SENSECAP Soil Moisture Temperature EC Sensor User Manual

SENSECAP Soil Moisture Temperature EC Sensor User Manual

1 Introduction

This S-Temp&VWC&EC-02 soil moisture & EC & temperature sensor is provided with high accurate and high sensitive. It is an important tool to observe and study the occurrence, evolution, improvement and the dynamics water of saline soil. By measuring the dielectric constant of the reaction of soil, soil direct stable real moisture content. This S-Temp&VWC&EC-02 sensor can measure the volume of soil moisture. The soil moisture measurement method is in line with international standards at present. Apply to the soil moisture monitoring, scientific experiment, water-saving irrigation, greenhouse vegetables, flowers, grass, soil, plant cultivation, measured speed of sewage treatment, grain storage, greenhouse control, precision agriculture.

Features:

  • Soil moisture content, electrical conductivity and temperature all in one
  • One solution can also be used for fertilizer, and other nutrient solution conductivity matrix
  • Electrode using special treatment of the alloy material, can withstand a strong external impact, not easy to damage
  • Completely sealed, acid and alkali corrosion, can be buried in the soil or directly into the water for long-term dynamic testing
  • High precision, fast response, good compatibility, the probe insert design to ensure accurate measurement, reliable performance
  • Perfect protection circuit

2 Specifications

  Soil Temperature
  Range   -40 ℃ to +80 ℃
  Accuracy   ±0.5℃
  Resolution   0.1 ℃
  Soil Moisture
  Range   From completely dry to fully saturated (from 0% to 100% of saturation)
  Accuracy   ±2% (0~50%); ±3% (50~100%)
  Resolution   0.03% (0~50%); 1% (50~100%)
  Electrical Conductivity
  Range   0 ~ 10000 μs/cm
  Accuracy   ±3%
  Resolution   10 μs/cm
  Temperature Compensation   Built in temperature compensation sensor, range 0-50℃
General Parameters
  Product Model   S-Temp&VWC&EC-02
  Interface   RS-485
  Protocol   MODBUS-RTU RS485
  Power Supply   3.6 ~ 30V DC
  Current Consumption   DC (quiescent dissipation)
  IP Rating   IP68
  Cable Length   5 meters
Operating Temperature -40 ~ 85℃
  The material of the probe   Anti-corrosion special electrode
  Sealing material   The black flame retardant epoxy resin
  Installation   All embedded or probe inserted into the measured medium
  Device Weight   210g

3 Wiring

  Yellow   RS485+/A/T+
  White   RS485-/B/T-
  Red   VCC+, power supply
  Black   VCC-, power ground
  Green    SET, V+ (power) when boot module into the “setting mode”. Not connected or connected with the     GND when boot into “mode of operation”.

Module configuration parameters such as Modbus address, baud rate, parity, communication                protocol is composed of module inside the EEPROM (power down storage device stores). The            specific configuration sometimes forget these parameters that cannot communicate with the                module. In order to prevent this problem, the module has a special mode called “mode”. When the      module is based on the “mode” electric start, the module communicates with the following                  parameters:  Fixed Modbus address  Communication configuration is 9600, N, 8,1 (9600bps, no          parity bit, 8 data bits, a stop bit) Communication protocol for Modbus-RTU Configuration                  parameters in EEPROM will not because the module into the

“mode” and “will change, when the module is in communication with in the EEPROM                        configuration   parameter is still running mode”.

4 Installation

Because of the direct determination of the soluble salt ions in the soil, the water content of the soil can be higher than about 20%, and the soluble ions in the soil can correctly reflect the electrical conductivity of the soil. In the long-term observation, after irrigation or rainfall measured values are close to the true level. If the velocity measurement, first in the tested soil watering, to be full of water permeability were measured.

  1. Rapid measurement method: selected measurement locations, avoid the rocks, to ensure that the needle will not touch the stones like hard object, according to the required depth of cut open the surface soil, maintain the tightness degree of the original soil below the sensor body, clenched vertically inserted into the soil, can not be inserted before and after shaking, ensure the close contact with the soil. A measuring point within a small range test should repeatedly averaging.
  2. Buried in the underground measurement method: vertical drilling diameter greater than 20 cm depth of pit, according to the measurement needs, then the sensor wire inserted into the pit wall in a given level of depth, the pit landfill compaction, ensure the close contact with the soil. Stable after a period of time, can be last for days, months or even longer to measure and record.

