down position) to conserve gas.

CAUTION: Failure to activate the bleed as described above may result in inaccurate oxygen concentrations from the blender. However, the MaxBlend 2 will always display the actually delivered concentration.CAUTION: The outlets of this device are capable of delivering pressure as high as the inlet pressure. Ensure that the devices that carry the gas from the blender to the patient prevent excessive pressure on the patient.

PERFORMANCE CHECK

Prior to placing the MaxBlend 2 into clinical use, perform the following tests.WARNING: If the MaxBlend 2 does not function as described on the next page, contact your Maxtec Distributor or contact a Maxtec Certified Service Technician at:Maxtec2305 South 1070 WestSalt Lake City, UT 84119(385) 549-8000 or (800) 748-5355DO NOT use the MaxBlend 2 until the correct performance has been verified.3.1 Blender Safety CheckNOTE: Before proceeding, ensure that the high alarm setpoint control is OFF [display reads (–)] and that the low alarm setpoint control is below 20%.

 

PROCEDURE

BLENDER RESPONSE

1. Conned the blender to 50 ±5 PSIG air/oxygen gas sources. Adjust the blender control knob to 60%. Turn the bleed toggle switch on (up position) and adjust the flowmeter to a minimum of 2 LPM. No response. Monitor display reads 60% ±3%.
2. Disconnect the 50 PSIG AIR source from the MaxBlend 2. NOTE: The blender must be flowing

gas for the alarm to activate.

 Audible alarm sounds. Monitor display reads 100%3

±%.

3. Reconnect 50 PSIG AIR source to the MaxBlend 2. The audible alarm stops.Verify the monitor display reads 60% ±3%.
4. Disconnect 50 PSIG OXYGEN source from the MaxBlend 2. Audible alarm sounds. Monitor display reads 20.9% ±3%.
5. Reconnect 50 PSIG OXYGEN to the MaxBlend 2. The audible alarm stops.Verify the monitor display reads 60% ±3%.
6. Adjust both air and oxygen inlet regulators to 0 PSIG. No response.
7. Remove the air inlet hose at the regulator and insert the end into a beaker of water. No response.
8. Slowly raise the pressure of the oxygen regulator to 50 PSIG and back to 0 PSIG while observing the air hose end in a beaker. No bubbles should be observed. Audible alarm sounds.
9. Dry and reattach air inlet hose to regulator. No response.
10. Remove oxygen inlet hose at the regulator, and insert the end into a beaker of water. No response.
11. Slowly raise the pressure of the air regulator to 50 PSIG and back to 0 PSIG while observing the oxygen hose end in a beaker. No bubbles should be observed. Audible alarm sounds.
12. Dry and reattach oxygen inlet hose to regulator. No response.

TROUBLESHOOTING

PROBLEM: Oxygen concentration discrepancy between oxygen concentration selection knob and actual reading on display, greater than 3%.POTENTIAL CAUSES AND SOLUTIONS:

  • The bleed is turned off. Turn bleed toggle switch on. Refer to section 2.9, Flowmeter
  • Monitor out of calibration. Calibrate. Refer to section 2.8, Calibration Procedure.
  • Sensor exhausted. Replace sensor. Refer to section 6.2.
  • Gas supply contaminated. Contact Maxtec for the repair of the MaxBlend 2.
  • Blender out of calibration. Contact Maxtec for repair.

PROBLEM: Pressure differential alarm sounding.POTENTIAL CAUSES AND SOLUTIONS:

  • Inlet pressure differences of 20 PSI or more. Correct pressure difference.
  • The pressure alarm was not calibrated properly. Contact Maxtec for repair.
  • MaxBlend 2 blender operation out of calibration. Contact Maxtec for repair.

PROBLEM: Inlet pressure has supply loss, no audible pressure differential alarm.POTENTIAL CAUSES AND SOLUTIONS:

  • Reed alarm cap damaged or defective. Contact Maxtec for repair.

PROBLEM: Selected oxygen concentration is accurate only when gas pressures are equal.POTENTIAL CAUSES AND SOLUTIONS:

  • MaxBlend 2 balance module not functioning properly. Contact Maxtec for repair.

