RND 320-KA Series single Channel Remote Control Syntax V3.0
Command format: VSET<X>:<NR2>
- VSET: Command header
- X: output channel
- separator
- NR2: parameter
Command Details:
- ISET<X>:<NR2>Description: Sets the output current.Example: ISET1 :2.225Sets the CH 1 output current to 2.225A
- ISET<X>?Description: Returns the output current setting.Example: ISET1?Returns the CH 1 output current setting
- VSET<X>:<NR2>Description: Sets the output voltage. Example:VSET1:20.50Sets the CH 1 voltage to 20.S0V
- VSET<X>?Description: Returns the output voltage setting.Example: VSET1?Returns the CH1 voltage setting.
- IOUT<X>?Description: Returns the actual output current.Example: IOUT1?Returns the CH1 output current
- VOUT<X>?Description: Returns the actual output voltage.Example: VOUT1?Returns the CH 1 output voltage
- BEEP<Boolean>Description: Turns on or off the beep. Boolean: boolean logic.Example: BEEP1 Turns on the beep.
- OUT<Boolean>Description: Turns on-off the output.Boolean: 0 OFF, 1 ONExample: OUT1 Turns on the output
- STATUS?Description: Returns the POWER SUPPLY status.Contents 8 bits in the following formatBit Item Description0 When output is ON,0=CC mode, 1 =CVmode4 Beep 0=Off, 1 =On5 OCP 0=Off, 1 =On6 Output 0=Off, 1 =On7 OVP 0=Off, 1 =On
- * IDN?Description: Returns the RND 320-KA3005P identification.Example *IDN?Contents RND 320-KA3005P V1 .3 (Manufacturer, model name,).
- RCL<NR1>Description: Recalls a panel setting.NR 1 1-5: Memory numbers 1 to 5Example RCL 1 Recalls the panel setting stored in memory number1
- SAV<NR1>Description: Stores the panel setting.NR 1 1-5: Memory numbers 1 to 5Example SAV1 Stores the panel setting stored in memory number1
- OCP< Boolean >Description: Stores the panel setting.Boolean: 0 OFF, 1 ONExample: OCP1 Turns on the OCP
- OVP< Boolean >Description: Turns on the OVP.Boolean: 0 OFF, 1 ONExample: OVP1 Turns on the OVP
[xyz-ips snippet=”download-snippet”]