OMRON 5 series Upper Arm Blood Pressure Monitor Instruction Manual

OMRON 5 series Upper Arm Blood Pressure Monitor

Introduction

Thank you for purchasing the OMRON BP7250 Blood Pressure Monitor. This monitor can be used by multiple users. This blood pressure monitor uses the oscillometric method of blood pressure measurement. This means this monitor detects your blood movement through your brachial artery and converts the movements into a digital reading.

Safety Instructions

This instruction manual provides you with important information about the OMRON BP7250 Blood Pressure Monitor. To ensure the safe and proper use of this monitor, READ and UNDERSTAND all of the safety and operating instructions. If you do not understand these instructions or have any questions, contact 1-800-634-4350 before attempting to use this monitor. For specific information about your own blood pressure, consult with your physician.

Intended Use

This device is a digital monitor intended for use in measuring blood pressure and pulse rate in adult patient population. The device detects the appearance of irregular heartbeats during measurement and gives a warning signal with readings.Environments of Use: HomePatient Population: Adult

Receiving and InspectionRemove this monitor and other components from the packaging and inspect for damage. If this monitor or any other components is damaged, DO NOT USE and contact1-800-634-4350.

Symbols GlossaryFor symbol information, visit: OmronHealthcare.com/symbols-glossary

Contraindications

  • This monitor is contraindicated for use in ambulatory environments.
  • This monitor is contraindicated for use on aircraft.

Important Safety Information

Read the Important Safety Information in this instruction manual before using this monitor. Follow this instruction manual thoroughly for your safety.Keep for future reference. For specific information about your own blood pressure, CONSULT WITH YOUR PHYSICIAN.

Warning:

  • DO NOT use this monitor on infants, toddlers, children or persons who cannot express themselves.
  • DO NOT adjust medication based on readings from this blood pressure monitor. Take medication as prescribed by your physician. ONLY a physician is qualified to diagnose and treat high blood pressure.
  • DO NOT use this monitor on an injured arm or an arm under medical treatment.
  • DO NOT apply the arm cuff on your arm while on an intravenous drip or blood transfusion.
  • DO NOT use this monitor in areas containing high frequency (HF) surgical equipment, magnetic resonance imaging (MRI) equipment, computerized tomography (CT) scanners. This may result in incorrect operation of the monitor and/or cause an inaccurate reading.
  • DO NOT use this monitor in oxygen rich environments or near flammable gas.
  • Consult with your physician before using this monitor if you have common arrhythmias such as atrial or ventricular premature beats or atrial fibrillation; arterial sclerosis; poor perfusion; diabetes; pregnancy; pre-eclampsia or renal disease. NOTE that any of these conditions in addition to patient motion, trembling, or shivering may affect the measurement reading.
  • NEVER diagnose or treat yourself based on your readings. ALWAYS consult with your physician.
  • To help avoid strangulation, keep the air tube and AC adapter cable away from infants, toddlers or children.
  • This product contains small parts that may cause a choking hazard if swallowed by infants, toddlers or children.

Data Transmission

  • This product emits radio frequencies (RF) in the 2.4 GHz band. DO NOT use this product in locations where RF is restricted, such as on an aircraft or in hospitals. Turn off the Bluetooth® feature in this monitor and remove batteries and/or unplug the AC adapter when in RF restricted areas. For further information on potential restrictions refer to documentation on the Bluetooth usage by the FCC.

AC Adapter (optional accessory) Handling and Usage

  • DO NOT use the AC adapter if this monitor or the AC adapter cable is damaged. If this monitor or the cable is damaged, turn off the power and unplug the AC adapter immediately.
  • Plug the AC adapter into the appropriate voltage outlet. DO NOT use in a multi-outlet plug.
  • NEVER plug in or unplug the AC adapter from the electric outlet with wet hands.
  • DO NOT disassemble or attempt to repair the AC adapter. Battery Handling and Usage
  • Keep batteries out of the reach of infants, toddlers or children.