PROBLEM: Blank display.POTENTIAL CAUSES AND SOLUTIONS:

  • Batteries not installed. Install batteries. Refer to section 2.1.1, Battery Installation.
  • The battery is completely dead. Replace batteries. Refer to section 2.1.1, Battery Installation.
  • Monitor defective. Contact Maxtec for repair.

PROBLEM: Partial or distorted display.POTENTIAL CAUSES AND SOLUTIONS:

  • Monitor damaged. Contact Maxtec for repair.

PROBLEM: The sensor will not calibrate.POTENTIAL CAUSES AND SOLUTIONS:

  • Sensor cell exhausted. Replace sensor. Refer to section 6, Replacing 02 sensor.
  • Sensor cable defective. Return to Maxtec.
  • Monitor defective. Contact Maxtec for repair.

PROBLEM: The sensor will calibrate, but takes too long or does not return to 21%1’2% oxygen in the air (2 to 5 minutes) when performing calibration.POTENTIAL CAUSES AND SOLUTIONS:

  • Disposable oxygen sensor damaged or defective. Replace sensor. Refer to section 6.2, Replacing 02 sensor.

PROBLEM: The sensor will calibrate, but reading at any constant level drifts more than ±3% over a 24 hour period.POTENTIAL CAUSES AND SOLUTIONS:

  • Barometric pressure changed since the last calibration. Recalibrate. Refer to section 2.8, Calibration Procedures.
  • Room or gas temperature went below 59°F (15°C) or above 104°F (40°C). The correct temperature and recalibrate.

PROBLEM: Low battery icon.POTENTIAL CAUSES AND SOLUTIONS:

  • If at any time, the low battery icon is displayed on the LCD readout, the batteries should be replaced as quickly as possible.

PROBLEM: E01: Sensor voltage is too low to perform a valid calibration.POTENTIAL CAUSES AND SOLUTIONS:

  • Manually attempt a new calibration. Refer to section 2.8, Calibration Procedures.
  • If the unit repeats this error more than three times, contact Maxtec’s Customer Service Department for possible sensor replacement.

PROBLEM: E02: No sensor attached.POTENTIAL CAUSES AND SOLUTIONS:

  • Disconnect the sensor and reconnect, making sure the male plug is fully inserted into the receptacle. The analyzer should now perform a new calibration with the error
  • If the error still persists, remove the batteries, wait 30 seconds, then reinstall, to perform a factory reset and diagnostic on the analyzer. The analyzer should again perform a new calibration with the error cleared.
  • Contact Maxtec Customer Service Department if the error code cannot be cleared.

PROBLEM: E03: No valid calibration data available.POTENTIAL CAUSES AND SOLUTIONS:

  • Make sure the unit has reached thermal equilibrium and perform a calibration routine. Refer to section 2.8, Calibration Procedures.

PROBLEM: E04: Battery below minimum operating voltage.POTENTIAL CAUSES AND SOLUTIONS:

  • Replace batteries. A medium priority alarm will sound every 25 seconds until the batteries are replaced or become too dead to sound the alarm. Refer to section 2.1.1, Battery Installation.

PROBLEM: E05: Sensor voltage is too high to perform a valid calibration.POTENTIAL CAUSES AND SOLUTIONS:

  • Manually attempt a new calibration. Refer to section 2.8, Calibration Procedures.
  • If the unit repeats this error more than three times, contact Maxtec’s Customer Service Department for possible sensor replacement.

PROBLEM: E06: Non-compatible oxygen sensor.POTENTIAL CAUSES AND SOLUTIONS:

  • Disconnect the sensor and reconnect, making sure the male plug is fully inserted into the receptacle. The analyzer should now perform a new calibration with the error cleared.
  • If the error still persists, remove the batteries, wait 30 seconds, then reinstall, to perform a factory reset and diagnostic on the analyzer. The analyzer should again perform a new calibration with the error cleared.
  • Contact Maxtec Customer Service Department if the error code cannot be cleared.

PROBLEM: E07: The sensor signal is not stable enough to perform a valid calibration.POTENTIAL CAUSES AND SOLUTIONS:

  • Wait for the displayed oxygen reading to stabilize when calibrating the device at 100% oxygen.
  • Wait for the unit to reach thermal equilibrium. Please note that this can take up to one-half hour if the device is stored in temperatures outside the specified operating temperature range.

