INSTRUCTION MANUAL Blood pressure watch BPW 300
Device and controls
Device and controls
- Display
- START/STOP button
- MEM button
- Wrist cuff
- Wristband
- USB port (for charging)
Explanation of symbols
This instruction manual belongs to this device. It contains important information about starting up and operation. Read the instruction manual thoroughly. Non-observance of these instructions can result in serious injury or damage to the device.
IP22 Ingress protection rating
LOT Lot number
51215 05/2020 Ver. 2.1
IMPORTANT INFORMATION! RETAIN FOR FUTURE USE!
Safety notes
- Contact your physician if you suspect you have a medical issue.
- Not for use by children, pregnant women, or pre-eclamptic patients.
- If you suffer from arrhythmia, diabetes, blood circulation problems or apoplexy, only use the device under a physician’s care.
- Prolonged over-inflation of the monitor will result in harmful injury to the patient.
- Do not use this device over a wound, as this can cause further injury.
- Too frequent measurements can cause injury due to blood flow interference.
- This device possibly cause improper result due to external interference, such as acceleration during transport or transport in general.
- Do not attempt to modify this device in any way.
- Store this device in a cool, dry place. Do not subject this device to extreme temperatures, humidity or sunlight. This device might not meet performance specifications if stored or used outside the ranges specified in the “Specifications” section. Keep away from rain.
- Do not immerse this device in water or clean with cleaning products, alcohol or solvents. Carefully follow cleaning instructions provided.
- Ensure that the device is used in the environment specified in the EMC declaration in this instruction manual, otherwise, may result in improper operation.
- Use of this device adjacent to or stacked with other device should be avoided because it could result in improper operation.
- This electrical medical equipment requires specific precautions regarding electromagnetic compatibility. It must be installed and used according to the electromagnetic information.
- Do not use this device where flammable gases or liquids are present.
- Never drop the device.
- Do not replace the wristband with the one not from the manufacturer, or may result in inaccurate measurement result. The cuff fits wrist circumference ranging from 5.3 inches to 8.5 inches (approx.13.5 cm to 21.5 cm), the user should confirm if the cuff fits the wrist before purchasing/using the device.
Calibration and Service
The accuracy of this blood pressure monitor has been carefully tested and is designed for a long service life. It is generally recommended to have the device inspected and calibrated every two years to ensure correct functioning and accuracy. Please consult your authorized distributor.
Intended use
This wrist blood pressure monitor is intended to be used to measure blood pressure (systolic and diastolic) and heart rate from the wrist by using the oscillometric method.The device is intended for using in only adult population, not applied to the other populations such as neonatal baby. It can’t be used while the wrist (arm) has bleeding or wound to avoid the blood flowing from the wound in pressurizing. The cuff circumference is limited to 13.5 cm – 21.5 cm.
Blood pressure indicator chart
When systolic and diastolic pressures fall into different categories, the higher pressure determines color code. Blood pressure measurements determined with this device are equivalent to those obtained by a trained observer using the cuff/stethoscope auscultatory method, within the limits prescribed by European Society of Hypertension International Protocol revision 2010 for the validation of blood pressure measuring devices in adults.
7 Systolic blood pressure (mmHg)8 Diastolic blood pressure (mmHg)9 Optimal blood pressure (target value)10 Normal blood pressure11 Normal systolic value12 Mild hypertension13 Moderate hypertension14 Severe hypertension
Scope of supply
Please check first of all that the device is complete and is not damaged in any way. In case of doubt, do not use it and contact your dealer or your service centre. The following parts are included:
- 1 medisana Blood pressure watch BPW 300
- 1 Instruction manual
- 1 USB cable
- 1 Soft ruler
ApplicationBefore Taking a Measurement
- For most accurate results, measurements should be taken in the same circumstances and at the same time every day.
- The device can be applied to either wrist, but preferably the left. Use the same wrist for each measurement.
- Any blood pressure recording can be affected by the position and physiological condition of the subject, as well as other factors such as background noise, stress, talking and body movement etc.
- Avoid taking measurements while under stress.
- Do not eat, drink alcohol, smoke, exercise, or bathe 30 minutes before using the device, as these activities may change blood pressure.
- For an accurate indication of true blood pressure, take and record several measurements over a period of time.
- If consecutive measurements are desired, allow two minutes between uses to allow blood to flow through the arm normally.
- This device is used to monitor blood pressure. However, it cannot treat or prevent high blood pressure. If you suffer from high blood pressure, consult your doctor.
Correct wearing and measuring precaution
- Remove sleeve, wristwatch and other jewelry. The device should be used on bare skin. Loosely roll up your sleeve to expose wrist if necessary.
Note: Tightening by your sleeve may result in inaccurate measurement result.
Measure wrist circumference
- Measure wrist circumference by using a soft ruler at the measuring point which is 20mm away from root of styloid process of ulna as the figure on the right:
- Note: Inaccurate measuring wrist circumference may result in inaccurate measurement results.Determine the hole number by which the wristband will be locked
- Determine the number of hole by which the wristband will be locked according to the wrist circumference measured above.
wrist circumference (mm) | 135~140 | 141~147 | 148~157 | 158~167 | 168~178 | 179~184 | 185~192 | 193~200 | 201~208 | 209~215 |
hole number | H4 | H5 | H6 | H7 | H8 | H9 | H10 | H11 | H12 | H13 |
Note: The incorrect number of hole for lock may result in inaccurate measurement results.