If the surface measurement is hard, should first hole (diameter should be less than the diameter of the probe), and then inserted into the soil and the soil compaction and measurement; sensor should prevent violent vibration and impact, but not with a hard object percussion. Because the sensor for black package, in the strong sunlight will make the sensor to make sharp warming (up to over 50 ℃℃), in order to prevent the temperature measurement of high temperature impact sensor, please pay attention to sun protection in the field or fields.

5 Modbus Communication Protocol

5.1 Serial communication parameters

Modbus is a serial communication protocol, Modicon programmable logic controller (PLC) for the use of published. It has become the industry standard communication protocols, and is now quite common connection between industrial electronic equipment. Modbus has ex-tensive application in the industrial field. Modbus protocol is a master slave framework agreement. A node is the master node, other nodes using the Modbus protocol in communication from node. Each slave device has a unique address.

S-Temp&VWC&EC-02 sensor with RS485 interface, support Modbus protocol. The communication parameters to factory default values for: baud rate 9600 BPS, one start bit, 8 data bits, no parity, one stop bit. Communication protocol is Modbus RTU protocol. Communication parameters can be changed by the setup program or MODBUS command, after the communication parameters are changed, the sensor is required to re – enter the sensor to be effective.

5.2 Modbus Register

Parameter name   Register address

  (HEX / DEC)

  Parameter type   Modbus function

   number

  Parameter range and description
  TEMPRATURE   0x0000 /0   INT16, read   3/4   -4000-8000 corresponds to -40.00 ~

   80.00℃.

  VWC   0x0001 /1   UINT16, read   3/4   0-10000 corresponds to 0-100%
EC 0x0002 /2 UINT16, read   3/4   0-20000 corresponds to 0-20000us/cm
SALINITY 0x0003 /3   UINT16, read   3/4   0-20000 corresponds to 0-20000mg/L
  TDS   0x0004 /4   UINT16, read   3/4   0-20000 corresponds to 0-20000mg/L
  EPSILON    0x0005 /5   UINT16, read  3/4

 

 

  0-8200 corresponds to

0.00~82.00

  SOIL TYPE    0x0020 /32   UINT16, read-write   3/6/16   0-3

0: Mineral soil

1: sandy soil

2: clay

3: organic soil

  TEMP UNIT   0x0021 /33   UINT16, read-write    3/6/16   0:℃

1:℉

  EC&TEMP COFF   0x0022 /34   UINT16, read-write   3/6/16   0-100 corresponds to 0.0%-10.0%
  SALINITY COFF   0x0023 /35   UINT16, read-write   3/6/16   0-100 corresponds to 0.00-1.00
  TDS COFF   0x0024 /36   UINT16, read-write   3/6/16   0-100 corresponds to 0.00-1.00
  Modbus ADDRESS   0x0200 /512   UINT16, read-write   3/6/16   0-255
  BAUDRATE   0x0201 /513   UINT16, read-write   3/6/16     0-6

0:1200bps 1:2400bps 2:4800bps             3:9600bps 4:19200bps

  5:38400bps
PROTOCOL 0x0202 /514 UINT16, read-write 3/6/16   0~1

0: Modbus RTU

1: Modbus ASCII

  PARITY   0x0203 /515   UINT16, read-write   3/6/16   0-2

0: No parity bit

1: even parity check 2: Odd Parity bit

   DATABITS   0x0204 /516   UINT16, read-write   3/6/16    1

1:8 data bits

  STOPBITS   0x0205 /517  

3/6/16

  0-1

0:1 Stop bit

1:2 Stop bit

  RESPONSE    DELAY   0x0206 /518   UINT16, read-write   3/6/16   0-255 corresponds to the 0-2550   milliseconds sensor to receive the host   request for a period of time and then     the   delay response. The time delay for    setting the value of *10 milliseconds.         Set to 0 when no delay.
 ACTIVE   OUTPUT   INTERVAL   0x0207 /519   UINT16, read-write   3/6/16   0-255 corresponds to 0-255 seconds        does not require the host to request,        the   sensor to send data at a fixed time     interval. The time interval is set value * 1 second. Set to 0 when the active             output

function is prohibited.