Indicates a potentially hazardous situation which, if not avoided, may result in minor or moderate injury to the user or patient, or cause damage to the equipment or other property.

  • Stop using this monitor and consult with your physician if you experience skin irritation or discomfort.
  • Consult with your physician before using this monitor on an arm where intravascular access or therapy, or an arterio-venous (A-V) shunt, is present because of temporary interference to blood flow which could result in injury.
  • Consult with your physician before using this monitor if you have had a mastectomy.
  • Consult with your physician before using this monitor if you have severe blood flow problems or blood disorders as cuff inflation can cause bruising.
  • DO NOT take measurements more often than necessary because bruising, due to blood flow interference, may occur.
  • ONLY inflate the arm cuff when it is applied on your upper arm.
  • Remove the arm cuff if it does not start deflating during a measurement.
  • DO NOT use this monitor for any purpose other than measuring blood pressure.
  • During measurement, make sure that no mobile device or any other electrical device that emits electromagnetic fields is within 12 inches (30 cm) of this monitor. This may result in incorrect operation of the monitor and/or cause an inaccurate reading.
  • DO NOT disassemble or attempt to repair this monitor or other components. This may cause an inaccurate reading.
  • DO NOT use in a location where there is moisture or a risk of water splashing this monitor. This may damage this monitor.
  • DO NOT use this monitor in a moving vehicle such as in a car.
  • DO NOT drop or subject this monitor to strong shocks or vibrations.
  • DO NOT use this monitor in places with high or low humidity or high or low temperatures. Refer to section 10.
  • During measurement, observe the arm to ensure that the monitor is not causing prolonged impairment to blood circulation.
  • DO NOT use this monitor in high-use environments such as medical clinics or physician offices.
  • DO NOT use this monitor with other medical electrical (ME) equipment simultaneously. This may result in incorrect operation of the monitor and/or cause an inaccurate reading.
  • Avoid bathing, drinking alcohol or caffeine, smoking, exercising and eating for at least 30 minutes before taking a measurement.
  • Rest for at least 5 minutes before taking a measurement.
  • Remove tight-fitting, thick clothing from your arm while taking a measurement.
  • Remain still and DO NOT talk while taking a measurement.
  • ONLY use the arm cuff on persons whose arm circumference is within the specified range of the cuff.
  • Ensure that this monitor has acclimated to room temperature before taking a measurement. Taking a measurement after an extreme temperature change could lead to an inaccurate reading. OMRON recommends waiting for approximately 2 hours for the monitor to warm up or cool down when the monitor is used in an environment within the temperature specified as operating conditions after it is stored either at the maximum or at the minimum storage temperature. For additional information of operating and storage/transport temperature, refer to section 10.
  • DO NOT use this monitor after the durable period has ended. Refer to section 10.
  • DO NOT crease the arm cuff or the air tube excessively.
  • DO NOT fold or kink the air tube while taking a measurement. This may cause an injury by interrupting blood flow.
  • To unplug the air plug, pull on the plastic air plug at the base of the tube, not the tube itself.
  • ONLY use the AC adapter, arm cuff, batteries and accessories specified for this monitor. Use of unsupported AC adapters, arm cuffs and batteries may damage and/or may be hazardous to this monitor.
  • ONLY use the approved arm cuff for this monitor. Use of other arm cuffs may result in incorrect readings.
  • Inflating to a higher pressure than necessary may result in bruising of the arm where the cuff is applied.NOTE: refer to “If your systolic pressure is more than 210 mmHg” in section 3 of this instruction manual for additional information.

Data Transmission

  • DO NOT replace batteries or unplug the AC adapter while your readings are being transferred to your smart device. This may result in incorrect operation of this monitor and failure to transfer your blood pressure data.