PROBLEM: E08: Battery voltage is too low to perform a valid calibration.POTENTIAL CAUSES AND SOLUTIONS:

  • Replace batteries. Refer to section 2.1.1, Battery Installation.

NOTE: Use only the Maxtec-approved Max-550E sensor described in section 10.0, Spare Parts and Accessories. The Max550E sensor is equipped with an authentication chip to ensure the monitor is used with an approved sensor.NOTE: The operator must be facing the device and positioned within 4 meters to distinguish the visual alarm indicators. Audible alarms can be distinguished as long as the operator is in the same room, and the ambient noise level is typical for a clinical setting.

CLEANING AND DISINFECTING

The external surfaces of the device, and its accessories, can be cleaned and disinfected using the process detailed below. Under normal use conditions, the sensing surfaces of the sensor should not become contaminated. If you suspect that the sensing face of the sensor or internal surfaces of the flow diverter has become contaminated, these items should be discarded and replaced. Store the device in a clean, dry location when not in use.To clean and disinfect:

  1. Ensure the battery drawer is closed and sensor/diverter is inserted into their port.
  2. Using disposable Super Sani-Cloth germicidal wipes (medical-grade 2-in-1 cleaning/disinfecting wipes), remove all visible contamination from the external surfaces of the device and its accessories. Be sure to closely inspect and remove contamination from seams and recesses on the device that may trap contaminants. Wipe with a clean paper towel to remove debris and bioburden.
  3. After all visible contamination is removed, use a second germicidal wipe to thoroughly wet the surfaces of the device and accessories. Allow remaining wet for 4 minutes. Use additional wipes, if needed, to assure surfaces are wetted continuously for 4 minutes.
  4. Allow the device to air dry completely.
  5. Visually inspect the device for visible contamination. Repeat cleaning/disinfection process if visible soil remains.

DO NOT allow the liquid or spray to penetrate the device.

DO NOT spray cleaning solution directly onto the sensor port, bleed muffler, or buzzer openings.Be sure to thoroughly clean and disinfect the areas depicted in the images below. These regions are contacted during normal use and may contribute to cross-contamination if not disinfected properly.CAUTION: Excessive rubbing of labels may cause them to become illegible.DO NOT immerse the device or sensor into liquid decontamination agents.DO NOT use strong solvent cleaners.DO NOT allow cleaning liquids to contact the face of the sensor as this may impair the readings of the sensor.DO NOT Attempt to sterilize the device with steam, ethylene oxide, or irradiation.

SERVICE AND MAINTENANCE

6.1 MaintenanceMaxtec recommends that the Performance Check listed in section 3.0 be performed at least once a year.When using the MaxBlend 2 with a medical-grade compressed air source, it is recommended that an air inlet water trap/filter be attached to the air inlet of the MaxBlend 2 prior to use. Contaminants from hospital airlines may compromise the function of the MaxBlend 2.Maxtec recommends that the MaxBlend 2 be overhauled and serviced at a minimum of every three years.Repair of this equipment must be performed by a Maxtec Certified Service Technician experienced in the repair of this device.6.2 Replacing O2 SensorThe oxygen sensor is designed to operate for two years under normal use conditions. The oxygen sensor should be replaced whenever any of the problems listed in section 4.0, Troubleshooting dictates the need to do so.To replace the O2 sensor:

  1. Remove the sensor from the sensor monitor port.
  2. Remove the sensor from the sensor cable.
  3. Install a new O2 sensor with a flow diverter and attach it to the sensor cable.
  4. Calibrate the sensor following the instructions for calibration listed in section 2.8.

6.3 Monitor Alarm Testing

  • Testing of alarms should be performed on a yearly basis.
  • To check the low alarm, adjust the low alarm setting to 23%, or higher, and expose the sensor to room air (20.9%). The low alarm LED should flash with the alarm sound.
  • To check the high alarm, adjust the low alarm setting to 17%, or lower, and the high alarm setting to 18% and expose the sensor to room air (20.9%). The high alarm LED should flash with the alarm sound. If one, or both alarms malfunction, contact a Maxtec Certified Service Technician.