5.3 Detail of Modbus Register

  TEMPERATURE
  Parameter range   -4000-8000 corresponds to -40.00~80.00℃   Default: none
  Parameter storage   none
  Meaning: the measured value of the temperature, negative for complement representation.

For example: if the return value is 0702H (of 16 hexadecimal, source code), the first byte is 07, the second byte and the low byte is   02h, then the temperature for the measured value (07H*256 02h) /100=17.94℃.If the return value is FF05H (16 – band, the complement), the first byte is FFH, low second byte is 05H, then temperature     measurement value ((FFH*256 05H) -FFFFH-1H) / 100 = FF05H-FFFFH-1H Celsius /100=-2.5℃.

  VWC — volumetric water content
  Parameter range   0-10000 corresponds to 0-100%   Default: none
Parameter storage   None
  Significance: volumetric water content measurements.For example: if the return value is 071DH (16 Decimal), the first byte of the high byte is 07H, second bytes of low byte is 1DH,      then the measured value is (1DH 07H*256) /10000= (7*256 29) =1821. representative volume water content is 18.21%.
  EC — electrical conductivity
  Parameter range   0-20000 corresponds to 0-20000us/cm   Default: none
  Parameter storage   None
    Significance: electrical conductivity measurement.

For example: if the return value is 071DH (in hexadecimal), the first byte is 07, the second byte and the low byte is 1dh, then          conductivity measurement value (07H*256 1dh) / 10000 = (7*256 29) on behalf of the =1821. soil conductivity 1821us/cm

SALINITY
  Parameter range   0-20000 corresponds to 0-20000mg/L   Default: none
  Parameter storage   None
  Significance: Salinity Measurement.

For example: if the value returned is 071DH (16 Decimal), the first byte of the high byte is 07H, the second byte low byte is   1DH, then the salinity measurement value (1DH 07H*256) /10000= (7*256 29)

=1821. on behalf of the soil salinity is 1821mg/L

  TDS— total dissolved solids
  Parameter range   0-20000 corresponds to 0-20000mg/L   Default: none
  Parameter storage   None
  Significance: TDS measurement value.

For example: if the value returned is 071DH (16 Decimal), the first byte of the high byte is 07H, second bytes of low byte is           1DH, then the TDS measurement value (1DH 07H*256) /10000= (7*256 29) =1821. on behalf of TDS 1821mg/L.

  EPSILON— dielectric constant
  Parameter range   0-8200   corresponds to 0.00-82.00   Default: none
  Parameter storage   None
  Meaning: dielectric constant.

For e xample: if the value returned is 071DH (16 Decimal), the first byte is 07H, the second byte low byte is 1DH, then the            measured value is (1DH 07H*256) /10000= (7*256 29) =1821. to represent the dielectric constant of 18.21.

TEMP UNIT— degree unit
  Parameter range   0:℃

1:℉

  Default: 0
  Parameter storage   None
  Significance: unit of temperature.
  EC TEMP COFF
  Parameter range   0-100   corresponds to 0.0%-10.0%   Default: 20(2%)
  Parameter storage   None
  Significance: the temperature compensation coefficient of electrical conductivity
  SALINITY COFF
  Parameter range   0-100 corresponds to 0.00-1.00   Default: 55(0.55)
  Parameter storage   None
  Significance: Salinity / conductivity compensation coefficient
  TDS COFF
  Parameter range   0-100 corresponds to 0.00-1.00   Default: 50(0.50)
  Parameter storage   None
  Significance: TDS/ conductivity compensation coefficient
  SLAVE ADDR — Modbus address
  Parameter range   0-255   Default:1
  Parameter storage   Immediate storage
  Modbus address can be set to 0-255. When outside of the module address the dip switch setting to address 0, using the             contents   of the register as a slave address. After setting need to re power or use the rst command restart module, the entry into    force of this address. The use of the command to

change the module address does not need to open the cabinet can be arranged.