AC Adapter (optional accessory) Handling and Usage

  • Fully insert the AC adapter into the outlet.
  • When unplugging the AC adapter from the outlet, be sure to safely pull from the AC adapter. DO NOT pull from the AC adapter cable. When handling the AC adapter cable:
  • DO NOT damage it.
  • DO NOT break it.
  • DO NOT tamper with it.
  • DO NOT forcibly bend or pull it.
  • DO NOT twist it.
  • DO NOT use it if it is gathered in a bundle.
  • DO NOT pinch it.
  • DO NOT place it under heavy objects.
  • Wipe any dust off of the AC adapter.
  • Unplug the AC adapter when not in use.
  • Unplug the AC adapter before cleaning this monitor.

Battery Handling and Usage

  • DO NOT insert batteries with their polarities incorrectly aligned.
  • ONLY use 4 “AA” alkaline or manganese batteries with this monitor.
  • DO NOT use other types of batteries.
  • DO NOT use new and used batteries together.
  • DO NOT use different brands of batteries together.
  • Remove batteries if this monitor will not be used for a long period of time.
  • If battery fluid should get in your eyes, immediately rinse with plenty of clean water. Consult with your physician immediately.
  • If battery fluid should get on your skin, wash your skin immediately with plenty of clean, lukewarm water. If irritation, injury or pain persists, consult with your physician.
  • DO NOT use batteries after their expiration date.
  • Periodically check batteries to ensure they are in good working condition.

Know Your Monitor

Contents/Product IncludesMonitor, 4 “AA” batteries, Wide-Range D-Ring Cuff, instruction manual, quick start guideMonitor and Arm CuffFront of the monitor

  • Display
  • [Forward/Backward] buttons [Connection] button
  • (Date/Time)
  • [Memory] button
  • [START/STOP] button

  • Air jack
  • AC adapter jack (for optional AC adapter)
  • Battery compartment
Arm cuff

(Upper arm circumference: 9″ – 17″ (22 – 42 cm))

Display and Symbols
    Memory symbol

    Appears when viewing readings stored in the memory.

      Average value symbol

      Flashes when viewing the average of the latest 2 or 3 readings taken within the most recent 10 minute timeframe.

         

        Sync symbol

        Flashes/appears when your data needs to be transferred because the stored memory is either almost, or completely full. Once you pair your monitor with your smart device, transfer your readings immediately before the monitor deletes the oldest reading. Up to 60 readings can be stored in the internal memory of your monitor.

        Systolic blood pressure reading
        Diastolic blood pressure reading
          Deflation symbol

          Appears during cuff deflation.

            Heartbeat symbol

            Flashes while taking a measurement.

            Date/time display
              Bluetooth ON symbol

              Appears when your readings are being transferred.

                Bluetooth OFF symbol

                Appears when Bluetooth is disabled.

                  High symbol

                  Appears if your systolic blood pressure is ≥ 130 mmHg or the diastolic blood pressure is ≥ 80 mmHg. Refer to sub- section 1.4 for more information.

                  • or

                   

                   

                   

                   

                   

                   

                  Irregular heartbeat symbol

                  Appears when irregular rhythm is detected 2 or more times during a measurement. The symbol with the number appears as below:

                  Irregular heartbeat is detected 2 times:

                  Irregular heartbeat is detected 3 times or more:

                   

                  An irregular heartbeat rhythm is defined as a rhythm that is 25% less or 25% more than the average rhythm detected while your monitor is measuring blood pressure.

                  Short Long

                  Irregular Heartbeat Normal Heartbeat

                  Irregular heartbeat symbol may be displayed when the monitor detects arm movements. Refer to section 7 for more information.

                  If it continues to appear, we recommend you to consult with and follow the directions of your physician.

                     

                    Movement error symbol

                    Appears with your reading when your body is moving during a measurement. If it appears, remove the arm cuff and wait for 2 to 3 minutes. Then, reapply the cuff, remain still and take another measurement.

                      OK symbol

                      Flashes when your monitor is connected to your smart device or readings are transferred successfully.

                         Low battery symbol

                      Flashes when batteries are low.

                        Depleted battery symbol

                        Appears when batteries are depleted.

                        Pulse display / Memory number

                        Pulse rate appears after the measurement.

                        When pressing the  button, the memory number will also appear for approximately 1.5 seconds before the pulse rate shows again on the display.