ABBREVIATION GUIDE

TERM

DESCRIPTION

Air/02 A mixture of compressed air and oxygen
°C Degrees Celsius
CGA Compressed Gas Association
DISS Diameter Indexed Safety System
°F Degrees Fahrenheit
Fi02 Fractional Concentration of Inspired Oxygen
02 Oxygen
LPM Liters Per Minute
PSIG Pounds Per Square Inch Gauge

SPECIFICATIONS

8.1 Instrument Specifications

Weight (unpackaged) 5.3 lbs. (2.4 kg.)
Power Source Four “AA” alkaline batteries, 1.5 V each
Battery Life 5000 hours (continuous operation, no alarming)
Oxygen Measurement Range 0% to 100% oxygen
Display Resolution 0.1% oxygen
O2 Concentration Adjustment Range 21% to 100% O2
Gas Supply Pressure The gas supplies must provide clean, dry, medical-grade air and oxygen at a pressure of 30 to 75 PSIG (2.0 to 5.2 BAR).Air and oxygen must be within 20 PSI (1.3 BAR). Optimal performance is achieved at 50 PSIG inlet pressures.
Pressure Drop Less than 6 PSIG (0.4 BAR) @ 50 PSIG (3.4 BAR) supply pressures and 10 LPM flow rate
Sensor Bleed Flow 0.1LPM @ 50 PSIG (3.4 BAR)
Bleed Flow (toggle ON) 3 LPM for low flow blender and 13 LPM for high flow blender
Outlet Flow Range 0-30 LPM for low flow blender and 2-100 LPM for high flow blender with inlet pressures at 50 PSIG (3.4 BAR)
Mixed Gas Stability* Ambient Operating Conditions ±1% oxygen
Operating Temperature Range 59°F to 104°F (15°C to 40°C)
Relative Humidity Range 0-95%, non-condensing
Ambient Storage Conditions Temperature Range 5°F to 122°F (-15°C to 50°C)
Flowmeter accuracy** +/-10% of indicated value or 0.5 LPM whichever is greater, with inlet pressure set to 50PSIG.
Flush Flow 3 LPM Flowmeter: 20-30 LPM15 LPM Flowmeter: 20-30 LPM30 LPM Flowmeter: 35-45 LPM70 LPM Flowmeter: 70-80 LPM

*The delivered oxygen concentration will remain constant within ±1% of the setpoint value with constant inlet pressures. The displayed value may vary more than this due to the sensor accuracy, age, environmental conditions, and the length of time since the last sensor calibration.**Position the device such that the flowmeters are vertical to ensure accuracy.8.2 Alarm Specifications

Pressure Differential Alarm Activation When supply pressures differ by 20 PSI (1.3 BAR) or more alarm is activated
Low Oxygen Alarm Range 15%-99% (>1% lower than high alarm)
High Oxygen Alarm Range 16%-100% (>1% higher than low alarm) (according to IEC 60601-1-8 Audible Alarms in Medical Equipment)

8.3 O2 Sensor Specifications

Total accuracy* ±3% actual oxygen level over the full operating temperature range
Oxygen Measurement Accuracy ±1% oxygen
Linearity ± 1% at constant temperature and pressure
Error Over Operating Temp Range ±3% oxygen, maximum
Response Time to 90% of Final Reading* @ 77°F (25°C) ≤20 seconds
Storage Temperature Range 5°F to 122°F (-15°C to 50°C)
Expected Useful Life 1,500,000 O2% hours (approx. 2 years average use)

*The accuracy of the oxygen monitor is not affected by supply gas inlet pressure to the blender, however, pressures below 50 PSIG may result in a longer response time.NOTE: All specifications assume the following standard environmental conditions unless specified otherwise:

  • Ambient and sample gas temperatures of 77°F (25°C)
  • Barometric pressure of 30 inHg (102 kPa)
  • Ambient relative humidity of 50%
  • Sample gas relative humidity of 0%

FACTORS INFLUENCING CALIBRATION

9.1 Temperature EffectThe MaxBlend 2 Monitor will hold calibration and read correctly within +/-3% when in thermal equilibrium within the operating temperature range.The device accuracy will be better than +/-3% if operated at the same temperature at which it was calibrated. The device must be thermally stable when calibrated and allowed to thermally stabilize afterexperiencing temperature changes before reading is accurate. For these reasons, the following is recommended:

  1. Allow adequate time for the sensor to equilibrate to a new ambient temperature.Please note that this can take up to one-half hour if the device is stored in temperatures outside the specified operating temperature range.
  2. For best results, perform the calibration procedure at a temperature close to the temperature where analysis will occur.

