The H5i is indicated for the humidification of the air delivered from a CPAP or bilevel device. The H5i is for use only as recommended by a physician. The H5i is intended for single patient re-use in the home environment and re-use in a hospital/institutional environment.
Contraindications
The H5i is contraindicated for use with patients whose upper (supraglottic) airway has been bypassed. Please refer to the S9 Series Information Guide for contraindications associated with CPAP or bilevel therapy.
Troubleshooting
If there is a problem, try the following suggestions. If the problem cannot be solved, contact your equipment supplier or ResMed. Do not attempt to open these devices.
Problem/possible cause Solution
No displayPower is not connected. Ensure the power cable is connected and the power outlet (if available) is on.The DC plug is partially inserted into the back of the device. Fully insert the DC plug.The S9 and H5i are not connected correctly. Ensure that the S9 and H5i are securely attached.Insufficient air delivered from the deviceThe S9 is not working correctly. See the Troubleshooting section of the S9 Series Information Guide.Air tubing is not connected properly. Check air tubing.Air tubing is blocked, pinched or punctured. Unblock or free the air tubing. Check the air tubing for punctures.The H5i flip lid is not latched correctly. Close the flip lid ensuring that it clicks into place.The H5i flip lid seal is not fitted correctly. Make sure the flip lid seal is facing the right way up and fitted securely.Incorrect air tubing selected. If you are using the SlimLine or Standard air tubing ensure that you have the correct air tubing selected via the menu.Device has been left in a hot environment. There is a fault in your ClimateLine or ClimateLineMAX air tubing.Allow to cool before re-use. Disconnect the power cord and then reconnect it to restart the device. Discontinue using your ClimateLine or ClimateLineMAX air tubing and contact your clinician/service provider. Use Standard or SlimLine tubing in the interim.
There is a fault in your H5i heated humidifier. Device has been left in a hot environment. Refilling the humidifier with cold water while humidifier is still hot after therapy. Filling the humidifier with ice cold water on a warm day or with hot water.
Discontinue using your H5i and contact your clinician/ service provider. Allow to cool before re-use. Disconnect the power cord and then reconnect it to restart the device. Allow the humidifier to cool before re-filling. Use room temperature water.
Displays error message: Tube blocked, please check your tubeAir tubing is blocked. Check your air tubing and remove any blockages
Displays error message: High leak, please check system setup and all connectionsThere is excessive leak. Adjust position of mask and headgear. Air tubing is not connected properly. Connect firmly at both ends.
The H5i flip lid is not latched correctly. Close the flip lid ensuring that it clicks into place.
Displays error message: Please close H5i flip lid, attach tube and press any key
H5i flip lid is not closed. Close the flip lid ensuring that it clicks into place.Air tubing is not connected properly. Connect firmly at both ends.
Water splashing on your face from the H5iThe water tub is overfilled. Check that the water level is below the maximum water level markCondensation is forming in the air tube and mask. Turn the H5i or relative humidity setting down via the menu.
Leaking water tub
The water tub may be damaged or cracked. The cleanable water tub is not assembled correctly.Contact your service provider for a replacement. Check for damage and reassemble the cleanable water tub correctly.
Air feels too warm/cold in the mask
The temperature of the Climate Lineor Climate Line MAX air tubing is set too high/low. Turn up/down the heated tubing temperature via the menu.
Technical specifications
Maximum heater plate temperature 65°CTemperature cut-out 74°CMaximum gas temperature £ 41°CNominal dimensions (L x W x H) Docking station and water tub: 153 mm x 145 mm x 86 mmWeight (Water tub) Docking station and unfilled water tub 0.67 kgWeight (Cleanable water tub) Docking station and unfilled water tub 0.77 kgWater capacity To maximum fill line 380 mLDocking station Flame retardant engineering thermoplastic, aluminiumCleanable water tub Injection molded plastic, stainless steel and silicone sealWater tub Injection molded plastic, aluminium and thermoplastic elastomerOperating temperature +5°C to +35°COperating humidity 10–95% non-condensingStorage and transport temperature -20°C to +60°CStorage and transport humidity 10–95% non-condensingClimateLine air tubing Flexible plastic and electrical components,1.98 m, 15 mm inner diameterClimateLineMAX air tubing Flexible plastic and electrical components,1.91 m, 19 mm inner diameterHeated tubing temperature cut-out £ 41°CIEC 60601-1 classification Class II (double insulation), Type BFElectromagnetic compatibility Please refer to the S9 Series Information Guide for details
Notes:
The manufacturer reserves the right to change these specifications without
The temperature and relative humidity settings displayed for ClimateLine or ClimateLineMAX are not measured
Check with your clinician/service provider before using the SlimLine air tubing with devices other than the S9 or H5i.
Humidifier performance
The following settings have been tested at 22°C ambient temperature:
CPAP mask RH output % Nominal system output AHa, BTPSb
pressure, cm H2O
Setting 3
Setting 6
Setting 3
Setting 6
4
90
100
10
18
10
95
100
11.5
21
20
95
100
11
18
25
100
100
12
13.5
a. AH – Absolute Humidity in mg/L.b. BTPS – Body Temperature Pressure Saturated.
Symbols
The following symbols may appear on your H5i, Climate Line or packaging.
