Nex Inno Tech PC859 Medical Infrared Thermometer User Manual

Nex Inno Tech PC859 Medical Infrared Thermometer

Introduction

Thank you for choosing our medical infrared thermometer. The PC859/PC859EV medical infrared thermometer is used to measure human forehead temperature or object’s surface temperature based on the relationship between temperature and measurable infrared radiation. Simply aim the unit’s probe toward the surface to be measured to obtain a quick and accurate temperature.To ensure proper use this product, please be sure to read this user  manual carefully before using, paying close attention to the section2 ‘safety precautions’.Please keep the instructions on the side for easy checking.Models covered for this manual and their differences are as follows:

Model

Difference

PC859 At the end of temperature measurement, the buzzer gives a ‘beep’ cue.
PC859EV At the end of temperature measurement, the speaker broadcasts the temperature in English

Aside from the above differences, the expected use of the two models is exactly the same.

Basic principles:All objects above absolute zero temperature emits certain percentage of infrared radiation energy based on its temperature. The amount of the radiation energy and the distribution of the wavelength have very closely relationship. When human forehead’s temperature in 36-37°C, it emits wavelength 9-13um of infrared radiation . Based on this principle, according to the relationship between surface forehead temperature and human forehead’s temperature, we are able to measure the human forehead’s actual temperature through measuring surface forehead temperature.

Safety precautions

Warning

  • The infrared thermometer is a kind of body temperature screening device, use of this thermometer is not intended as a substitute for consultation with your physician. It is dangerous for user to perform a self-evaluation and self-treatment based on the measuring result.
  • Keep the thermometer out of reach of children. For accidental swallow of battery or other component, please consult the doctor at once.
  • Don’t throw the battery into fire.
  • Keep the thermometer out of direct sunlight in a dry, dust-free, well-ventilated area. Do not use the thermometer in the following cases:a) Beyond the specified humidity range 15-95% RH to operate;b) Beyond the specified temperature and humidity range to storage or shipping:Temperature: -25.0- 50.0’C (-13.0- 122.0′ F);Humidity: ≤ 95% RH, non-condensing. Otherwise the measured temperature may be inaccurate.
  • ASTM laboratory accuracy requirements in the display range of 37°C to 39’C (98’F to 102’F) for IR thermometers is ±0.2′ C (±0.4’F), whereas for mercury-in-glass and electronic thermometers, the requirement per ASTM Standards E667-86 and E 1112-86 is ±0.1 ‘ C(±0.2’F).Notice
  • The device is precision instrument, don’t drop, tramp or impose any vibration or impact on the thermometer.
  • Do not touch the lens of the probe with your fingers, and do not dissemble the device by yourself.
  • Before measuring forehead temperature, make sure the hair removed, sweat dried.
  • After you do some exercise, eating and bathing, you should stay still indoor about 30 minutes before measurement.
  • To make the measurement data reliable and stable, when ambient temperature varies a lot, the thermometer should be placed indoors for about 30 minutes before using.
  • When we measure somebody continuously, the temperature should be measured every minute, if you need to measure yourself continuously for a short time, there are some slight errors when you read the temperature, which is a normal phenomenon. At this time, we should choose the average. We recommend that you measure yourself continuously maximum of three in a unit of time, and because the temperature of the human will conduct to the thermometer, it may affect the accuracy of measurement.
  • There is no absolute standard about the temperature of the human, so please try to collect the recording of individual temperature in the usual, as a reference for having a fever or not.
  • Do not measure the sites of scarred tissue or tissue compromised by skin disorders, because sensing body temperature is not accurate from sites of scarred tissue or tissue compromised by skin disorders.
  • Do not measure forehead temperature if the patient has trauma on forehead.
  • Do not immerse the device into water or any other liquid, and not directly sunlight exposure.
  • Do not use a mobile phone or cordless phone near the thermometer when measuring.
  • During the therapeutic period of the drug, the body temperature may rise, please don’ t measure.
  • In order to ensure the accuracy of measurement data, please don’t take measurement of body temperature in strong electromagnetic interference environment (such as microwave, high frequency equipment operation environment).
  • Do not disassemble, repair, or modify the thermometer.
  • Do not touch the battery output when measuring.
  • The thermometer must be stored according to the technicalspecifications. The materials (ABS) of expect contact with patient had passed the ISO 10993-5 and ISO 10993-10 standard test, no toxicity, allergy and irritation reaction. Based on the current science and technology, other potential allergic reactions are unknown.
  • The patient is an intended operator. The patient can measure, read data and replace battery under normal circumstances and maintain the device and its accessories according to the user manual.
  • The device is not intended use for PATIENT transport outside a health care facility.Recommendations
  • Don’t use this thermometer for other purposes. It is forbidden to leave the product exposed to any chemical solvent, direct sunshine or high temperature.
  • Do not measure while talking on the phone.
  • Please report to MANUFACTURER if any unexpected operation or events occur