                        Preparation

                        1. Push down the hook of the battery cover and pull downward.
                        2. Insert 4 “AA” batteries as indicated in the battery compartment.
                        3. Close the battery cover. As soon as batteries are correctly installed, the year flashes on the display. Refer to sub-section 2.2 to set the date and time manually, or refer to sub-section 4.1 to pair your monitor to set the date and time automatically.

                        Note

                        • When the “ ” symbol flashes on the display, it is recommended you replace batteries.
                        • To replace batteries, turn your monitor off and remove all batteries. Then replace with 4 new batteries at the same time.
                        • Replacing batteries will not delete previous readings.
                        • The supplied batteries may have a shorter life span than new batteries.
                        • Disposal of used batteries should be carried out in accordance with local regulations.

                        1. If nothing is displayed, press the button to display the year.If the year is already displayed, skip this step.
                        2. Press the or button to adjust the year. Press and hold the button to quickly advance the years.Press and hold the button to quickly go backwards through the years.
                        3. Press the button to confirm the year, then the month will flash. Repeat the same steps to adjust the month, day, hour, and minute.
                        4. Press the button to store the setting. Your monitor automatically turns off 3 seconds after the minutes have been set in this sequence.

                        Blood Pressure Measurement Tips

                        To help ensure an accurate reading, follow these directions:

                        • Stress raises blood pressure. Avoid taking measurements during stressful times.
                        • Measurements should be taken in a quiet place.
                        • It is important to take measurements at the same times each day. Taking measurements in the morning and in the evening is recommended.
                        • Remember to have a record of your blood pressure and pulse readings for your physician. A single measurement does not provide an accurate indication of your true blood pressure. You need to take and record several readings over a period of time.

                        Caution

                        • Avoid bathing, drinking alcohol or caffeine, smoking, exercising and eating for at least 30 minutes before taking a measurement.
                        • Rest for at least 5 minutes before taking a measurement.
                        Applying the Arm Cuff
                        • The following steps are for applying the arm cuff to your left arm. When you take a measurement on your right arm, follow the cuff wrap instructions for a right arm use mentioned under “To take a measurement on your right arm” located at the end of this sub-section.
                        • The blood pressure can differ between the right arm and the left arm, and the measured blood pressure readings can be different. OMRON recommends to always use the same arm for measurement. If the readings between both arms differ substantially, check with your physician to determine as to which arm to use for your measurements.

                        Caution: Remove tight-fitting, thick clothing from your arm while taking a measurement.

                        1. Plug the arm cuff into your monitor by inserting the air plug into the air jack securely until it clicks.
                        2. Place your hand through the cuff loop. Pull the cuff until it reaches your upper left arm.
                        3. Make sure that the air tube is positioned on the inside of your arm and wrap the cuff securely, so it can not move around your arm.

                        To take a measurement on your right armWhen you take a measurement on your right arm, the air tube should run along the side of your elbow, along the bottom of your arm. Be careful not to rest your arm on the air tube.

                        Sitting Correctly

                        To take a measurement, you need to be relaxed and comfortably seated in a room at a comfortable temperature. Place your arm on the table.

                        • Sit in a comfortable chair with your back and arm supported.
                        • Keep your feet flat and your legs uncrossed.
                        • The arm cuff should be placed on your arm at the same level as your heart, with the arm resting comfortably on a table.

                        Using Your Monitor without a Smart Device

                        Taking a MeasurementYou can use your monitor without pairing to a smart device. To pair your monitor with a smart device, refer to sub-section 4.1.

                        Note: To stop the measurement, press the [START/STOP] button once to deflate the arm cuff.

                        • DO NOT use this monitor with other medical electrical (ME) equipment simultaneously. This may result in incorrect operation of the monitor and/ or cause an inaccurate reading.
                        • Remain still and DO NOT talk while taking a measurement.
                        1. Press the [START/STOP] button. All symbols appear on the display before starting the measurement.
                        2. Remain still and do not talk until the entire measurement process is complete.The “ ” symbol flashes at every heartbeat. After your monitor has detected your blood pressure and pulse rate, the cuff automatically deflates. Your blood pressure and pulse rate are displayed.If either the systolic or the diastolic reading is high (refer to subsection 1.3), the “ ” symbol appears.
                        3. Press the [START/STOP] button to turn your monitor off.

