ThermoWorks IFST Infrared Food Safety Thermometer
Operating Instructions
The thermometer is a non-contact infrared thermometer, and always with Real-Time Maximum (MAX) data. Please remember to keep away from children and don’t use it for safety related applications.
(Default Screen) |
|
ADD VALUE
°C or °F |
Press Mode key () for °C or °F switching. |
LCD Backlight: always on. | |
ClassII Laser: always enabled. |
EMC/RFIReadings may be affected if the unit is operated within radio frequency electromagnetic field strength of approximately 3 volts per meter, but the performance of the instrument will not be permanently affected.* Note: Under the electromagnetic field of 3V/m from 350 to 550MHZ, the maximum error is 8°C (14.4°F).
- WHEN DEVICE IS IN USE, DO NOT LOOK DIRECTLY INTO THE LASER BEAMPERMANENT EYE DAMAGE MAY RESULT.
- USE EXTREME CAUTION WHEN OPERATING THE LASER.
- NEVER POINT THE DEVICE TOWARDS ANYONE’S EYES.
- KEEP OUT OF REACH OF ALL CHILDREN.
STORAGE & CLEANING
The sensor lens is the most delicate part of the thermometer. The lens should be kept clean at all times, care should be taken when cleaning the lens using only a soft cloth or cotton swab with water or medical alcohol. Allowing the lens to fully dry before using the thermometer. Do not submerge any part of the thermometer. The thermometer should be stored at room temperature between 20 to +65°C (-4~149°F).
LCD ERROR MESSAGES
The thermometer incorporates visual diagnostic messages as follows:
Er2’ is displayed when the thermometer is exposed to rapid changes in the ambient temperature. ‘Er3’ is displayed when the ambient temperature exceeds 0o C(32°F) or +50o C (122°F). | |
The thermometer should be allowed plenty of time (minimum 30 minutes) to stabilize to the working/room temperature. | |
For all other error messages it is necessary to reset the thermometer. To reset it, turn the instrument off, remove the battery and wait for a minimum of one minute, reinsert the battery and turn on. If the error message remains please contact the Service Department for further assistance. |
BATTERIES
The thermometer incorporates visual low battery indication as follows:
|
|
|
SPECIFICATION
Measurement Range | –67 to 536°F (–55 to 280°C) |
Operating Range | 32 to 122°F (0 to 50°C) |
Accuracy, Tobj =0 to 65°C, Tamb=23±2°C | 0 to 65°C= ±1°C (1.8°F) |
Accuracy Tamb=23±2°C | <0°C=±1°C (±2°F) ±0.1 degree/degree>149°F (65°C)=1.5% of reading |
Emissivity | 0.97 Fixed |
Resolution (–9.9 to 199.9°C) | 0.1°C/°F |
Response Time (90%) | 1sec |
Distance: Spot | 12:1 |
Battery Life | Typ. 180, min 140 hours continuous use |
Dimensions | 175.2 x 39.0 x 71.9 mm |
Weight | 179 grams including batteries (AAA*2pcs) |
** The thermometer will automatically shut off if left idle for more than 15sec. |
CERTIFICATE OF CONFORMANCEThe manufacturer of this instrument has implemented a quality assurance system under ISO 9001:2000 certified quality system and fully follow ISO GUM (Guide to the Expression of Uncertainty in Measurement) to evaluate the uncertainty of temperature and resistance standards, guarantee performance as below;
Note: Assume the operation ambient temperature under 23 ± 3°C Furthermore, we certify that this infrared thermometer has been inspected and found to comply with published specifications. This device has been calibrated by temperature and/or resistance standards that are traceable to NIST (U.S. National Institute of Standards and Technology) or NML National Measurement Laboratory) of Taiwan, and the calibration procedure corresponds with generally accepted regulations and standards. ThermoWorks, Inc.741 E. Utah Valley Dr. American Fork, UT 84003www.thermoworks.com |