  BAUDRATE
  Parameter range   0-5

0:1200bps 1:2400bps 2:4800bps 3:9600bps

  Default:3
  4:19200bps

5:38400bps

  Parameter storage   Immediate storage
  PROTOCOL — Serial communication Protocol
  Parameter range   0~1

0:Modbus RTU 1:Modbus ASCII

  Default:0
   Parameter storage   Immediate storage
  PARITY — Serial communication Check bit
  Parameter range   0-2

0:none

1: even parity check 2: Odd parity check

  Default:0
  Parameter storage   Immediate storage
  DATA BITS
  Parameter range   1

1:8 data bits

  Default:1,Only supports 8    data bits,

the other is invalid

  Parameter storage   Immediate storage
STOP BITS
Parameter range   0-1

0:1 stop bit

1:2 stop bits

  Default:0
Parameter storage Immediate storage
  RESPONSE DELAY
  Parameter range   0-255 Default:0
  Parameter storage   Immediate storage
  Serial communication delay response used in the following circumstances: when the host sends a request command, delay          module (RESPONSEDELAY*10 milliseconds), then the response data is returned to the host. For example, to set up                RESPONSEDELAY=5, so delay module 5*10=50 millisecond response requesting host. Set to 0 for no delay an immediate     response. This command is mainly used to host from RS485 transmission switch state to the receiving state relatively slow speed of occasions.
  ACTIVE OUTPUT INTERVAL
  Parameter range   0-255   Default:0
  Parameter storage   Immediate storage
Serial communication active output time interval used in the following circumstances: hosts that do not need to send a request command module active output response data and output interval for ACTIVEOUTPUTINTERVAL second, such as setting ACTIVEOUTPUTINTERVAL=5. So module every 5 seconds according to set up the communication protocol of a debate output data. Set to 0 when the active output is invalid, the main request before response. This command is mainly used in GPRS wireless transmission, terminal active node data transmission occasions.Note: when the active output data is set, only one module can be connected on the RS485 bus.

5.4 Communication Sample

In the following instructions, the data at the beginning of the 0x or the ending of the H is a 16 – band data. Modbus protocol with two common types of registers:(1) To maintain the register, storage data is not lost, it is read and write. Usually with function number 3 (0x03) read, use function number 6 (0x06) or 16 (0x10) write.(2) The input registers are used to store a number of read – only physical variables, such as temperature values, that are read – only and usually read with a function number 4 (0x04).

5.4.1 Function number 3 communication sample

Common request format:AA 03 RRRR NNNN CCCC

  AA   1 byte   Address, 0-255
  03   1byte   Function number 3
  RRRR   2byte   Start register address, high byte in front
  NNNN   2byte   read the number of registers N, high byte in the front
  CCCC   2byte   CRC CHECK

Common request format:AA 03 MM VV0 VV1 VV2 VV3… CCCC

  AA   1byte   Address, 0-255
  03   1byte   Function number 3
MM   1byte   Returns the number of data byte in the register value
VV0,VV1   2byte   Returns the first register value
  VV2,VV3   2byte   Returns the second register value
  Returns the “N” register value (N=MM/2)
  CCCC   2byte   CRC CHECK

For example: to read register 0x0200-0x0201, namely from the machine address and baud rate for example

  Address   1byte   0x01
  Function number   1byte   0x03
  Start register

address

  2byte   0x0200
  Register number   2byte   0x0002
  Check   2byte   0xC5B3
Respond:01 03 04 00 01 00 03 EB F2
  Address   1byte   0x01
  Function number   1byte   0x03
  Effective byte

number

  1byte   0x04
  Slave address

register value

  2byte   0x00 (From machine address high byte)
  0x01 (From machine address low byte)
  The baud rate

register value

  2byte   0x00 (High baud rate byte)
  0x03 (low baud rate byte)
  Check   2byte   0xEBF2
5.4.2 Function number 4 communication sample