Intended use

This non-contact infrared thermometer is intended to measure forehead temperature at home or hospital, including anyone, such as infants, children and adults. Infants or children’s temperature must be measured by their parents or adults. It’ s intended used in the home healthcare environment.

Temperature measurement mode and range description 

The infrared thermometer has the following measurement mode:

  1. Forehead measurement mode – measure the skin surface to obtain the human forehead temperature accurately, take the place of traditional mercury thermometer and electrical thermometer.
  2. Object measurement mode – measure the object’s surface temperature, such as ambient, bathwater and milk temperature etc.

Normal temperature range for different measuring position

Measuring position

Normal temperature(°C)

Normal temperature(°F)

Anus 36.6-38.0 97.9-100.4
Oral 35.5-37.5 95.9-99 .5
Armpit 34.7-37.3 94.5-99.1
Forehead 35.8-38.0 96.4-100.4

Normal forehead temperature range for different ages

Ages

Normal temperature(°C)

Normal temperature(°F)

0-2 years old 36.4-38.0 97.5-100.4
3-1 0 years old 36. 1- 37.8 97.0-100.0
11- 65 years old 35.9-37.6 96.6-99.7
> 65 years old 35.8-37.5 96.4-99.5

NOTE:

  • The normal temperature and difference between the different body parts is individual. To define yours, measure your temperature for at least 2 weeks at the same forehead position and time.
  • When consulting your physician, communicate with him/her what measured site is important, note the individual’s normal temperature range as additional reference.
  • Because the forehead temperature is affected obviously by the external environment (e.g., environment, air convection and skin tone, etc.), we advice that you take the forehead temperature only as reference.

Feature

  • Precise non-contact measurement
  • Convertible °C and °F
  • Set warning temperature value
  • Automatic data hold and auto power save
  • Automatic selection range and display resolution 0.1 °C (0.l °F)
  • 3 colors backlight
  • Green LED indicates ready for use
  • Orange LED indicates low fever status
  • Red LED indicates a warning of high fever status
  • 2 measurement mode: Forehead and surface mode

Overall description

[Main component description]

  1. IR sensor
  2. LCD display
  3.  Buzzer Key
  4. Mode Key
  5. Memory Key
  6. Power/Scan button
  7.  Battery cover

[LCD display description]

  1. Surface mode
  2. Body mode
  3. Temperature value
  4. Low battery symbol
  5. buzzer
  6. °C(Celsius)/°F(Fahrenheit)
  7. Saved data number
  8. Memory symbol

Operation Instruction

[Preparation]

  1. Check battery: replace the batteries to ensure power supply is  enough if there is low voltage icon showed on the LCD screen.
  2. Check sensor: if have pollution and spray, clean the sensor before using, the cleaning method see the chapter 7 Care and Cleaning for detailed; If the sensor’ s lens is damaged, please stop using.
  3. Check thermometer: when you press the [Power/Scan] button, the system will have self testing of software and hardware. If there are problems, LCD will display ‘Err’ symbol.
  4. In order to obtain accurate measuring results, put the thermometer in the room temperature environment 30 minutes before using.
  5. Unexpected fluctuations of ambient temperature may decrease the measurement results accuracy. If the ambient temperature various a lot, or test temperature in front of the air conditioner, it will not be able to obtain accurate results.
  6. Before you measure forehead temperature, clean forehead, arrange hair and wipe sweat, make sure your forehead is clean and not be covered, in order to ensure the accuracy of measurement.