9.2 Pressure EffectChanges in barometric pressure can affect the oxygen reading. A 1% change in the barometric pressure results in an error of 1% of actual reading (Example: If you are reading a 50% oxygen mix and the barometric pressure drops from 30kPa to 29kPa, the reading will drop to 50% x (29/30) = 48.3%. Maxtec recommends that you recalibrate after changing point-of-use elevation by more than 500 feet(150m).9.3 Humidity EffectHumidity in the sample gas will affect the oxygen reading. Maxtec recommends that the gas delivered to the MaxBlend 2 be medical-grade, clean, and dry. Refer to ISO 7396-1 for further details.9.4 Exposure to Anesthetic GasesBecause of the unique chemistry of the oxygen sensors provided with the MaxBlend 2, there are no significant effects when exposed to commonly used anesthetic gases; however, the monitor is not designed for exposure to flammable gas mixtures (See WARNING page 2).

DEMENT

ROME % DRY

INTERFERENCE IN 02%

Nitrous Oxide 60%, balance 02 <1.5%
Halothane 4% <1.5%
Enflurane 5% <1.5%
Isoflurane 5% <1.5%
Helium 50%, balance 02 <1.5%
Sevoflurane 5% <1.5%
Desflurane 15% <1.5%

NOTE: Balance mixture 30% 02/70%N20, unless otherwise specified.

SPARE PARTS AND ACCESSORIES

DESCRIPTION PART NUMBER
MAX550E Oxygen Sensor R140P02-001

ACCESSORIES DESCRIPTION

Monitor Cable R228P49
Rail Mount Bracket R100P09
Adjustable Pole Mount Bracket R100P22
Pole Mount Bracket R100P26
Compact Wall Mount RP05P07
Maxtec-Approved Power Supply R230P10
Wall Mount Large Bracket RPO5P09
10′ Dual Blender hose (DISS) R129P01

Repair of this equipment must be performed by a Maxtec Certified Service Technician, experienced in the repair of this device.Equipment in need of repair shall be sent to:MaxtecService Department2305 South 1070 WestSalt Lake City, Ut 841191.800.748.5355(Include RMA number issued by Customer Service)