Caution;Follow instructions for use; Manufacturer; IP21Protection against insertion of fingers and against vertically dripping water; IP20 Not drip proof;European Authorised Representative;European RoHS;Batch code; Catalogue number;Serial number;Not for use on more than one patient; Keep dry;Maximum water level; Disinfectable up to 93°C; China pollution control logo1; China pollution control logo 2;Lock/unlock;Remove tub to fill;
Environmental information
WEEE 2002/96/EC is a European Directive that requires the proper disposal of electrical and electronic equipment. These devices should be disposed of separately, not as unsorted municipal waste. To dispose of your device, you should use appropriate collection, reuse and recycling systems available in your region. The use of these collection, reuse and recycling systems is designed to reduce pressure on natural resources and prevent hazardous substances from damaging the environment.If you need information on these disposal systems, please contact your local waste administration. The crossed-bin symbol invites you to use these disposal systems. If you require information on collection and disposal of your ResMed device please contact your ResMed office, local distributor or go to www.resmed.com/environment.
Servicing
The H5i device is intended to provide safe and reliable operation when operated in accordance with the instructions provided by ResMed. ResMed recommends that the H5i be inspected and serviced by an authorised ResMed Service Centre if there is any sign of wear or concern with device function. Otherwise, service and inspection of the devices generally should not be required during the five year design life of the device.
Limited warranty
ResMed Pty Ltd (hereafter ’ResMed’) warrants that your ResMed product shall be free from defects in material and workmanship from the date of purchase for the period specified below.
Product Warranty period
Mask systems (including mask frame, cushion, headgear and tubing) excluding single-use devices
Accessories— 90 daysexcluding single-use devices
Flex-type finger pulse sensors
Humidifier water tubs
Batteries for use in ResMed internal and external battery systems 6 months
Clip-type finger pulse sensors 1 year
CPAP and bilevel device data modules
Oximeters and CPAP and bilevel device oximeter adapters
Humidifiers and humidifier cleanable water tubs
Titration control devices
CPAP, bilevel and ventilation devices (including power supply units) 2 years
Battery accessories
Portable diagnostic/screening devices
This warranty is only available to the initial consumer. It is not transferable
If the product fails under conditions of normal use, ResMed will repair or replace, at its option, the defective product or any of its components. This limited warranty does not cover: a) any damage caused as a result of improper use, abuse, modification or alteration of the product; b) repairs carried out by any service organisation that has not been expressly authorised by ResMed to perform such repairs; c) any damage or contamination due to cigarette, pipe, cigar or other smoke; and d) any damage caused by water being spilled on or into an electronic device.Warranty is void on product sold, or resold, outside the region of original purchase.Warranty claims on defective product must be made by the initial consumer at the point of purchase. This warranty replaces all other expressed or implied warranties, including any implied warranty of merchantability or fitness for a particular purpose. Some regions or states do not allow limitations on how long an implied warranty lasts, so the above limitation may not apply to you.ResMed shall not be responsible for any incidental or consequential damages claimed to have resulted from the sale, installation or use of any ResMed product. Some regions or states do not allow the exclusion or limitation of incidental or consequential damages, so the above limitation may not apply to you.This warranty gives you specific legal rights, and you may also have other rights which vary from region to region. For further information on your warranty rights, contact your local ResMed dealer or ResMed office.
WARNINGS
Read the entire manual before using the
Use the device only as directed by your physician or healthcare
Use the device only for the intended use as described in this manual. Advice contained in this manual should not supersede instructions given by the prescribing
If you notice any unexplained changes in the performance of the device, if it is making unusual or harsh sounds, if the device or the power supply are dropped or mishandled, if water is spilled into the enclosure, or if the enclosure is broken, discontinue use and contact your ResMed Service
Beware of Do not immerse the device, humidifier, power supply or power cord in water. In the event of a spill, disconnect the device from the power supply and let the parts dry. Always unplug the device before cleaning and make sure that all parts are dry before plugging in the device.
Explosion hazard—do not use in the vicinity of flammable anesthetics.
Make sure the power cord and plug are in good condition and the equipment is not
Keep the power cord away from hot
Do not operate the H5i if it is not working properly or if any part of the device or H5i has been dropped or
Do not leave long lengths of air tubing around the top of your It could twist around your head or neck while you are sleeping.
Only ResMed air tubing and accessories should be used with the A different type of air tubing or accessory may alter the pressure you actually receive, reducing the effectiveness of the treatment.
Only use the ResMed 90W power supply
CAUTIONS
Do not open the device There are no user serviceable parts inside. Repairs and servicing should only be performed by an authorised ResMed service agent.
Do not use bleach, chlorine, alcohol, or aromatic-based solutions, moisturising or antibacterial soaps or scented oils to clean the device, humidifier or air These solutions may cause damage and reduce the life of these products.
The H5i should only be used with tubing or accessories recommended by ResMed. Connection of other delivery tubes or accessories could result in injury, or damage to the
Do not open the H5i enclosure. There are no user serviceable parts inside. Repairs and servicing should only be performed by an authorised ResMed service
Do not overfill the water tub as water may enter the device and air
Do not use any additives (eg, scented oils and perfumes). These may reduce the humidification output of the H5i and/or cause deterioration of the water tub
Take care when handling the H5i as the water/water tub may be Allow 10 minutes for the heater plate and any excess water to cool.
The H5i should only be connected or disconnected when the water tub is
Make sure that the water tub is empty before transporting the
Do not operate the H5i on an aircraft as water may enter the device and air tubing during
Always place the H5i on a level surface below the level of the user to prevent the mask and tubing from filling with
If liquids are inadvertently spilled into or on the H5i, unplug the device from the power Disconnect the H5i from the device and allow it to drain and dry before re-using.