[Instruction for use]

1) Advice for the first useFor stable and reliable results, it is essential to check the infrared thermometer and changes as needed, as follows:

1st step: Take the temperature of a person using a conventional thermometer, you will get 37.5°C(99.5°f ) for instance.
2nd step: Take the temperature of the same person using the device king the 3 to 5cm distance between the thermometer and forehead ( e care to remove any obstacle which could alter the measurement (hair, perspiration … )

2) Use

  1. Install batteriesNotice Please take out the battery if the device is not used for a long period.● Please don’ t put the battery in the fire.● Please observe the related national laws of disposing the abandoned battery and don’t litter to the garbage can .To protect the environment, dispose of empty battery at your retail store or at appropriate collection sites according to national or local regulations.
  2. For the first use or when inserting new batteries, please wait about 10 minutes for the warm -up of the apparatus.
  3. If the device is not used for a long time, once you turn it on again, the device will test the room temperature first and will delay turning on for one or two seconds.
  4. Select body mode, aim towards the forehead , from distance of Scm (2in), press the measuring key, the temperature is displayed immediately, accompanied by a beep sound. Making sure there is no hair, perspiration, cosmetic or cap covered on the forehead.
  5. The cutoff temperature of low fever is 37.5ºC. If the temperature is higher than 37.5 ºC, the back light lit orange, please double check. We recommend measurement for 3 times to reference.
  6. The cutoff temperature of high fever is 38.0ºC. If the measurement temperature is higher than 38.0ºC, the sound should be “be-be-be-be-be” 5 rapid sounds and the back light lit red, please double check. We recommend measurement for 3 times to reference.

    TEMPERATURE TAKING HINTS

    Before take the temperature, please fix the forehead hair to prevent measured deviation.The sweat on head or cosmetic can affect the accuracy of measurement, please maintain the cleanness of the forehead when measuring.It is normal that there may be temperature difference depending on various skin types and color, since different skin type will reflect different voltage of infrared ray.Don’t use the thermometer outdoors.

  7. Select surface mode for measuring room temperature, surface temperature etc.
TEMPERATURE TAKING HINTS

For heat-insulated object, please don’t measure the surface.For the hot liquid, please do not directly measure the surface, because the hot steam fog can condense in the sensor lens and cause the measurement deviation.

3) Setting operation

Function Operation step
1 Set buzzer on/off Press  button to select the buzzer is on/off.
2 Measurement mode Press MODE key can select measurement modes . The are 2 measurement modes for this device. That is body and surface. It is able to take the body temperature of human begins, take the surface temperature of an area or an object, a food, a liquid or a room.
Body mode Press the MODE key, the screen will display “Body” on  the upper left, measurement range 32.0 to 43.0°C(89.6 to 109.4″F).
Surface mode Press the MODE key, the screen will display “Surface” on the upper left, measurement range 0 to 100°C(32.0 to 212.0°F ).
3 3 colors backlight In body mode, the device is able to judge whether a fever or not by different color of backlight:32.0 to 37.4°C (89.6 to 99.3°F )- Green37.5 to 37.9°C (99.5 to 100.2°F )- Orange38.0 to 42.9°C (100.4 to 109.4°F) – RedIn surface mode, the backlight always keeps green.
4 Data memory Date memory automatically after temperature measurements, which will display at the bottom of LCD. Press the “MEM” key to display the last recorded human body temperature.
5 °C and °F conversion Hold pressing  button, until °C/”F symbol appears on the screen, press MODE/MEM button to convert the using temperature unit to another one.

4) Automatic shutdown If you do nothing after the temperature measurement completed,   the thermometer will automatically shutdown within 20 seconds.