Wear the devicePosition the device and leave width of your forefinger (approx. 1.0 – 1.5 cm) between the wristband and the bottom of your palm. wrap and fasten wristband with cuff around your wrist properly.
Take a Measurement
- Comfortably seated • Legs uncrossed • Feet flat on the floor • Back and arm supported
- Keep the device at the same level as your heart
- Relax as much as possible and not talk during the measurement.While sitting upright in a chair with your feet flat on the floor, press START/STOP button 2.Cuff will begin to inflate. Do not move, activate arm muscles, or talk during measurement.Measurement can be stopped at anytime by pressing START/STOP button 2.When the device senses a pulse, a heart symbol will flash on screen.When measurement is complete, the cuff deflates automatically and the systolic pressure, diastolic pressure, and pulse rate will display on screen.MEM button 3 : press the button to recall last measurement.
Connecting to VitaDock+
Free download of the VitaDock+ app in the app store or Google Play store.Activate Bluetooth® on your iOS or Android appliance. Start the app. Touch the Bluetooth® symbol to open the list of devices. Scroll down the device list until you see the BPW 300.Touch the arrow beside the device to start the setup process.
Cleaning and maintenance
- For regular maintenance, the device only needs to be wiped gently with a soft, dry cloth while it is not being used. Never immerse the device or any components in water.
- Do not carry out repairs of any kind yourself. If a defect occurs, please contact your local authorized distributor. Use only authorized parts and accessories.
- The device’s battery should be recharged periodically, even during storage since the rechargeable battery installed in the device will self-discharge, leading to an unacceptably low voltage, thus damaging the battery.
Troubleshooting
Screen text | Possible cause | Solution |
Wearing loose | wearing is too loose | Check if the cuff is leaking.Maintenance may be needed if leaking.or: Wrap the cuff properly. |
Body motion interference | body motion | Do not move your arm and body, keep quiet |
Wearing tight | wearing is too tight | Wrap the cuff properly |
Wearing inappropriate | wearing is not appropriate | Wrap the cuff properly |
C6:D5:6D:94:ED:4A | Bluetooth code for binding | Please bind the collateral device and the BPW 300 |
The battery voltage is too low | Please charge in time | |
Charging mode | Device is charging | |
Wrong values | Change measurement unit | Press and hold button 2 and button 3 for 5 sec., then change unit with button 3 and confirm with button 2 |
Technical specifications | |
Name and model | medisana blood pressure watch BPW 300 |
Display | E-Ink LCD |
Measuring Principle | Oscillometric Method |
Measurable Circumference of Wrist | 13.5~21.5cm |
Electric Shock Protection | Internal power supply appliance type BF |
Measurement Range(Static pressure; Pulse) | 0 to 299 mmHg (0 to 39.9kPa); 40 to 180 Pulses/min |
Accuracy(Static pressure; Pulse) | ±3 mmHg (±0.4 kPa); ±5% of reading |
LCD Indication (Pressure; Pulse) | Yes; Yes |
Inflation | Automatic by Internal Air Pump |
Rapid Air Release | Automatic by Air Valve |
Power Source | Built-in 3.7V DC, 210mAh Li-ion Polymer Battery which should be charged by a charger of 5V DC, 500mA or more |
USB | USB 2.0 Port only for battery charge |
Bluetooth® | Bluetooth 4.0(data transmission), Frequency range:2.4GHz(2402 – 2480MHz),Modulation: GFSK, Effective radiated power: <20dBm |
Automatic turn off | within 2 minutes |
Weight | approx. 65 g |
Lifespan | 2 Years |
Operating conditions | 10 °C to 40 °C, 15 to 90 % relative humidity, non condensing, atmospheric pressure: 80-106 kPa |
Storage and transport conditions | – 20 °C to + 55 °C, 0 to 90 % relative humidity, non condensing |
Article No. | 51215 |
EAN number | 40 15588 51215 5 |
The device complies with IEC 60601-1:2005+A1:2012, IEC60601-1-2:2014, IEC60601-1-11:2015, IEC80601-2-30:2009+A1:2013, EN 1060-3: 1995+A2:2009
In accordance with our policy of continual product improvement, we reserve the right to make technical and visual changes without notice.
The current version of this instruction manual can be found under www.medisana.com
Warranty and repair termsPlease contact your dealer or the service centre in case of a claim under the warranty.If you have to return the unit, please enclose a copy of your receipt and state what the defect is.The following warranty terms apply:
- The warranty period for medisana products is three years from date of purchase. In case of a warranty claim, the date of purchase has to be proven by means of the sales receipt or invoice.
- Defects in material or workmanship will be removed free of charge within the warranty period.
- Repairs under warranty do not extend the warranty period either for the unit or for the replacement parts.
- The following is excluded under the warranty:a. All damage which has arisen due to improper treatment, e.g. non-observance of the user instructions.b. All damage which is due to repairs or tampering by the customer or unauthorised third parties.c. Damage which has arisen during transport from the manufacturer to the consumer or during transport to the service centre.d. Accessories which are subject to normal wear and tear.
- Liability for direct or indirect consequential losses caused by the unit are excluded even if the damage to the unit is accepted as a warranty claim.
imported & distributed bymedisana GmbHJagenbergstr. 1941468 NEUSSGERMANY
ShangHai International Holding Corp. GmbH (Europe)Address: Eiffestrasse 80, 20537 Hamburg Germany