THEORY OF OPERATION

11.1 Operational Diagram

11.2 Mixing OperationThe MaxBlend 2 is designed to utilize two 50 PSIG (3.4 BAR) gas sources. The two gas sources enter through the air and oxygen inlet connectors located on the bottom of the MaxBlend 2.Each inlet connector incorporates a 30-micron particulate filter. Once through the filters, each gas passes through a duckbill check valve, which prevents possible reverse gas flow from either the air or the oxygen supply systems.The two gases then pass through a two-stage balanced regulator. The purpose of this regulator is to equalize the operating pressures of the air and oxygen gas sources. Once these pressures have been balanced, the gases are proportioned according to the oxygen concentration selected on the oxygen concentration selection knob. The oxygen concentration knob allows the clinician to select the desired oxygen concentration from 21% to 100% O2. From this point, the mixed gas flows to the outlet port.11.3 Gas OutletThere are three gas outlets on the MaxBlend 2. One is a fixed acrylic flowmeter, the other two are auxiliary ports (one on the bottom of the unit, and one on the right side). These outlets are capable of delivering combined metered flows of 0-30 LPM for the low flow model and 0-120 LPM for the high flow model. The auxiliary outlet ports are fitted with an automatic shut-off valve. The flow of gas from either outlet port is automatically initiated by attaching a pneumatic device (such as a flowmeter) to the outlet port.Regardless of whether or not the outlet has any device connected to it, a minimal gas bleed flow of 0.1 LPM flows from the MaxBlend 2 sensor port on the left side of the blender. It is from this bleed flow that the gas is analyzed by the oxygen sensor. In addition, a toggle switch is provided allowing the user to activate an additional gas bleed, which ensures the lender has sufficient flow to function accurately when the total flow delivered to the patient is below a certain minimum threshold. For a low flow model, this additional bleed should be activated if the total flow delivered to the patient is less than 3 LPM. For a high flow model, the additional bleed should be activated if the total flow delivered to the patient is less than 15 LPM. At delivered flows greater than these limits, the bleed togglecan be deactivated to conserve oxygen.CAUTION: Failure to activate the bleed as described above may result in a significant drift in the oxygen concentration delivered to the patient.11.4 Alarm/Bypass FunctionThe MaxBlend 2 includes a pressure differential alarm which provides an audible alarm if gas source pressures differ by 20 PSI (1.3 BAR) (nominal) or more, or if there is a gas supply failure of one of the source gases. This alarm is generated by a reed alarm located in a cap on the bottom of the MaxBlend 2.The primary purpose of the alarm is to audibly warn the operator of an excessive pres­sure drop or depletion of either source gas pressure. Should both gas pressures increase or decrease simultaneously, an alarm will not activate.If either source gas pressure drops, the outlet pressure will drop similarly, as the mixed gas is always balanced to that of the lower gas source.The gas bypass function operates in unison with the alarm. Once the pressure alarm is acti­vated, the bypass function is actuated and the gas with the higher pressure flows directly to the outlet port, bypassing the mixing function of the MaxBlend 2. The oxygen concentration flowing out of the MaxBlend 2 will be that of the gas with the higher pressure. In the alarm/ bypass mode, the blender will deliver oxygen (100%) or medical air (21%) until pressures have been restored to a differential of 6 PSI or less (0.4 BAR).If the MaxBlend 2 is set to deliver 21% and the OXYGEN source pressure is reduced enough to produce a 20 PSI (1.3 BAR) differential, the unit may not alarm because it will continue to deliver 21% concentration according to the setting. If the setting is moved slightly from 21%, the pressure differential alarm will sound. Similarly, if the MaxBlend 2 is set to deliver 100% and the AIR source pressure is reduced or lost, the unit may not alarm because it will continue to deliver 100% concentration.

FLOW CHARACTERISTICS

The outlet pressure of the MaxBlend 2 decreases as the total flow rate increases. The total flow rate is the measurement of the total flow from all outlet ports. The charts below indicate the pressure drop that occurs for both low flow, and high flow, models at 3 inlet pressure settings; 30 PSIG (2.07 BAR), 50 PSIG (3.45 BAR), and 75 PSIG (5.17 BAR).The fixed acrylic flowmeter on the left side of the MaxBlend 2 has been pressure compensated to accommodate for the pressure loss through the blender at each flow rate, using an inlet pressure of 50 PSIG.Flow Rate -vs- Pressure Drop:

ELECTROMAGNETIC COMPATIBILITY

The information contained in this section (such as separation distances) is in general specifi­cally written with regard to the MaxBlend 2. The numbers provided will not guarantee fault­less operation but should provide reasonable assurance of such. This information may not be applicable to other medical electrical equipment; older equipment may be particularly susceptible to interference.NOTE: Medical electrical equipment requires special precautions regarding electromagnetic compatibility (EM() and needs to be installed and put into service according to the [MC infor­mation provided in this document and the remainder of the instructions for use this device.Portable and mobile RF communications equipment can affect medical electrical equipment.Cables and accessories not specified within the instructions for use are not authorized. Using other cables and/or accessories may adversely impact safety, performance, and electromag­netic compatibility (increased emission and decreased immunity).Care should be taken if the equipment is used adjacent to or stacked with other equipment; if adjacent or stacked use is inevitable, the equipment should be observed to verify normal operation in the configuration in which it will be used.ELECTROMAGNETIC EMISSIONS

This equipment is intended for use in the electromagnetic environment specified below. The user of this equipment should assure that it is used in such an environment.

EMISSIONS

COMPLIANCE ACCORDING TO

ELECTROMAGNETIC ENVIRONMENT

RF Emissions (CISPR 11) Group 1 The MaxBlend 2 uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.
CISPR Emissions Classification (lass A The MaxBlend 2 is suitable for use in all establishments other than domestic and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes.NOTE: The EMISSIONS characteristics of this equipment make it suitable for use in industrial areas and hospitals (CISPR 11 class A). If it is used in a residential environment (for which CISPR11 class B is normally required) this equipment might not offer adequate protection to radio-frequency communication services. The user might need to take mitigation measures, such as relocating or re-orienting the equipment.
Harmonic Emissions (1E( 61000-3-2) Class A
Voltage Fluctuations Complies

ELECTROMAGNETIC IMMUNITYThis equipment is intended for use in the electromagnetic environment specified below.The user of this equipment should assure that it is used in such an environment.