Cleaning and disinfection

The probe tip and lens are the most delicate and precise part of the thermometer. Please keep clean and entirety in order to assure the accuracy of readings

Clean the probe and lens as follows:

  • Use a cotton swab or soft cloth moistened with absolute alcohol to gently wipe the surface, and wait for the alcohol completely dry out.
  • If the lens is damaged, contact the distributor. Clean and disinfect the unit body:
  • Use a soft cloth moistened with clean water to wipe the thermometer display and unit body, and wait for the water completely dry out.
  • If the unit body is very dirty, it is need to be disinfected, use a soft cloth moistened with 75% medical alcohol to wipe the surface.NOTES:
  • Do not use abrasive cleaners.
  • Do not use other non-recommended methods to perform clean and disinfect.
  • Non-waterproof, do not drop the thermometer in the water or the other liquid.

Maintenance

  1. We do not authorize any agency or individual to maintain and repair the product. If you have any quality issues with the product, please contact the manufacturer or distributor for handling.
  2. Manufacturer will make available on request circuit diagrams, component part lists, descriptions, calibration instructions, etc., to assist to service personnel in parts repair.
  3. The users must not attempt to repair the device or its accessories, please contact the retailer for repair.
  4. Equipment disassembled by unauthorized agencies and individuals is not permitted and any warranty will be terminated in advance.
  5. Warning: Remove the batteries if the thermometer will not be used for more than one month.
  6. Warning: No servicing/ maintenance while the ME equipment is in use.
  7. Warning: No modification of this equipment is permitted!

Calibration

The thermometer was first calibrated at the time of manufacture, and if the user uses the product correctly according to the instructions, it does not need to be re-calibrated on a regular basis. If you have questions about the accuracy of the product measurement, you can contact the dealer or manufacturer at any time, the manufacturer information please see the last page.

Storage

  1. Don’t directly put the thermometer under the sunshine, high temperature and moist environment or someplace which maybe get in touch with fire or is vulnerable to vibration.
  2. Take out the battery if don’t use the device in a long time.

Packing list

Only use original accessories. Check that the contents of the delivery are complete.

Quantity Parts
1 pc PC859/PC859EVdevice
1 pc User’s Manual
1 pc Storage bag
2 pcs AAA battery

Trouble-shooting

Symptom Possible Cause Solution
No response/ Automatically reset The battery level is too low Replace the batteries.
The polarities of the batteries are reversed Ensure that the batteries are inserted correctly.
The thermometer is damaged Contact the dealer or manufacturer.
Hi
  1. First make sure that the robe lens is clean and free rom dirt.
  2. If the environmental temperature is out of range, then both patient and thermometer should stay in a room  between  10.0- 40.0°C (50.0 – 104.0 “F) for at least 30 minutes before taking a  temperature measurement.
  3. If problem persists after trying the above method , please contact the dealer or manufacturer.
Lo Human body temperature below 32 . 0°C/ 89.6 °F.
  1. First make sure that the probe lens is clean and free rom dirt.
  2. If the environmental temperature is out of range, then both patient and thermometer should stay in a room between 10.0- 40.0°C(50 . 0- 104.0 “F) for at least 30 minutes before taking a temperature measurement.
  3. If problem persists after trying the above method, please con tact the dealer or manufacturer.
Err Environmental temperature is not in the 10.0- 40.0°C (S0.0 – 104 . 0 °F) range .
  1. First make sure that the probe is clean and free from dirt.
  2. If the environmental temperature is out of range, then both patient and thermometer should stay in a room between 10.0 – 40 .0°C (50.0 – 104.0 “F) for at least 30 minutes before taking a temperature measurement.
  3. If problem persists after trying the above method , please contact the dealer or manufacturer .
Low battery, but you can still use it Keep an eye on power and continue to use.
Lower battery, however you can ‘t use it Replace the new battery .

