Thermofocus ®0700A2 model01500A3 modelNon contact thermometer
USER MANUALRead carefully the instruction before using the thermometer
See also the multi-languagewww.tecnimed.it/movie.html
THERMOFOCUS®0700A2 / 01500A3 mod.Class IIa medical device, clinically testedPatents no. MI 1.284.119,EP 0909.377, US 6.196.714, IL127.876, JP 504769/98, EP1.283.983, US 7.001.066, and others international Patents Pending
0051Manufactured in Italy byTECNIMED srl12, P.le Cocchi21040 Vedano O. (VA) – ITALY[email protected]www.tecnimed.comTel. +39 0332 402350
TECHNICAL CHARACTERISTICSTHERMOFOCUS®0700A2 mod.01500A3 mod.Non-contact infrared thermometerResolution: 0.1Body temperature readingsMeasurement range: 34.0/42.5°C (93.2-108.5°F)Room temperature working range: 10/40°C 55/104.0°F
AccuracyºC | from 34.0 to 35.9°C: | ±0.3°C |
from 36.0 to 39.0°C: | ±0.2°C (1) | |
from 39.1 to 42.5°C: | ±0.3°C |
Accuracy °F | from 93.2 to 96.7°F: | ±0.5°F |
from 96.8 to 102.2°F: | ±0.4°F (1) | |
from 102.3 to 108.5: | ±0.5°F |
Other readings(2)Measurement range: 1.0/55.0°C (33.8-131°F)Room temperature working range: 10/40°C 55/104.0°F (3)
Accuracy °C | from 1,0 to 19,9°C: | ±1,0°C |
from 20,0 to 35,9°C: | ±0,3°C | |
from 36,0 to 39,0°C: | ±0,2°C | |
from 39,1 to 42,5°C: | ±0,3°C | |
from 42,5 to 55°C: | ±1,0°C |
Accuracy
°F |
from 33.8 to 67.9°F | ±1.8°F |
from 68.0 to 96.7°F | ±0.5°F | |
from 96.8 to 102.2 °F | ±0.4°F | |
from 102.3 to 108.5°F | ±0.5°F | |
from 108.7 to 131 °F | ±1.8°F |
(1)The ASTM E1965-98-2009 standard requires ±0.2°C (±0.4°F) accuracy for infrared thermometers in the 37-39°C (98.6-102.2°F) temperature range while the ASTM E667-86 and E1112-86 standards require ±0.1°C (±0.2°F) precision for mercury and electronic thermometers in that same temperature range.2)Only with the 0700A2 model: subtract 2°C (3°F) from the temperature shown on the display to get a rough value.If the thermometer is set in rectal or internal temperature, subtract 3°C (5°F).3) Thermofocus can also take surface temperatures in rooms where the temperature is lower than 10°C (55°F) although accuracy and the operating range are not guaranteed and the display shows “Lo.5” and the value alternately.
Power supply: | 4 AAA (LR03) alkaline batteries – 1.5 V |
Life of high-quality batteries: | up to 3 years or 10,000 readings (depending on use) |
Dimensions: | m 165 x 40 x 22 (6.5 x 1.45 x 0.86 inches) – including cap |
Weight: | gr. 92 (3.17 oz.) – batteries included |
Distance from the subject: calculated using an optical aiming system (approximately 3 cm / 1.2 inches).Large, clearly visible display.Atmospheric pressure range of operating conditions: from 700 hPa to 1,060 hPa. Relative humidity range of operating conditions:from 15 % to 93 %, non-condensing.Keep in a clean, dry place, preferably at a temperature ranging between +16 and +40°C (60.8 and 104°F). Store at a temperature included between -10 and +60°C (14 and 140°F) and in any case not lower than -18°C (0.4°F) or higher than +70°C (158°F).Expected life: 10 years.Their focus is a class II medical device (IIa as per 93/42/EEC Directive and subsequent amendments and integrations) tested in hospitals, private clinics, and medical offices.Contact the manufacturer for clinical tests.Tecnimed Srl, P.le Cocchi, 12 – 21040 Vedano Olona (VA) – Italy undertakes full responsibility for this product’s compliance with the reference standards.Their focus is compliant with the pertinent ASTM standard (E1965-98:2009) and CEI EN 60601-1, CEI EN 60601-1-11 standards.Internally powered equipment for continuous operation.The aiming LEDs of Thermofocus emit low luminous radiation largely under the level foreseen from IEC 60825- 1:1993+A1:1997+A2:2001 Standard for Class I lights (class I, wavelength: 610 nm; maximum power: Pmax<1mW).
Follow instructions for use | |
ATTENTION: Read the warnings | |
A device designed for continuous use | |
Applied part: type BF | |
Recyclable material |
Dispose of the device in compliance with applicable legislation regarding electrical equipment and batteries.