IMMUNITY AGAINST

1E( 60601-1-1: (4TH EDITION) TEST LEVEL

ELECTROMAGNETIC ENVIRONMENT

Professional Healthcare Facility Environment Home Healthcare Environment
Electrostatic discharge, ESD (1E( 61000-4-2) Contact discharge: ±8 kV Air discharge: ±2 kV, ±4 kV, ±8 kV, ±15 kV Floors should be wood, concrete, or ceramic tile. If floors are covered with synthetic material, the relative humidity should be kept at levels to reduce the electrostatic charge to suitable levels.The main power quality should be that of a typical commercial or hospital environment.Equipment that emits high levels of power line magnetic fields (in excess of 30A/m) should be kept at a distance to reduce the likelihood of interference.If the user requires continued operation during power mains interruptions, ensure that batteries are installed and charged.Ensure that battery life exceeds the longest anticipated power outages or provide an additional uninterruptible power source.
Electrical fast transients / bursts (1E( 61000-4-4) Power supply lines: ±2 kVLonger input/output lines: ±1 kV
Surges on AC mains lines (1E( 61000-4-5) Common mode: ±2 kV Differential mode: ±1 kV
3 Alm power frequency magnetic field 50/60 Hz (1E( 61000-4-8) 30 A/m 50 Hz or 60 Hz
Voltage dips and short interruptions on A( mains input lines (1E( 61000-4-11) Dip>95%, 0.5 periods Dip 60%, 5 periods Dip 30%, 25 periods Dip >95%, 5 seconds
report this ad

Recommended separation distances between portable and mobileRF communications equipment and the equipment

RATED MAXIMUM OUTPUT POWER OF TRANSMITTER W Separation distance according to the frequency of transmitters in meters
150 kHz to 80 MHzd=1.2/V1] √P 80 MHz to 800 MHzd=1.2/1/1] √P 800MHz to 2.5 GHzd=2.3 √P
0.01 0.12 0.11 0.13
0.1 0.38 038 0.73
1 1.1 1.1 1.3
10 3.8 3.8 7.3
100 11 11 23

For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects, and people.This equipment is intended for use in the electromagnetic environment specified below. The customer or the user of this equipment should assure that it is used in such an environment.

IMMUNITY TEST IEC 60601-1-2: 2014 (4TH EDITION) TEST LEVEL ELECTROMAGNETIC ENVIRONMENT – GUIDANCE
Professional Healthcare Facility Environment Home Healthcare Environment
Conducted RF coupled into lines (IEC 61000-4-6) 3V (015 – 80 MHz) 6V (ISM bands) 3V (015 – 80 MHz) 6V (ISM & Amateur bands) Portable and mobile RF communications equipment (including cables) should be used no closer to any part of the equipment than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter as below.Recommended separation distance:d=1.2 √Pd=1.2 √P 80 MHz to 800 MHzd=2.3 √P 800 MHz to 2.7 GHzWhere P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in meters (m).Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey a, should be less than the compliance level in each frequency range b. Interference may occur in the vicinity of equipment marked with the following symbol:
Radiated RF immunity (IEC 61000-4-3) 3 V/m 80 MHz – 2.7 GHz 80% @ 1 kHz AM Modulation d=1.2 VP  10 V/m 80 MHz – 23 GHz 80% @ 1 kHz AM Modulation

The ISM (industrial, scientific and medical) bands between 150 kHz and 80 MHz are 6,765 MHz to 6,795 MHz; 13,553 MHz to 13,567 MHz; 26,957 MHz to 27,283 MHz; and 40,66 MHz to 40,70 MHz.Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM, and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the equipment is used exceeds the applicable RF compliance level above, the equipment should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the equipment.

2305 South 1070 WestSalt Lake City, Utah 84119(800) 748-5355www.maxtec.com

References

[xyz-ips snippet=”download-snippet”]


Posted

in

by