                        Note

                        • Your monitor automatically turns off after 2 minutes.
                        • Wait 2 – 3 minutes between measurements. The wait time allows the arteries to decompress and return to their pre-measurement state.
                        • You may need to increase the wait time depending on your individual physiological characteristics.

                        Error Messages and Troubleshooting

                        If any of the below problems occur during measurement, check to make sure that no other electrical device is within 12 inches (30 cm). If the problem persists, refer to the table below.

                        Display/Problem Possible Cause Solution
                         

                         

                         

                        appears or the arm cuff does not inflate.

                        Air plug is not completely plugged into the monitor. Insert the air plug securely. Refer to sub- section 2.4.
                         

                        The arm cuff is not applied correctly.

                        Apply the arm cuff correctly, then take another measurement. Refer to sub-section 2.4.
                        Air is leaking from the arm cuff. Replace the arm cuff with a new one. Refer to section 9.
                         

                         

                         

                         

                        appears or a measurement cannot be complete after the arm cuff inflates.

                         

                         

                         

                        You move or talk during a measurement and the arm cuff does not inflate sufficiently.

                        Remain still and do not talk during a measurement. If “E2” appears repeatedly, inflate the arm cuff manually until the systolic pressure is 30 to 40 mmHg above your previous readings. Refer to sub-section 3.1.
                         

                        The systolic pressure is above 210 mmHg and a measurement cannot be taken.

                        Inflate the arm cuff manually until the systolic pressure is 30 to 40 mmHg above your previous readings. Refer to sub-section 3.1.
                         

                         

                        appears

                         

                         

                        The arm cuff is inflated exceeding the maximum allowable pressure.

                        Do not touch the arm cuff and/or bend the air tube while taking a measurement.

                        If inflating the arm cuff manually, refer to the end of sub-section 3.1.

                         

                         

                        appears

                        You move or talk during a measurement.

                        Vibrations disrupt a measurement.

                        Remain still and do not talk during a measurement.
                        Display/Problem Possible Cause Solution
                         

                         

                        appears

                         

                         

                         

                         

                        The pulse rate is not detected correctly.

                        Apply the arm cuff correctly, then take another measurement. Refer to sub-section 2.4. Remain still and sit correctly during a measurement.

                        If the “              ” symbol continues to appear, we recommend you to consult with your physician.

                         / appears
                         

                        does not flash during a measurement

                         

                         

                        appears

                         

                        The monitor has malfunctioned.

                        Press the [START/ STOP] button again. If “Er” still appears, contact customer service. Refer to section 12.
                         

                         

                        appears

                         

                         

                        The monitor cannot connect to a smart device or transmit data correctly.

                        Follow the instructions shown in the “OMRON connect US/CAN” app. If the “Err” symbol still appears after checking the app, contact customer service. Refer to section 12.
                         

                        flashes

                        The monitor is waiting for pairing with the smart device. Refer to sub-section 4.1 for pairing your monitor with your smart device.
                         

                        flashes

                        The monitor is ready to transfer your readings to the smart device. Open the “OMRON connect US/CAN” app to transfer your readings.
                         

                        flashes

                        There are more than 40 readings in memory to be transferred.  

                        Pair or transfer your readings to the “OMRON connect US/CAN” app so you can keep them in memory in the app, and this error symbol disappears.

                        The date and time is not set.
                         

                        appears

                        There are 60 readings in memory to be transferred.
                         

                        flashes

                         

                        Batteries are low.

                        Replacing all batteries with 4 new batteries is recommended. Refer to sub-section 2.1.
                         

                        appears or the monitor is turned off unexpectedly during a measurement.

                         

                         

                        Batteries are depleted.

                         

                        Immediately replace all batteries with 4 new batteries. Refer to sub- section 2.1.