Common request format:AA 04 RRRR NNNN CCCC

  AA   1byte   Address, 0-255
  04   1byte   Function number4
  MM   1byte   Returns the number of data byte in the register value
  VV0,VV1   2byte   Returns the first register value
  VV2,VV3   2byte   Returns the second register value
  Returns the “N” register value (N=MM/2)
  CCCC   2byte   CRC CHECK
Common request format:AA 04 RRRR NNNN CCCC
  Address   1byte   0x01
  Function number   1byte   0x04
  Start register

address

  2byte   0x0000
  Register number   2byte   0x0003
  Check   2byte   0xB00B

Common request format:AA 04 MM VV0 VV1 VV2 VV3… CCCC

  Address   1byte   0x01
  Function number   1byte   0x04
  Effective byte

number

  1byte   0x06
  Temperature

register value

  2byte   0x08
  0x90
  Volume water

content register value

  2byte   0x0E
  0x93
  Conductivity

register value

  2byte   0x02
  0x4E
  Check   2byte   0xD257

For example: to read the register 0x0000-0x0003, that reads the temperature, water content, electrical conductivity valueAsk:01 04 0000 0003 B00B

  AA   1byte   Address, 0-255
06   1byte   Function number6
  RRRR   2byte   Register address, high byte in front
  VVVV   2byte   To write the value of the register, the high byte is vin the front

Common request format:AA 06 RRRR VVVV CCCC

  AA   1byte   Address, 0-255
  06   1byte   Function number6
  RRRR   2byte   Register address, high byte in front
  VVVV   2byte   To write the value of the register, the high byte is

in the front

  CCCC   2byte   CRC CHECK

Respond:01 04 06 08 90 0E 93 02 4E D2 57

  Address   1byte   0x01
  Function number   1byte   0x06
  Start register

address

  2byte   0x0021
  Register number   2byte   0x0001
  Check   2byte   0x1800

5.4.3 Function number 6 communication sample

Common request format:AA 06 RRRR VVVV CCCC

  AA   1byte   Address, 0-255
  10 (HEX)   1byte   Function number16 (10 binary system)
  RRRR   2byte   Start register address, High byte in front
  NNNN   2byte   To read the number N Register, high byte in the

front

  MM   1byte   The number of byte to write the value of the

register

 VVVV1   2byte   To write the value of the first register, the high

byte is in the front.

 VVVV2   2byte   To write the value of the second register, the high

byte is in the front.

To write the value of the “N” register, the high byte

is in the front. N=MM/2

CCCC   2byte CRC CHECK

For example: to write register 0x0021, namely the temperature unit for Fahrenheit cases Ask:01 06 0021 0001 1800

  Address   1byte   0x01
  Function number   1byte   0x06
Start register

address

  2byte   0x0021
Register number   2byte   0x0001
Check   2byte   0x1800

Respond:01 06 0021 0001 1800

  Address   1byte   0x01
  Function number   1byte   0x06
  Start register

address

  2byte   0x0021
  Register number   2byte   0x0001
  Check   2byte   0x1800

For example: to write register 0x0021, namely the temperature unit for Fahrenheit casesAsk:01 06 0021 0001 1800

5.4.4 Function number 16 communication sample

Common request format:AA 10 RRRR NNNN MM VVVV1 VVVV2 …CCCC

  AA   1byte   Address, 0-255
  10 (HEX)   1byte   Function number16 (10 binary system)
  RRRR   2byte   Start register address, High byte in front
  NNNN   2byte   To read the number N Register, high byte in the

front

MM 1byte   The number of byte to write the value of the

register

  VVVV1   2byte   To write the value of the first register, the high

byte is in the front.

VVVV2   2byte   To write the value of the second register, the high

byte is in the front.

  To write the value of the “N” register, the high byte

is in the front. N=MM/2

  CCCC   2byte   CRC CHECK

Common request format:AA 10 RRRR NNNN CCCC

  AA   1byte   Address, 0-255
  10(HEX)   1byte   Function number16 (10 binary system)
  RRRR   2byte   Start register address, High byte in front
  NNNN   2byte   To read the number N Register, high byte in the

front

  CCCC   2byte   CRC CHECK
Common request format:AA 10 RRRR NNNN CCCC
AA 1byte Address, 0-255
10(HEX) 1byte Function number16 (10 binary system)
RRRR   2byte Start register address, High byte in front
NNNN   2byte To read the number N Register, high byte in the

front

CCCC   2byte CRC CHECK
  AA   1byte   Address, 0-255
  10(HEX)   1byte   Function number16 (10 binary system)
  RRRR   2byte   Start register address, High byte in front
  NNNN   2byte   To read the number N Register, high byte in the

front

  CCCC   2byte   CRC CHECK

For example: to write register 0x0200-0x0201 is set from the machine address is 1, the baud rate is 19200bps as an exampleAsk:01 10 0200 0002 04 0001 0004 BACC