Specifications

Product name Medical Infrared Thermometer
Product model PC859/PC859EV
Applicable regulations and laws ASTM E 1965 -98 / EN12470-5 / GB/T 19146
Power supply DC 3.0V (2 x AAA batteries)
Battery life More than 2,000 measurements
Low battery indicator Appears on the screen when the voltage is below 2.7V
Measurement modes Forehead & object surface temperature modes
Measurement time 1 second
Temperature units “C and “F (convertible)
Measurement range 32.0-43.0″C (89.6-109.4″F)
Accuracy (At laboratory conditions) 35.0- 42.0″C (95.0-107.6″F) ± 0.2″C/ 0.4″F
Other ranges ± 0.3″C/0S F
Clinical repeatability Within ± 0.3″C
Measuring site Forehead (keep distance 3-5cm fr om forehead)
Reference body site Human core
Auto shut down within 20 seconds
Display resolution 0.1″C/°F
Memory 99 saved readings
Operation environment Temperature : 10.0 – 40.0″C (50.0- 104 . 0″F)Humidity: 15- 95% RH , non-condensing Atmospheric pressure: 70-106kPa
Storage and shipping environment Temperature: -25.0- 50.0″C (-13.0- 122.0 “F)Humidity: 95% RH, non-condensingAtmospheric pressure : 70-106 kPa
Date of manufacture Sealable
Expected service life 5 years
Net weight 97 g (not including batteries)
Dimension 140 x 105 x 42mm
Grade of waterproof IP22
Electric shock Internally powered ME equipment
Software version B04
Note:

  1. Not intended to be sterilized;
  2. Not intended for use in an oxygen rich environment;
  3. Not intended for use in conjunction with flammable an aesthetics or flammable agents; 4. Surface temperature mode is direct mode, forehead temperature mode is adjust mode, surface temperature mode is test mode of forehead temperature mode

The above specifications are subject to change without prior notice.Shenzhen PACOM Medical Instruments Co., Ltd declares that the PC859/PC859EV complies with following applicable standards:

EN 980 Symbols for use in the labeling of medical devices
EN 1041 In formation supplied by the manufacturer with medical devices
EN 60601-1 Medical electrical equipment Part 1: General requirements for basic safety and essential performance
EN 60601-1-2 Medical electrical equipment — Part 1-2: General requirements for basic safe ty and essential performance – Collateral standard : Electromagnetic compatibility ­ Requirements and tests
I EN 60601-1-6 Medical electrical equipment – Part l -6: General requirements for basic safety and essential performance – Collateral standard: Usability
EN 60601 -1-11 Medical electrical equipment – Part 1-11: General requirements for basic safe ty and essential performance – Collateral standard: Requirements for  medical electrical equipment and medical electrical systems used in home healthcare environment
EN 12 470-5 Clinical thermometers – Part 5: Performance of infra-red ear thermometers (with maximum device)
ISO 80601-2-56 Medical electrical equipment part 2-56: particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement
EN 62304 Medical device software – Software life-cycle processes
EN 62366 Medical devices – Application of usability engineering g to medical devices

Disposal

Dispose of the device in accordance with the regulation applicable at the place of operation.Dispose of at public collection point in the EU countries –– 2002/ 96/ EC WEEE Directive.If you have any queries, please refer to the local authorities responsible for waste disposal.

NOTES:

  • Handling of battery and wastes method, please act according to the native law to proceed to handle. Take out the battery if you are not going to use the unit for a long time.

To protect the environment, dispose of empty battery at your retail store or at appropriate collection sites according to national or local regulations.Dispose of at public collection point in the EU countries – 2006/ 66/EC Directive.

Normalized symbols

Symbol Description
Disposal io accordance with Directive 2012/19/EU (WEEE)
Please read the instructions carefully.
Manufacturer
Date of manufacture
Serial number
Batch code
IP22 IP code of the device: this device’ s grade of against ingress of solid foreign objects — a;; 12.5mm diameter (and the against access to hazardous parts with finger); the grade of waterproof is drippings (150 tilted).

Appendix A: EMC Information – Guidance  and Manufacturer’s Declaration

CAUTION!

  • This Medical Infrared Thermometer requires special precautions regarding EMC, and must be installed and put into service according to the EMC information provided in the ACCOMPANYING DOCUMENTS.
  • Portable and mobile RF communications equipment can affect the thermometer.
  • This device should not be used adjacent to or stacked with other electric equipment.