                         

                        Nothing appears on the display of the monitor.

                        Batteries are depleted.
                         

                        Battery polarities are not properly aligned.

                        Check the battery installation for proper placement. Refer to sub- section 2.1.
                         

                        Readings appear too high or too low.

                        Blood pressure varies constantly. Many factors including stress, time of day, and/or how you apply the arm cuff, may affect your blood pressure.

                        Review sub-sections 2.3 – 2.5 and section 3.

                         

                        Any other communication issue occurs.

                        Follow the instructions shown in the smart device, or visit the “Support” section in the “OMRON connect US/CAN” app for further help. If the problem still persists, contact customer service. Refer to section 12.
                         

                         

                        Any other problems occur.

                        Press the [START/STOP] button to turn the monitor off, then press it again to take a measurement. If the problem continues, remove all batteries and wait for 30 seconds. Then re-install batteries.

                        If the problem still persists, contact customer service. Refer to section 12.

                        FCC Statement

                        Changes or modifications not expressly approved by the party responsible for compliance could void the user’s authority to operate the equipment.This device complies with part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) This device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation.

                        Note

                        • This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. This equipment generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures:
                        • Reorient or relocate the receiving antenna.
                        • Increase the separation between the equipment and receiver.
                        • Connect the equipment into an outlet on a circuit different from that to which the receiver is connected.
                        • Consult the dealer or an experienced radio/TV technician for help.

                        This transmitter must not be co-located or operated in conjunction with any other antenna or transmitter.This equipment complies with FCC radiation exposure limits set forth for an uncontrolled environment and meets the FCC radio frequency (RF) Exposure Guidelines. This equipment has very low levels of RF energy that are deemed to comply without testing of specific absorption ratio (SAR).

                        Limited Warranty

                        Your BP7250 Blood Pressure Monitor, excluding the arm cuff and batteries is warranted to be free from defects in materials and workmanship appearing within 2 years from the date of purchase, when used in accordance with the instructions provided with the monitor. The arm cuff is warranted to be free from defects in materials and workmanship appearing within 1 year from the date of purchase when the monitor is used in accordance with the instructions provided with the monitor. The above warranty extends only to the original retail purchaser.

                        We will, at our option, replace without charge any monitor or arm cuff covered by the above warranty. Replacement is our only responsibility and your only remedy under the above warranty.To obtain warranty service contact Customer Service by calling 1-800-634-4350 for the address of the inspection center and the return shipping and handling fee.Enclose the original printed receipt. Include a letter, with your name, address, phone number, and description of the specific problem. Pack the product carefully to prevent damage in transit. Because of possible loss in transit, we recommend insuring the product with return receipt requested.

                        THE FOREGOING IS THE SOLE WARRANTY PROVIDED BY OMRON IN CONNECTION WITH THIS PRODUCT, AND OMRON HEREBY DISCLAIMS ANY OTHER WARRANTIES, EXPRESS OR IMPLIED, INCLUDING IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE. IMPLIED WARRANTIES AND OTHER TERMS THAT MAY BE IMPOSED BY LAW, IF ANY, ARE LIMITED IN DURATION TO THE PERIOD OF THE ABOVE EXPRESS WARRANTY.

                        OMRON SHALL NOT BE LIABLE FOR LOSS OF USE OR ANY OTHER SPECIAL, INCIDENTAL, CONSEQUENTIAL OR INDIRECT COSTS, EXPENSES OR DAMAGES.

                        This warranty provides you with specific legal rights, and you may have other rights that vary by jurisdiction. Because of special local requirements, some of the above limitations and exclusions may not apply to you.

                        OMRON HEALTHCARE Co., Ltd53, Kunotsubo, Terado-cho, Muko, Kyoto, 617-0002 JAPAN

                        Distributed by:OMRON HEALTHCARE, INC.1925 West Field Court Lake Forest, IL 60045 U.S.A.OmronHealthcare.com© 2019 OMRON HEALTHCARE, INC.Made in Vietnam