  0x01   1byte   Address
  0x10 (HEX)   1byte   Function number 16 (DEC)
  0x0200   2byte   Start register address, High byte in front
0x0002   2byte   To read the number N Register, high byte in the

front

  0x04   1byte   The number of byte to write the value of the

register

  0x0001 2byte   To write such as from the station address register

value is 1

  0x0004   2byte   To write such as from the station address register
  value is 4
  0xBACC   2byte   CRC CHECK

Respond:01 10 0200 0002 4070

  0x01   1byte   Address
  0x10 (HEX)   1byte   Function number16 (DEC)
  0x0200   2byte   Start register address, high byte in the front
  0x0002   2byte   To read the number N Register, high byte in the

front

  0x4070   2byte   CRC CHECK

5.5 CRC16 Check Algorithm

// CRC calculation of C51 language function is as follows// Enter the parameter 1:snd, to be the name of the byte Check array// Input parameters 2:num, the total number of Check to be byte// Function return value: Check and//—————————————————————————–unsigned int calc_crc16 (unsigned char *snd, unsigned char num){unsigned char i, j;unsigned int c,crc=0xFFFF;for(i = 0; i < num; i ++){c = snd[i] & 0x00FF;crc ^= c;for(j = 0;j < 8; j ++){if (crc & 0x0001){crc>>=1;crc^=0xA001;}else{crc>>=1;}}}return(crc);

For example: to read the register 0x0000-0x0002, that reads the temperature, water content, electrical conductivity valueHost Ask:01 0400000003 B00B (8 byte)

  Address   1byte   0x01
  Function number   1byte   0x04
  Start register address   2byte   0x0000
  Register number   2byte   0x0003
  Check   2byte   0xB00B

When the host needs to send data to the sensor, it will need to send Check data stored in the snd array. (01 04 00 00 00 03 A total of 6 byte), Among them num=6Pseudo code as follows:unsigned char request[8]={01,04,00,00,00,03,00,00};// The last two 00,00 are CHECK CRCunsigned char num=6;// Calculate the array of the first 6 CRC CHECK byteunsigned int crc16=0;crc16= calc_crc16 (request, num);request[6]= crc16%256;// Store check CRC in an array to be sentrequest[7]= crc16/256;Comm Port .Send(request, 8);// Send data through serial port

Sensor Respond:01 04 06 08 90 0E 93 02 4E D2 57 (11 byte)
  Address   1byte   0x01
  Function number   1byte   0x04
  Effective byte number   1byte   0x06
  Temperature register value   2byte   0x08
  0x90
  Volume water content register value   2byte   0x0E
  0x93
  Conductivity register value   2byte   0x02
  0x4E
  Check   2byte   0xD257

When the host receives the 11 byte data returned by the sensor, the following CRC calculation is performed, where num=11Pseudo code as follows:unsigned char response[11]={ 01 04 06 08 90 0E 93 02 4E D2 57};// The last two byte are the CHECK CRC that the sensor returnsunsigned char num=11;// Calculate the entire return of the 11 CRC CHECK byteunsigned int crc16=0;crc16= calc_crc16 (response, num);if(crc16==0){// Check CRC correctly, you can use the returned data }else{// Check CRC error, can not be used to return the data }To get results back to 0 so the success of Check, if Check fails to return to a nonzero value.If the Check does not succeed, it shows that the transmission process is wrong, should give up the collected data, re collection.The success of the Check, use the following formula to calculate the temperature (negative to complement representation) and conductivity of H at the end of the 16 hexadecimal data:temperature=(08H*256+90H)/100=2192/100=21.92 ℃volumetric water content =(0EH*256+93H)/100=3731/100=37.31%conductivity =02H*256+4EH=2*256+78 =590 μs/cm

 