Guidance and manufacturer’s declaration – electromagnetic emissions – for all EQUIPMENT and SYSTEMS

Guidance and manufacturer’s declaration – electromagnetic emissions
The Medical Infrared Thermometer is intended for use in the electromagnetic environment specified below. The customer or user of the device should ensure that it is only used in such an environment.
Emissions test Compliance Electromagnetic environment – guidance
RF emissions CISPR 11 Group I As this device only uses RF energy for its internal functions, its RF emissions are very low and not likely to cause any interference to nearby electronic equipment.
RF emissions CISPR 11 Class B The device is suitable for use in all establishments, including domestic establishments and those directly connected to the public low-voltage power supply network which supplies buildings for domestic purposes.

Guidance and manufacture’ s declaration – electromagnetic immunity – for all EQUIPMENT and SYSTEMS

Guidance and manufacturer’s declaration – electromagnetic immunity
The Medical Infrared Thermometer is intended for use in the electromagnetic environment specified below. The customer or user of the device should ensure that it is only used in such an environment.
Immunity test !EC 60601 test level Compliance level Electromagnetic environment – guidance
Electrostatic discharge (ESD)!EC 61000-4-2 ±6KY contact

±SKY air

±6KY contact±SKY air Floors should be wooden, concrete or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30%.
Power frequency (50/60 Hz) magnetic field!EC 6 I 000-4-8 3Nm 3 A /m Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment.

Guidance and manufacture’ s declaration – electromagnetic immunity – for all EQUIPMENT and SYSTEMS

Guidance and manufacturer ‘ s declaration – electromagnetic immunity
The Medical Infrared Thermometer is intended for use in the electromagnetic environment specified below. The customer or user of the device should ensure that it is only used in such an environment.
Immunity test IEC60601 test level Compliance level Electromagnetic environment – guidance
Radiated RF !EC 61000-4-3 3V/m 80MHz To 2.5 GHz

3V/m

Portable and mobile RF communications equipment should not be used closer to any part of the device.
Infrared Thermometer, including cables, than the recommended separation distance calculated by the equation applicable to the frequency of the transmitter. Recommended separation distance:Where P is the  maximum  output  power rating of the  transmitter in watts (W) according to the  transmitter  manufacturer, and d is the  recommended  separation distance in meters (m).  b  Field  strengths from mixed  RF transmitters,  as determined by an electromagnetic site survey,  a  should be less than the compliance level in each frequency range. b interference may occur in the vicinity of equipment marked with the following symbol:
Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.Note 2: These guidelines may not apply in all situations. Electromagnetism is affected by absorption and reflection from structures, objects and people.
Theoretically, the field strengths of such fixed transmitters as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radios, AM and FM radio signals and TV signals cannot be predicted with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. The field strength in the location in which the thermometer is to be used should be measured to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating the thermometer. Over the frequency range of 150 kHz to 80MHz the field strength should be less than 3 V/m.

Recommended separation distances between portable and mobile RF communications equipment and the EQUIPMENT or SYSTEM – for EQUIPMENT and SYSTEMS that are not LIFE-SUPPORTING

Recommended separation distances between portable and mobile RF communications equipment and the Medical Infrared Thermometer.
The Infrared Thermometer is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or user of the infrared thermometer can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the thermometer as recommended below, according to the maximum output power of the communications equipment.
 

Rated maximum output power of transmitterw

Separation distance according to frequency of transmitter m
150 kHz to 80 MHz

80 MHz to 800

800 MHz to 2.5 GHz
0.01 / 0.12 0.23
0.1 / 0.38 0.73
1 / 1.2 2.3
10 / 3.8 7.3
100 / 12 23
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.Note 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.Not e 2: These guidelines may not apply in all situations . Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.

Guarantee

This device is under warranty for one year from the date of acquisition. Applications for repairs should be presented during the warranty period. Damage caused by improper use does not fall under the scope of the warranty. Batteries and packaging are also not covered by the warranty.

Manufacturer information

report this ad

Shenzhen Pacom Medical Instruments Co., Ltd.Add: Zone B, 8th floor, Building B, No.5, Gongye 5th Road, Jiangbian Community, Songgang Street, Bao’ An District, Shenzhen City, Guangdong Province, ChinaTel: +86-755-32920339Email: [email protected]

[xyz-ips snippet=”download-snippet”]


Posted

in

by